510(k) Report

510(k) · Cleared

Cleared

Ermis Sterilization Container System

K123234510(k) TraditionalCleared Jul 5, 2013263-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ermis Medizintechnik · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Jul 5, 2013 as substantially equivalent to K091720 (Sklar Corp., 2010) after a 263-day review. Its family tree reaches back 5 generations to 2000. 1 later device descends from it.

Decided
Jul 5, 2013
Review
263 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K991023 (2000). 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK991023 Steritite Perforated Bottom Rigid Reusable Sterilization Container…Case Medical, Inc. · 2000
  2. 4 generations backK012931 Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp. · 2002
  3. 3 generations backK023614 Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc. · 2003 · Recalled
  4. 2 generations backK050570 Miltex Rigid Sterilization Container SystemMiltex, Inc. · 2005
  5. 1 generation backK091720 Sklarlite Rigid Sterilization Container SystemSklar Corp. · 2010
  6. This deviceK123234 Ermis Sterilization Container SystemErmis Medizintechnik · 2013
  7. Descendant · cites it directly · 0 of its ownK180528 InstruSafe (R) Sterilization ContainerSummit Medical, Inc. · 2018

Most influential descendants

Showing 1 of 1 · See all
  • K180528InstruSafe (R) Sterilization ContainerSummit Medical, Inc. · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K991023Steritite Perforated Bottom Rigid Reusable Sterilization Container System with…Case Medical, Inc.2000—
−4K012931Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp.2002—
−3K023614Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc.20031
−2K050570Miltex Rigid Sterilization Container SystemMiltex, Inc.2005—
−1K091720Sklarlite Rigid Sterilization Container SystemSklar Corp.2010—
0 (this device)K123234Ermis Sterilization Container SystemErmis Medizintechnik2013—
+1K180528InstruSafe (R) Sterilization ContainerSummit Medical, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K123234, 2000–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK121909FDA’s K121909 is Inspire 6 Dual Hollow Fiber Oxygenator with Integrated Hardshell…
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Sorin Group Italia S.R.L.2012—
*K091720Sklarlite Rigid Sterilization Container System
Summary page 2
…r Sahadir Sterilization Trays K 12190 Bahadir USA Corn. SkiarLite Rigid Slerilizatioon Container K091 720 Sklar Instruments Systems 001 510K Sunriajvl 5,doc…
Sklar Corp.2010—

Also cleared against K091720: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180528InstruSafe (R) Sterilization ContainerSummit Medical, Inc.20180

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K180528 (Summit Medical, Inc., 2018) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K091720.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The ERMIS Sterilization Container System Is Intended to allow sterilization of the enclosed medical device and also maintain sterility during transport and storage for 30 days. The system consists of different models, such as 1/1 Size, % Size, %1size, mini and dental containers which may vary in size, perforations and color. All models are available with perforated lids and either perforated or non-perforated bottoms. To enable proper organization of…

Read the full statement (145 words)

The ERMIS Sterilization Container System Is Intended to allow sterilization of the enclosed medical device and also maintain sterility during transport and storage for 30 days. The system consists of different models, such as 1/1 Size, % Size, %1size, mini and dental containers which may vary in size, perforations and color. All models are available with perforated lids and either perforated or non-perforated bottoms. To enable proper organization of sterilized goods, different wire- or sleeve baskets exactly adapted to the specific container dimensions are available. All models are intended to be used with Ermis single use paper filters.

The containers are reusable devices designed to be used with the following sterilization cycle parameters:

Pre Vacuum cycle:

4 minutes 3 minutes

132 0 C (270-F) 135-C (273-F)

Drying Time minimum 20 minutes Drying Time minimum 16 minutes

Loading: Metal surgical instruments (scissors, clamps, forceps) and textiles

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K123234: Ermis Sterilization Container System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K123234

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026