Datex-Ohmeda Aestiva/5 with 7100 Ventilator

K000706510(k) SpecialCleared May 11, 200070-day reviewClass II21 CFR 868.5160Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on May 11, 2000 after a 70-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 46 of the 159 devices in its family, counting itself, have a Class I recall. 158 later devices descend from it. It has been recalled 4 times.

Decided
May 11, 2000
Review
70 days
Class
Class II
Product code
BSZ
Recalls
4
Predicates
0

158 later devices descend from it, 4 citing it directly. 46 of the 159 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2002: 2 descendants cleared, 1 with a Class I recall12004: 5 descendants cleared2017: 12 descendants cleared, 5 with a Class I recall52008: 11 descendants cleared, 1 with a Class I recall12012: 14 descendants cleared, 6 with a Class I recall62007: 4 descendants cleared, 1 with a Class I recall12006: 5 descendants cleared2005: 3 descendants cleared2011: 14 descendants cleared, 8 with a Class I recall82014: 7 descendants cleared, 2 with a Class I recall22009: 7 descendants cleared, 1 with a Class I recall12010: 6 descendants cleared, 1 with a Class I recall12025: 4 descendants cleared2021: 7 descendants cleared, 3 with a Class I recall32015: 9 descendants cleared, 1 with a Class I recall12013: 11 descendants cleared, 4 with a Class I recall42016: 8 descendants cleared, 3 with a Class I recall32019: 7 descendants cleared, 3 with a Class I recall32024: 5 descendants cleared2018: 4 descendants cleared, 1 with a Class I recall12026: 5 descendants cleared2020: 5 descendants cleared, 4 with a Class I recall42023: 2 descendants cleared, 1 with a Class I recall12022: 1 descendant cleared200020052010201520202025
  1. This deviceK000706 Datex-Ohmeda Aestiva/5 with 7100 VentilatorDatex-Ohmeda, Inc. · 2000
  2. Descendant · cites it directly · 133 of its ownK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  3. Descendant · 2 generations on · 114 of its ownK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  4. Descendant · 3 generations on · 112 of its ownK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  5. Descendant · 3 generations on · 87 of its ownK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  6. Descendant · 4 generations on · 86 of its ownK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006

Showing 5 of 158 descendants. See all descendants

Descendants cleared per year

2002: 22004: 52005: 32006: 52007: 42008: 112009: 72010: 62011: 142012: 142013: 112014: 72015: 92016: 82017: 122018: 42019: 72020: 52021: 72022: 12023: 22024: 52025: 42026: 5
2002high 14 (2011) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 158 · See all
  • K013642AVEABird Products Corp. · 2002 · cites it directly133descendants
  • K032451Bird Tbird Vela VentilatorBird Products Corp. · 2004 · 2 generations on114descendants
  • K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005 · 3 generations on112descendants
  • K053270IVENT 201Versamed Medical Systems, Inc. · 2006 · 3 generations on87descendants
  • K061627Modification to I Vent 201Versamed Medical Systems, Inc. · 2006 · 4 generations on86descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K000706Datex-Ohmeda Aestiva/5 with 7100 VentilatorDatex-Ohmeda, Inc.2000—
+1K013642AVEABird Products Corp.2002Class I
+2K032451Bird Tbird Vela VentilatorBird Products Corp.20041
+3K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
+3K053270IVENT 201Versamed Medical Systems, Inc.2006—
+4K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200221
200450
200530
200650
200741
2008111
200971
201061
2011148
2012146
2013114
201472
201591
201683
2017125
201841
201973
202054
202173
202210
202321
202450
202540
202650
FIG. 1 — Lineage of K000706, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

46 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (158)

See all 158

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K013642AVEA Class IBird Products Corp.2002133
K032803Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda200439
K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire… Class IDatex-Ohmeda, Inc.20170
K081447Fourth Gas Module with CO2 for Primus US ApolloDraeger Medical AG & Co. KG20080

Has anything in its family been recalled? (46 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 19, 2019Class IIZ-0116-2020Software in the Use EnvironmentCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that…
Jul 19, 2019Class IIZ-0117-2020Software in the Use EnvironmentCertain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that…
Sep 27, 2005Not classifiedZ-0037-06OtherTwo conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen…
Apr 16, 2004Not classifiedZ-0968-04OtherIncreased circuit pressure as a result of a misassembled Aestiva/5 Active Gas Scavenging System.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K013642AVEABird Products Corp. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 113 Class I recalls
  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 3 generations on · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K093261MOVESThornhill Research, Inc. · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K131774Hamilton-G5Hamilton Medical AG · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K131252Puritan Bennett 980 Series Ventilator SystemCovidien · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 13 Class I recalls
  • K072885Carina System, Model 8419350; Carina Basic Device, Model 5704110Draeger Medical AG & Co. KG · 2008 · 5 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · 5 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 46 shown.

What does this lineage say?

  • Of its descendants, K013642 (Bird Products Corp., 2002) has gone furthest, with 133 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—
Cite this page

510(k) Report. “K000706: Datex-Ohmeda Aestiva/5 with 7100 Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K000706

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026