510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Fourth Gas Module with CO2 for Primus US Apollo

K081447510(k) TraditionalCleared Oct 8, 2008139-day reviewClass II21 CFR 868.5160Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical AG & Co. KG · BSZ Gas-Machine, Anesthesia

Cleared on Oct 8, 2008 as substantially equivalent to K000706 (Datex-Ohmeda, Inc., 2000) and 1 other predicate after a 139-day review. Its family tree reaches back 1 generation to 1998. It has been recalled once.

Decided
Oct 8, 2008
Review
139 days
Class
Class II
Product code
BSZ
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K973896 (1998). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2000: 1 other ancestorK973896 · Ohmeda Excel 3000 Anesthesia System · Ohmeda Medical · 1998 · recalled1995200020052010
Other ancestors, by generation 1 generation back: 1 other ancestor K973896 · Ohmeda Excel 3000 Anesthesia System · Ohmeda Medical · 1998 · recalled0
  1. Root · 1 generation backK973896 Ohmeda Excel 3000 Anesthesia SystemOhmeda Medical · 1998 · Recalled
  2. This deviceK081447 Fourth Gas Module with CO2 for Primus US ApolloDraeger Medical AG & Co. KG · 2008
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K973896Ohmeda Excel 3000 Anesthesia SystemOhmeda Medical19981
0 (this device)K081447Fourth Gas Module with CO2 for Primus US ApolloDraeger Medical AG & Co. KG2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
FIG. 1 — Lineage of K081447, 1998–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BSZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K000706Datex-Ohmeda Aestiva/5 with 7100 Ventilator
Summary page 2
…the following devices: DEVICE 510(k) NUMBER Aestiva/5 Anesthesia System K000706 Ohmeda Excel 3000 Anesthesia System K973896 The Fourth Gas Module with C02 for Primus US "Apollo"…
Datex-Ohmeda, Inc.20004
*K973896Ohmeda Excel 3000 Anesthesia System
Summary page 2
…510(k) NUMBER Aestiva/5 Anesthesia System K000706 Ohmeda Excel 3000 Anesthesia System K973896 The Fourth Gas Module with C02 for Primus US "Apollo" provides, as does the Aestive/5 or the Excel 3000, a mechanical means of…
Ohmeda Medical19981

Also cleared against K000706: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2018Class IIZ-2253-2018Process change controlThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K000706.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Primus US "Apollo" is indicated as a continuous flow anesthesia system. The Primus US "Apollo" may be used for manually assisted, or automatic ventilation, and delivery of gases (02, N 2 0 and CO 2 - in combination with the CO 2 module [i.e. Fourth Gas Module]), anesthetic vapor, and monitoring of oxygen and CO 2 concentrations, breathing pressure, respiratory volume, and anesthetic agent identification and concentration. Federal law restricts this device to sale by or on the order of a physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K081447: Fourth Gas Module with CO2 for Primus US Apollo.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K081447

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026