510(k) Report

510(k) · Cleared

Recalled · Class I · 2012

Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40

K112114510(k) TraditionalCleared Jan 13, 2012172-day reviewClass II21 CFR 868.5160Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet Critical Care AB · BSZ Gas-Machine, Anesthesia

Cleared on Jan 13, 2012 as substantially equivalent to K102182 (Maquet Critical Care AB, 2011) and 1 other predicate after a 172-day review. Its family tree reaches back 9 generations to 1994. 4 of the 39 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Jan 13, 2012
Review
172 days
Class
Class II
Product code
BSZ
Recalls
1 (1 Class I)
Predicates
2

Its longest line of descent runs back 9 generations to K930017 (1994). 35 ancestors in all, the oldest K811102 (1981). 3 later devices descend from it, 1 citing it directly. 4 of the 39 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2004: 2 other ancestors2002: 1 other ancestor2000: 2 other ancestors1999: 1 other ancestor1997: 1 other ancestor1991: 1 other ancestor1989: 2 other ancestors1981: 1 other ancestor1996: 3 other ancestors1995: 2 other ancestors1993: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor1985: 2 other ancestors1984: 2 other ancestors1998: 2 other ancestors2014: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall1K930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 1994K023366 · Datex-Ohmeda 7900 Ventilator Enhancements · Datex-Ohmeda, Inc. · 2003 · Class I recall×2: 2 generations cleared close together, K040743 (2004), K042154 (2004) K061609 · GE Datex-Ohmeda Aisys · Datex-Ohmeda, Inc. · 2006 · recalledK071142 · GE Datex-Ohmeda Avance Anesthesia System · Datex-Ohmeda · 2007K081844 · GE Datex-Ohmeda Avance · Datex-Ohmeda, Inc. · 2008 · recalledK090233 · GE Datex-Ohmeda Aisys · Datex-Ohmeda, Inc. · 2009 · recalledK102182 · Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30 · Maquet Critical Care AB · 201119801990200020102020
  1. Root · 9 generations backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. 5 generations between
  3. 3 generations backK081844 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · Recalled
  4. 2 generations backK090233 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · Recalled
  5. 1 generation backK102182 Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30Maquet Critical Care AB · 2011
  6. This deviceK112114 Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012
  7. Descendant · cites it directly · 2 of its ownK133958 Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · Recalled (Class I)
  8. Descendant · 2 generations on · 1 of its ownK160665 FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i…Maquet Critical Care AB · 2017
  9. Descendant · 3 generations on · 0 of its ownK191027 Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e…Maquet Critical Care AB · 2020 · Recalled (Class I)
Show all 9 generations
  1. Root · 9 generations backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. 8 generations backK023366 Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Recalled (Class I)
  3. 7 generations backK040743 Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc. · 2004
  4. 6 generations backK042154 GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc. · 2004 · Recalled
  5. 5 generations backK061609 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2006 · Recalled
  6. 4 generations backK071142 GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda · 2007
  7. 3 generations backK081844 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · Recalled
  8. 2 generations backK090233 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · Recalled
  9. 1 generation backK102182 Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30Maquet Critical Care AB · 2011
  10. This deviceK112114 Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · cites it directly2descendants
  • K160665FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i Anesthesia System C40Maquet Critical Care AB · 2017 · 2 generations on1descendant
  • K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB · 2020 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K930017PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp.1994—
−8K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc.2003Class I
−7K040743Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc.2004—
−6K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc.20043
−5K061609GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20061
−4K071142GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda2007—
−3K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.20084
−2K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20092
−1K102182Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30Maquet Critical Care AB2011—
0 (this device)K112114Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB2012—
+1K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB2014Class I
+2K160665FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i Anesthesia…Maquet Critical Care AB2017—
+3K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB2020Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198420
198520
198610
198710
198920
199110
199310
199520
199630
199710
199820
199910
200020
200210
200420
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201411
201710
202011
FIG. 1 — Lineage of K112114, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102182Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30
Summary page 3
…to Predicate Devices The indications for use and fundamental technology for the modified FLOW-i Anaesthesia System is identical to the cleared FLOW-i Anaesthesia System K1 02182. Comparison of Intended Use The Intended Use for the modified FLOW-i Anaesthesia System (KI 12114) and the predicate devices FLO…
Maquet Critical Care AB2011—
*K090233GE Datex-Ohmeda Aisys
Summary page 1
…being claimed Maguet FLOW-i Anesthestia System GE Datex-Ohmeda, Aisys Anaesthesia System jKjK090233 510(k) #…
Datex-Ohmeda, Inc.20092

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133958Flow-I Anesthesia System C20, C30 & C40 Class IMaquet Critical Care AB20142

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 20, 2012Class IZ-1925-2012Device DesignA technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 11 months on median.
  • 12 of the 35 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K133958 (Maquet Critical Care AB, 2014) has gone furthest, with 2 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The indication for FLOW-i Anesthesia System is administering inhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRIf environment, by healthcare professionals trained in inhalation Anesthesia administration.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112114: Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112114

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026