510(k) Report

510(k) · Cleared

Recalled · 2011

Amsco V-Pro 1 Low Temperature Sterilization System

K083097510(k) TraditionalCleared Aug 3, 2009290-day reviewClass II21 CFR 880.6860Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MLR Sterilizer, Chemical

Cleared on Aug 3, 2009 as substantially equivalent to K062297 (STERIS Corporation, 2007) after a 290-day review. Its family tree reaches back 1 generation to 2007. 27 later devices descend from it. It has been recalled once.

Decided
Aug 3, 2009
Review
290 days
Class
Class II
Product code
MLR
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K062297 (2007). 27 later devices descend from it, 6 citing it directly.

Descendants, by year cleared 2011: 3 descendants cleared2013: 1 descendant cleared2014: 3 descendants cleared2012: 1 descendant cleared2016: 3 descendants cleared2021: 1 descendant cleared2019: 5 descendants cleared2018: 4 descendants cleared2017: 1 descendant cleared2023: 3 descendants cleared2022: 2 descendants clearedK062297 · VHP 136 Low Temperature Sterilization System · STERIS Corporation · 2007 · recalled20052010201520202025
Descendants, by generation 1 generation forward: 6 descendants2 generations forward: 5 descendants3 generations forward: 7 descendants4 generations forward: 3 descendants5 generations forward: 2 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants K062297 · VHP 136 Low Temperature Sterilization System · STERIS Corporation · 2007 · recalled0+5
  1. Root · 1 generation backK062297 VHP 136 Low Temperature Sterilization SystemSTERIS Corporation · 2007 · Recalled
  2. This deviceK083097 Amsco V-Pro 1 Low Temperature Sterilization SystemSTERIS Corporation · 2009
  3. Descendant · cites it directly · 18 of its ownK111810 Amsco V-Pro 1 and V-Pro 1 Plus Low Temperature Sterilization UnitsSTERIS Corporation · 2011
  4. Descendant · cites it directly · 18 of its ownK102394 Amsco V-Pro 1 Low Temperature Sterilization System; Amsco V-Pro 1…STERIS Corporation · 2011
  5. Descendant · cites it directly · 17 of its ownK120632 Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation · 2012
  6. Descendant · cites it directly · 16 of its ownK131120 Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature…STERIS Corporation · 2013
  7. Descendant · 2 generations on · 8 of its ownK160433 V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation · 2016

Showing 5 of 27 descendants. See all descendants

Descendants cleared per year

2011: 32012: 12013: 12014: 32016: 32017: 12018: 42019: 52021: 12022: 22023: 3
2011high 5 (2019) · low 0 (2015)2026

Most influential descendants

Showing 5 of 27 · See all
  • K111810Amsco V-Pro 1 and V-Pro 1 Plus Low Temperature Sterilization UnitsSTERIS Corporation · 2011 · cites it directly18descendants
  • K102394Amsco V-Pro 1 Low Temperature Sterilization System; Amsco V-Pro 1 Plus Low Temperature…STERIS Corporation · 2011 · cites it directly18descendants
  • K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation · 2012 · cites it directly17descendants
  • K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2013 · cites it directly16descendants
  • K160433V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation · 2016 · 2 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K062297VHP 136 Low Temperature Sterilization SystemSTERIS Corporation20071
0 (this device)K083097Amsco V-Pro 1 Low Temperature Sterilization SystemSTERIS Corporation2009—
+1K111810Amsco V-Pro 1 and V-Pro 1 Plus Low Temperature Sterilization UnitsSTERIS Corporation2011—
+1K102394Amsco V-Pro 1 Low Temperature Sterilization System; Amsco V-Pro 1 Plus Low…STERIS Corporation2011—
+1K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation2012—
+1K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2013—
+2K160433V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201130
201210
201310
201430
201630
201710
201840
201950
202110
202220
202330
FIG. 1 — Lineage of K083097, 2007–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062297VHP 136 Low Temperature Sterilization System
Summary page 2
…21 CFR 880.6860 Product Code: 80 MLR. 2. Predicate Device · Amsco V-PRO 1 Low Temperature Sterilization System: K062297. 3. Overview The purpose of this 510(k) is to obtain premarket clearance for the Amsco V-PRO 1 P…
STERIS Corporation20071

Also cleared against K062297: 7 other devices →

What descends from it? (27)

See all 27

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111810Amsco V-Pro 1 and V-Pro 1 Plus Low Temperature Sterilization UnitsSTERIS Corporation201118
K102394Amsco V-Pro 1 Low Temperature Sterilization System; Amsco V-Pro 1 Plus Low Temperature…STERIS Corporation201118
K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation201217
K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation201316
K140464Sterilucent PSD-85 Hydrogen Peroxide SterilizerSterilucent, Inc.20145
K112813Amsco V-ProSteris Corp20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 1, 2011Not classifiedZ-3041-2011OtherSTERIS learned that Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems users may experience premature wear of the vacuum pump and filtering assembly components. Users may also experience incomplete aeration of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K111810 (STERIS Corporation, 2011) has gone furthest, with 18 descendants of its own.
  • 7 other devices also cite its predicate K062297.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The Amsco V-PRO 1 Plus Low Temperature Sterilization System, with VAPROX HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The two pre-programmed sterilization cycles; the Lumen and the Non Lumen Cycles operate at low pressure and low temperature and are thus suitable for processing medical devices…

Read the full statement (285 words)

The Amsco V-PRO 1 Plus Low Temperature Sterilization System, with VAPROX HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The two pre-programmed sterilization cycles; the Lumen and the Non Lumen Cycles operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

The Amsco V-PRO 1 Plus Low Temperature Sterilizer System's (Lumen Cycle) was cleared under (K062297) can sterilize *'

· Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

* Medical, devices with a single stainless steel lumen with:

o an inside diameter of.1 mnm or larger and a length of a 125 mm or shorter

o an insider diameter of 2 mm or larger and a length of 250 mm or shorter

o an inside diameter of 3 mm or larger and a length of 400 mm or shorter

The validation testing for all lumen sizes was conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

The Amsco V-PRO 1 Plus Low Temperature Sterilization System's (Non Lumen Cycle) the subject of this (K083097) can sterilize**:

Non-lumened instruments including non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.

** The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083097: Amsco V-Pro 1 Low Temperature Sterilization System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K083097

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026