510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Modification to Autofuser

K090300510(k) SpecialCleared Apr 30, 200983-day reviewClass II21 CFR 880.5725Recalls 16

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ace Medical Us, LLC · MEB Pump, Infusion, Elastomeric

Cleared on Apr 30, 2009 as substantially equivalent to K060258 (Ace Medical Us, LLC, 2006) after an 83-day review. Its family tree reaches back 8 generations to 1990. It has been recalled 16 times.

Decided
Apr 30, 2009
Review
83 days
Class
Class II
Product code
MEB
Recalls
16
Predicates
1

Its longest line of descent runs back 8 generations to K896422 (1990). 21 ancestors in all, the oldest K884505 (1989). No later device cites it as a predicate yet.

  1. Root · 8 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 7 generations backK980558 Painbuster Infusion SystemI-Flow Corp. · 1998
  3. 6 generations backK982946 Painbuster Infusion KitI-Flow Corp. · 1998
  4. 5 generations backK984502 Nerve Block Infusion KitI-Flow Corp. · 1999
  5. 4 generations backK020862 I-Flow Elastomeric PumpI-Flow Corp. · 2002 · Recalled
  6. 3 generations backK023098 Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser…Mckinley Infuser · 2002
  7. 2 generations backK033039 Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc. · 2003
  8. 1 generation backK060258 AutofuserAce Medical Us, LLC · 2006 · Recalled
  9. This deviceK090300 Modification to AutofuserAce Medical Us, LLC · 2009
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K896422Burron Ambulatory Drug Delivery SystemBurron Medical, Inc.1990—
−7K980558Painbuster Infusion SystemI-Flow Corp.1998—
−6K982946Painbuster Infusion KitI-Flow Corp.1998—
−5K984502Nerve Block Infusion KitI-Flow Corp.1999—
−4K020862I-Flow Elastomeric PumpI-Flow Corp.20021
−3K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser2002—
−2K033039Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc.2003—
−1K060258AutofuserAce Medical Us, LLC200646
0 (this device)K090300Modification to AutofuserAce Medical Us, LLC2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199020
199310
199520
199810
199930
200120
200410
FIG. 1 — Lineage of K090300, 1990–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MEB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEB device

  1. #1 K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)score 14.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060258Autofuser
Summary page 1
…Regulation Number: 880.5725 Classification Panel: 80, General Hospital and Personal Use Device Predicate Device Ace Medical Autofluser Elastomeric Infusion Pump System (K060258) Intended Use / Indications for Use The Autofuser family of ambulatory infusion pumps with integrated administration set, eithe…
Ace Medical Us, LLC200646

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2018Class IIZ-0943-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0949-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0950-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0955-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0956-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0940-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0941-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0942-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0946-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0954-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0945-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0948-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0953-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0947-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0951-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…
Oct 2, 2018Class IIZ-0952-2019Process controlPossible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 6 of the 21 ancestors have a recall of any class.

How does its category compare?

All MEB devices

MEB Pump, Infusion, Elastomeric: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · MEB

2in 2024

2007: 52008: 22009: 12013: 12014: 12016: 12019: 12024: 22026 (partial year): 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEB

42in 2007

2007: 42 days median
2007high 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEB

17in 2018

2007: 12008: 32009: 482010: 22011: 8, 1 Class I2013: 2, 1 Class I2018: 17
2007high 48 (2009) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MEB

1,408in 2025

2007: 1102008: 4442009: 7982010: 1,8272011: 2,2652012: 1,9762013: 1,9562014: 2,6952015: 2,9802016: 2,1472017: 1,6532018: 1,4732019: 9742020: 2,7322021: 1,8022022: 1,2882023: 1,2482024: 1,4242025: 1,408
2007high 2,980 (2015) · low 110 (2007)2025

MAUDE reports naming product code MEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075421110
20082—3444
20091—48798
20100—21,827
20110—8 (1)2,265
20120—01,976
20131—2 (1)1,956
20141—02,695
20150—02,980
20161—02,147
20170—01,653
20180—171,473
20191—0974
20200—02,732
20210—01,802
20220—01,288
20230—01,248
20242—01,424
20250—01,408
20261—0—

What is it indicated for?

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites. Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button. General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090300: Modification to Autofuser.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K090300

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026