510(k) Report

510(k) · Cleared

Recalled · 2009

Autofuser

K060258510(k) TraditionalCleared May 30, 2006118-day reviewClass II21 CFR 880.5725Recalls 46

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ace Medical Us, LLC · MEB Pump, Infusion, Elastomeric

Cleared on May 30, 2006 as substantially equivalent to K041585 (Algos, LC, 2004) and 5 other predicates after a 118-day review. Its family tree reaches back 7 generations to 1990. 1 later device descends from it. It has been recalled 46 times.

Decided
May 30, 2006
Review
118 days
Class
Class II
Product code
MEB
Recalls
46
Predicates
6

Its longest line of descent runs back 7 generations to K896422 (1990). 20 ancestors in all, the oldest K884505 (1989). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor1995: 2 other ancestors1990: 2 other ancestors1993: 1 other ancestor2001: 2 other ancestors1999: 3 other ancestors1998: 1 other ancestor1989: 1 other ancestor2009: 1 descendant clearedK896422 · Burron Ambulatory Drug Delivery System · Burron Medical, Inc. · 1990K980558 · Painbuster Infusion System · I-Flow Corp. · 1998×2: 2 generations cleared close together, K982946 (1998), K984502 (1999) K020862 · I-Flow Elastomeric Pump · I-Flow Corp. · 2002 · recalledK023098 · Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion Kit · Mckinley Infuser · 2002K033039 · Modification to: Accufuser, Accufuser Plus, Standard Procedure Kits · Mckinley, Inc. · 20031990200020102020
  1. Root · 7 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 6 generations backK980558 Painbuster Infusion SystemI-Flow Corp. · 1998
  3. 5 generations backK982946 Painbuster Infusion KitI-Flow Corp. · 1998
  4. 4 generations backK984502 Nerve Block Infusion KitI-Flow Corp. · 1999
  5. 3 generations backK020862 I-Flow Elastomeric PumpI-Flow Corp. · 2002 · Recalled
  6. 2 generations backK023098 Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser…Mckinley Infuser · 2002
  7. 1 generation backK033039 Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc. · 2003
  8. This deviceK060258 AutofuserAce Medical Us, LLC · 2006
  9. Descendant · cites it directly · 0 of its ownK090300 Modification to AutofuserAce Medical Us, LLC · 2009 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K090300Modification to AutofuserAce Medical Us, LLC · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K896422Burron Ambulatory Drug Delivery SystemBurron Medical, Inc.1990—
−6K980558Painbuster Infusion SystemI-Flow Corp.1998—
−5K982946Painbuster Infusion KitI-Flow Corp.1998—
−4K984502Nerve Block Infusion KitI-Flow Corp.1999—
−3K020862I-Flow Elastomeric PumpI-Flow Corp.20021
−2K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser2002—
−1K033039Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc.2003—
0 (this device)K060258AutofuserAce Medical Us, LLC2006—
+1K090300Modification to AutofuserAce Medical Us, LLC200916

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199020
199310
199520
199810
199930
200120
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K060258, 1990–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MEB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEB device

  1. #1 K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)score 14.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041585Autofuser Ambulatory Infusion Pump
Summary page 1
…Regulation Number: 880.5725 Classification Panel: 80, General Hospital and Personal Use Device Product code: MEB - Elastomeric Infusion Pump Original cleared 510(k): K041585 Establishment Registration: pending assignment from FDA Owner/Operator Number: 9082449 5. Summary of Safety and Effectivene…
Algos, LC200446
*K033039Modification to: Accufuser, Accufuser Plus, Standard Procedure Kits
Summary page 2
…ser (K04 1585). The modified Autofuiser system also demonstrates substantial equivalence to the Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepunip/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546). 5 10(k) Ace M…
Mckinley, Inc.2003—
*K980558Painbuster Infusion System
Summary page 2
…stantial equivalence to the Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepunip/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546). 5 10(k) Ace Medical Autofuser System 5 10(k) Sunumary p. 2 of…
I-Flow Corp.1998—
*K982946Painbuster Infusion Kit
Summary page 2
…equivalence to the Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepunip/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546). 5 10(k) Ace Medical Autofuser System 5 10(k) Sunumary p. 2 of 2…
I-Flow Corp.1998—
*K944692C-Series Homepump
Summary page 2
…ce to the Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepunip/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546). 5 10(k) Ace Medical Autofuser System 5 10(k) Sunumary p. 2 of 2…
Block Medical, Inc.1995—
*K896546Homepump (Disp. Elastomeric Infusion Device)
Summary page 2
…Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepunip/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546). 5 10(k) Ace Medical Autofuser System 5 10(k) Sunumary p. 2 of 2…
Block Medical, Inc.1990—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090300Modification to AutofuserAce Medical Us, LLC20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 14, 2009Not classifiedZ-1846-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1848-2009Process controlTFailure to infuse and potential compromised sterility: he seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1834-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1840-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1845-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1829-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1862-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1852-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1832-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1868-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1836-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1838-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1859-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1855-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1865-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1847-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1839-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1858-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1863-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1867-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1844-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1856-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1827-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1830-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1837-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1831-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1842-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1826-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1841-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1824-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1861-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1854-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1828-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1851-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1857-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1864-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1835-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1866-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1823-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1849-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1843-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1825-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1860-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1850-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1853-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
May 14, 2009Not classifiedZ-1833-2009Process controlFailure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 10 months on median.
  • 5 of the 20 ancestors have a recall of any class.
  • Of its descendants, K090300 (Ace Medical Us, LLC, 2009) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MEB devices

MEB Pump, Infusion, Elastomeric: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MEB

2in 2024

2007: 52008: 22009: 12013: 12014: 12016: 12019: 12024: 22026 (partial year): 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEB

42in 2007

2007: 42 days median
2007high 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEB

17in 2018

2007: 12008: 32009: 482010: 22011: 8, 1 Class I2013: 2, 1 Class I2018: 17
2007high 48 (2009) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MEB

1,408in 2025

2007: 1102008: 4442009: 7982010: 1,8272011: 2,2652012: 1,9762013: 1,9562014: 2,6952015: 2,9802016: 2,1472017: 1,6532018: 1,4732019: 9742020: 2,7322021: 1,8022022: 1,2882023: 1,2482024: 1,4242025: 1,408
2007high 2,980 (2015) · low 110 (2007)2025

MAUDE reports naming product code MEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075421110
20082—3444
20091—48798
20100—21,827
20110—8 (1)2,265
20120—01,976
20131—2 (1)1,956
20141—02,695
20150—02,980
20161—02,147
20170—01,653
20180—171,473
20191—0974
20200—02,732
20210—01,802
20220—01,288
20230—01,248
20242—01,424
20250—01,408
20261—0—

What is it indicated for?

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidurar, and into intra-operative (soft tissue/body cavity) sites.

Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button.

General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060258: Autofuser.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K060258

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026