Trilogy 202 Ventilator

K093905510(k) AbbreviatedCleared May 12, 2010141-day reviewClass II21 CFR 868.5895Recalls 3 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on May 12, 2010 after a 141-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 7 of the 13 devices in its family, counting itself, have a Class I recall. 12 later devices descend from it. It has been recalled 3 times, including 2 Class I recalls.

Decided
May 12, 2010
Review
141 days
Class
Class II
Product code
CBK
Recalls
3 (2 Class I)
Predicates
0

12 later devices descend from it, 1 citing it directly. 7 of the 13 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2011: 1 descendant cleared, 1 with a Class I recall12019: 2 descendants cleared, 2 with a Class I recall22013: 2 descendants cleared, 1 with a Class I recall12017: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2025: 2 descendants cleared2021: 1 descendant cleared, 1 with a Class I recall12024: 2 descendants cleared2010201520202025
  1. This deviceK093905 Trilogy 202 VentilatorRespironics, Inc. · 2010
  2. Descendant · cites it directly · 11 of its ownK111610 Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Recalled (Class I)
  3. Descendant · 2 generations on · 8 of its ownK123144 VIVO 50Ge Healthcare, Breas Medical AB · 2013 · Recalled
  4. Descendant · 3 generations on · 7 of its ownK160481 Vivo 60Breas Medical AB · 2016 · Recalled
  5. Descendant · 2 generations on · 6 of its ownK162877 VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · Recalled (Class I)
  6. Descendant · 2 generations on · 5 of its ownK181166 Trilogy EvoRespironics, Inc. · 2019 · Recalled (Class I)

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2011: 12013: 22016: 12017: 12019: 22021: 12024: 22025: 2
2011high 2 (2013) · low 0 (2012)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · cites it directly11descendants
  • K123144VIVO 50Ge Healthcare, Breas Medical AB · 2013 · 2 generations on8descendants
  • K160481Vivo 60Breas Medical AB · 2016 · 3 generations on7descendants
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 2 generations on6descendants
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 2 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K093905Trilogy 202 VentilatorRespironics, Inc.2010—
+1K111610Trilogy Series Ventilator with OximetryRespironics, Inc.2011Class I
+2K123144VIVO 50Ge Healthcare, Breas Medical AB20131
+3K160481Vivo 60Breas Medical AB20162
+2K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc.2017Class I
+2K181166Trilogy EvoRespironics, Inc.2019Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201111
201321
201610
201711
201922
202111
202420
202520
FIG. 1 — Lineage of K093905, 2010–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (12)

See all 12

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111610Trilogy Series Ventilator with Oximetry Class IRespironics, Inc.201111

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 18, 2016Class IIZ-1065-2016Software designSoftware Issue
Feb 11, 2014Class IZ-1143-2014Component design/selectionInternal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
Apr 27, 2012Class IZ-2024-2012Component design/selectionThe specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K181170Trilogy Evo UniversalRespironics, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 112 Class I recalls
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls
  • K122715Stellar 150Resmed Germany, Inc. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K193586Vivo 45 LSBreas Medical AB · 2021 · 3 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K111610 (Respironics, Inc., 2011) has gone furthest, with 11 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Respironics Trilogy 202 system provides continuous or intermittent ventilatory

support for the care of individuals who require mechanical ventilation with or without

air/oxygen blending. Trilogy 202 is intended for pediatric through adult patients

weighing at least 5 kg (11 lbs.).

The device is intended to be used in hospitals and institutions, and for portable

applications such as wheelchairs and gurneys only when in an institutional setting. It may

be used for both invasive and non-invasive ventilation. It is not intended to be used as a

transport ventilator.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K093905: Trilogy 202 Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093905

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026