510(k) Report

510(k) · Cleared

Recalled · Class I · 2025

Zyno Z-800 Infusion System Model Z-800

K100705510(k) AbbreviatedCleared Jul 8, 2010118-day reviewClass II21 CFR 880.5725Recalls 8 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zyno Medical, LLC · FRN Pump, Infusion

Cleared on Jul 8, 2010 as substantially equivalent to K071545 (Zyno Medical, LLC, 2007) and 3 other predicates after a 118-day review. Its family tree reaches back 1 generation to 1995. 3 of the 6 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 8 times, including 5 Class I recalls.

Decided
Jul 8, 2010
Review
118 days
Class
Class II
Product code
FRN
Recalls
8 (5 Class I)
Predicates
4

Its longest line of descent runs back 1 generation to K950766 (1995). 4 ancestors in all. 1 later device descends from it, 1 citing it directly. 3 of the 6 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall1K950766 · Sigma 8000 and Sigma 8002 · Sigma Intl. General Medical Apparatus, Inc. · 19951995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant, 1 with a Class I recall1 K950766 · Sigma 8000 and Sigma 8002 · Sigma Intl. General Medical Apparatus, Inc. · 19950
  1. Root · 1 generation backK950766 Sigma 8000 and Sigma 8002Sigma Intl. General Medical Apparatus, Inc. · 1995
  2. This deviceK100705 Zyno Z-800 Infusion System Model Z-800Zyno Medical, LLC · 2010
  3. Descendant · cites it directly · 0 of its ownK130690 Z-800 Infusion SystemZyno Medical, LLC · 2013 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K130690Z-800 Infusion SystemZyno Medical, LLC · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K950766Sigma 8000 and Sigma 8002Sigma Intl. General Medical Apparatus, Inc.1995—
0 (this device)K100705Zyno Z-800 Infusion System Model Z-800Zyno Medical, LLC2010—
+1K130690Z-800 Infusion SystemZyno Medical, LLC2013Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199911
200010
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201311
FIG. 1 — Lineage of K100705, 1995–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071545Z-800 Volumetric Infusion Pump
Summary page 1
…Common Name: Volumetric Infusion Pump Classification Name: Infusion Pump Product code: FRN 880.5725 Predicate Devices -Z-800 Infusion pumps (K07 1545) -B. Braun Vista basic (formerly known as Infusomnat P, K003029) -Sigma 8000 (K950766) -Abbott Acclaim Encor…
Zyno Medical, LLC20071
*K003029Infusomat P
Summary page 1
…on Pump Product code: FRN 880.5725 Predicate Devices -Z-800 Infusion pumps (K07 1545) -B. Braun Vista basic (formerly known as Infusomnat P, K003029) -Sigma 8000 (K950766) -Abbott Acclaim Encore (K99150 1) Intended Use The Z-800 Infusion system is i…
B.Braun Medical, Inc.2000—
*K950766Sigma 8000 and Sigma 8002
Summary page 1
…FRN 880.5725 Predicate Devices -Z-800 Infusion pumps (K07 1545) -B. Braun Vista basic (formerly known as Infusomnat P, K003029) -Sigma 8000 (K950766) -Abbott Acclaim Encore (K99150 1) Intended Use The Z-800 Infusion system is intended to provide intravenou…
Sigma Intl. General Medical Apparatus, Inc.1995—
*K991501Acclaim Infusion Pump
Summary page 1
…-Z-800 Infusion pumps (K07 1545) -B. Braun Vista basic (formerly known as Infusomnat P, K003029) -Sigma 8000 (K950766) -Abbott Acclaim Encore (K99150 1) Intended Use The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood an…
Abbott Laboratories1999Class I

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130690Z-800 Infusion System Class IZyno Medical, LLC20130

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2025Class IZ-1867-2025Software design (manufacturing process)Unreleased software versions were installed on distributed devices without verification or validation.
May 7, 2025Class IZ-1868-2025Software design (manufacturing process)Unreleased software versions were installed on distributed devices without verification or validation.
May 7, 2025Class IZ-1865-2025Software design (manufacturing process)Unreleased software versions were installed on distributed devices without verification or validation.
May 7, 2025Class IZ-1866-2025Software design (manufacturing process)Unreleased software versions were installed on distributed devices without verification or validation.
Oct 18, 2024Class IIZ-0506-2025Device DesignPotential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
Sep 30, 2024Class IIIZ-0523-2025Software design (manufacturing process)For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
Sep 13, 2024Class IZ-0005-2025Process change controlThere is a defect in the air-in-line software algorithm.
Jun 13, 2011Class IIZ-1024-2013Process designUpdate Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K991501Acclaim Infusion PumpAbbott Laboratories · 1999 · Direct predicate · ◆ in the 1999 ancestors bar of FIG. 11 Class I recall
  • K130690Z-800 Infusion SystemZyno Medical, LLC · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 15 Class I recalls

What does this lineage say?

  • 2 of the 4 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K130690 (Zyno Medical, LLC, 2013) has gone furthest, with 0 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood and blood products to a human patient under the direction or supervision of physician or other certified health care professional.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100705: Zyno Z-800 Infusion System Model Z-800.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100705

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026