510(k) Report

510(k) · Cleared

Recalled · Class II · 2011

GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQ

K092004510(k) TraditionalCleared Aug 25, 200954-day reviewClass II21 CFR 892.1650Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · OWB Interventional Fluoroscopic X-Ray System

Cleared on Aug 25, 2009 as substantially equivalent to K052412 (GE Healthcare, 2005) and 5 other predicates after a 54-day review. Its family tree reaches back 11 generations to 1993. 1 of the 83 devices in its family, counting itself, has a Class I recall. It has been recalled 2 times.

Decided
Aug 25, 2009
Review
54 days
Class
Class II
Product code
OWB
Recalls
2
Predicates
6

Its longest line of descent runs back 11 generations to K926514 (1993). 82 ancestors in all, the oldest K882061 (1988). No later device cites it as a predicate yet. 1 of the 83 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2006: 5 other ancestors2005: 3 other ancestors2003: 10 other ancestors, 1 with a Class I recall12008: 2 other ancestors2007: 2 other ancestors2002: 7 other ancestors2000: 4 other ancestors1999: 3 other ancestors1998: 4 other ancestors2004: 4 other ancestors1995: 10 other ancestors1993: 2 other ancestors1989: 2 other ancestors2001: 4 other ancestors1997: 2 other ancestors1991: 2 other ancestors1994: 1 other ancestor1992: 2 other ancestors1996: 1 other ancestor1988: 1 other ancestorK926514 · Endoscopic Ultrasound System Type 20 · Olympus Corp. · 1993 · recalledK951994 · Olympus EU-M30 Endoscopic Ultrasound Center · Olympus America, Inc. · 1996K982323 · Olympus UM 2R/3R Ultrasonic Probes and Associated Ancillary Equipment, for Bronchial Use · Olympus America, Inc. · 1998K042438 · Superdimension/Bronchus · Superdimension, Ltd. · 2004 · recalled×2: 2 generations cleared close together, K052260 (2005), K052852 (2005) K062315 · Superdimension/Bronchus Premium 2 · Superdimension, Ltd. · 2006 · recalledK071473 · Modification to Superdimension/Bronchus Premium 2 · Superdimension, Ltd. · 2007×2: 2 generations cleared close together, K080271 (2008), K081379 (2008) K083745 · The IC-Pro System, Model Version 3.2 · Paieon, Inc. · 2009198519901995200020052010
  1. Root · 11 generations backK926514 Endoscopic Ultrasound System Type 20Olympus Corp. · 1993 · Recalled
  2. 7 generations between
  3. 3 generations backK080271 Inreach SystemSuperdimension, Ltd. · 2008
  4. 2 generations backK081379 Modification to Inreach SystemSuperdimension, Ltd. · 2008
  5. 1 generation backK083745 The IC-Pro System, Model Version 3.2Paieon, Inc. · 2009
  6. This deviceK092004 GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model…GE Healthcare · 2009
Show all 11 generations
  1. Root · 11 generations backK926514 Endoscopic Ultrasound System Type 20Olympus Corp. · 1993 · Recalled
  2. 10 generations backK951994 Olympus EU-M30 Endoscopic Ultrasound CenterOlympus America, Inc. · 1996
  3. 9 generations backK982323 Olympus UM 2R/3R Ultrasonic Probes and Associated Ancillary…Olympus America, Inc. · 1998
  4. 8 generations backK042438 Superdimension/BronchusSuperdimension, Ltd. · 2004 · Recalled
  5. 7 generations backK052260 Superdimension/Bronchus 4.1Superdimension, Ltd. · 2005 · Recalled
  6. 6 generations backK052852 Superdimension/Bronchus PremiumSuperdimension, Ltd. · 2005 · Recalled
  7. 5 generations backK062315 Superdimension/Bronchus Premium 2Superdimension, Ltd. · 2006 · Recalled
  8. 4 generations backK071473 Modification to Superdimension/Bronchus Premium 2Superdimension, Ltd. · 2007
  9. 3 generations backK080271 Inreach SystemSuperdimension, Ltd. · 2008
  10. 2 generations backK081379 Modification to Inreach SystemSuperdimension, Ltd. · 2008
  11. 1 generation backK083745 The IC-Pro System, Model Version 3.2Paieon, Inc. · 2009
  12. This deviceK092004 GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model…GE Healthcare · 2009

