510(k) Report

510(k) · Cleared

Recalled · Class I · 2013

Spacelabs Bleasesirius Anesthesia Workstation

K101850510(k) TraditionalCleared Mar 2, 2011244-day reviewClass II21 CFR 868.5895Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Healthcare, Ltd. · CBK Ventilator, Continuous, Facility Use

Cleared on Mar 2, 2011 as substantially equivalent to K051629 (Blease USA, Inc., 2005) after a 244-day review. Its family tree reaches back 3 generations to 1999. 3 of the 7 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Mar 2, 2011
Review
244 days
Class
Class II
Product code
CBK
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K982132 (1999). 4 ancestors in all. 2 later devices descend from it, 2 citing it directly. 3 of the 7 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor, 1 with a Class I recall12012: 2 descendants cleared, 1 with a Class I recall1K982132 · Blease 6200 Anaesthesia Ventilator · Blease Medical Equipment, Ltd. · 1999 · recalledK041430 · Blease Sirius Anesthesia System, Models 3000 and 2000 · Blease Medical Equipment, Ltd. · 2004K051629 · Bleas Esiriuc Anaesthsia System, Models 3000, 2000 · Blease USA, Inc. · 2005 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor, 1 with a Class I recall1 1 generation forward: 2 descendants, 1 with a Class I recall1 K051629 · Bleas Esiriuc Anaesthsia System, Models 3000, 2000 · Blease USA, Inc. · 2005 · recalledK041430 · Blease Sirius Anesthesia System, Models 3000 and 2000 · Blease Medical Equipment, Ltd. · 2004K982132 · Blease 6200 Anaesthesia Ventilator · Blease Medical Equipment, Ltd. · 1999 · recalled0
  1. Root · 3 generations backK982132 Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd. · 1999 · Recalled
  2. 2 generations backK041430 Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd. · 2004
  3. 1 generation backK051629 Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc. · 2005 · Recalled
  4. This deviceK101850 Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011
  5. Descendant · cites it directly · 0 of its ownK112729 BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · Recalled
  6. Descendant · cites it directly · 0 of its ownK113051 Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · cites it directly0descendants
  • K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K982132Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd.19991
−2K041430Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd.2004—
−1K051629Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc.20053
0 (this device)K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd.2011—
+1K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare20121
+1K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare2012Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200111

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201221
FIG. 1 — Lineage of K101850, 1999–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051629Bleas Esiriuc Anaesthsia System, Models 3000, 2000
Summary page 1
…dicate Predicate Equivalent Devices: 510(k) Number Manufacturer/ Model Spacelabs BleaseSirius K05 1629 Spacelabs Blease Anesthesia Workstation Frontline Sirius 2000,…
Blease USA, Inc.20053

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare20120
K113051Spacelabs Arkon Anesthesia Workstation Class ISpacelabs Healthcare20120

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 15, 2013Class IZ-0968-2013Process controlA defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003251Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690Blease Medical Equipment, Ltd. · 2001 · 3 generations back · ◆ in the 2001 ancestors bar of FIG. 12 Class I recalls
  • K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 5.4 years on median.
  • 3 of the 4 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K112729 (Spacelabs Healthcare, 2012) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Spacelabs BleaseSirius Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101850: Spacelabs Bleasesirius Anesthesia Workstation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101850

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026