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K926514Endoscopic Ultrasound System Type 20Olympus Corp.19931
−10K951994Olympus EU-M30 Endoscopic Ultrasound CenterOlympus America, Inc.1996—
−9K982323Olympus UM 2R/3R Ultrasonic Probes and Associated Ancillary Equipment, for…Olympus America, Inc.1998—
−8K042438Superdimension/BronchusSuperdimension, Ltd.20041
−7K052260Superdimension/Bronchus 4.1Superdimension, Ltd.20052
−6K052852Superdimension/Bronchus PremiumSuperdimension, Ltd.20051
−5K062315Superdimension/Bronchus Premium 2Superdimension, Ltd.20061
−4K071473Modification to Superdimension/Bronchus Premium 2Superdimension, Ltd.2007—
−3K080271Inreach SystemSuperdimension, Ltd.2008—
−2K081379Modification to Inreach SystemSuperdimension, Ltd.2008—
−1K083745The IC-Pro System, Model Version 3.2Paieon, Inc.2009—
0 (this device)K092004GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ…GE Healthcare2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198920
199120
199220
199320
199410
1995100
199610
199720
199840
199930
200040
200140
200270
2003101
200440
200530
200650
200720
200820
FIG. 1 — Lineage of K092004, 1993–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ
Summary page 1
…33130704399 Innova 4100, 3100, 2100' with StentViz Innova 4100'°, 3100', 2100'? with StentViz System X-F!ay, Angiographic : OWR, JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K03183…
GE Healthcare200550
*K050489Innova 2100
Summary page 1
…Innova 4100, 3100, 2100' with StentViz Innova 4100'°, 3100', 2100'? with StentViz System X-F!ay, Angiographic : OWR, JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K031836) The In…
Ge Medical Systems, LLC200527
*K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121
Summary page 1
…4100, 3100, 2100' with StentViz Innova 4100'°, 3100', 2100'? with StentViz System X-F!ay, Angiographic : OWR, JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K031836) The Innova 4100…
Ge Medical Systems, LLC20064
*K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…
Summary page 1
…00, 2100' with StentViz Innova 4100'°, 3100', 2100'? with StentViz System X-F!ay, Angiographic : OWR, JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K031836) The Innova 4100/, 3100',…
Ge Medical Systems, LLC200638
*K083745The IC-Pro System, Model Version 3.2
Summary page 1
…Innova 4100'°, 3100', 2100'? with StentViz System X-F!ay, Angiographic : OWR, JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K031836) The Innova 4100/, 3100', 2100" Systems are modif…
Paieon, Inc.2009—
*K031836Philips Fresco Release 1
Summary page 1
…JAk, 121 innova 41oo'°. 3100, 21oo'° K052412, KO050489, K060259, K061163 IC-PRO device (K083745) featuring StentOp - Philips Fresco release 1 (featuring StentBoost) (K031836) The Innova 4100/, 3100', 2100" Systems are modified with an optional software feature called StentViz. The StentViz feature enh…
Philips Medical Systems North America Co.2003—

Also cleared against K052412: 7 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 24, 2011Not classifiedZ-0133-2012Nonconforming Material/ComponentA potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
Feb 4, 2011Class IIZ-0934-2013Nonconforming Material/ComponentExposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 SeriesThe Olympus Optical Co. · 2003 · 9 generations back · ◆ in the 2003 ancestors bar of FIG. 114 Class I recalls

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 10 months on median.
  • 30 of the 82 ancestors have a recall of any class; 1 has a Class I recall.
  • 7 other devices also cite its predicate K052412.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K092004: GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQ.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K092004

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026