Spacelabs Arkon Anesthesia Workstation

K113051510(k) TraditionalCleared Mar 22, 2012161-day reviewClass II21 CFR 868.5895Recalls 6 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Healthcare · CBK Ventilator, Continuous, Facility Use

Cleared on Mar 22, 2012 as substantially equivalent to K101850 (Spacelabs Healthcare, Ltd., 2011) after a 161-day review. Its family tree reaches back 4 generations to 1999. 3 of the 6 devices in its family, counting itself, have a Class I recall. It has been recalled 6 times, including 3 Class I recalls.

Decided
Mar 22, 2012
Review
161 days
Class
Class II
Product code
CBK
Recalls
6 (3 Class I)
Predicates
1

Its longest line of descent runs back 4 generations to K982132 (1999). 5 ancestors in all. No later device cites it as a predicate yet. 3 of the 6 devices in this family, counting this one, have a Class I recall.

  1. Root · 4 generations backK982132 Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd. · 1999 · Recalled
  2. 3 generations backK041430 Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd. · 2004
  3. 2 generations backK051629 Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc. · 2005 · Recalled
  4. 1 generation backK101850 Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · Recalled (Class I)
  5. This deviceK113051 Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K982132Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd.19991
−3K041430Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd.2004—
−2K051629Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc.20053
−1K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd.2011Class I
0 (this device)K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200111
FIG. 1 — Lineage of K113051, 1999–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101850Spacelabs Bleasesirius Anesthesia Workstation
Summary page 2
…edicate Predicate Equivalent Devices: 510(k) Number ManufacturerI Model Spacelabs Arkon K10 1850 Spacelabs BleaseSirius Anesthesia Workstation IAnesthesia Workstation Devic…
Spacelabs Healthcare, Ltd.2011Class I

Also cleared against K101850: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 11, 2018Class IZ-2589-2018Nonconforming Material/ComponentArkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.
Oct 11, 2017Class IZ-0072-2018Software designArkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or…
Apr 14, 2017Class IIZ-2093-2017Device DesignReports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
Feb 8, 2017Class IIZ-1460-2017Software designSpacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.
Mar 10, 2014Class IZ-1442-2014Software in the Use EnvironmentThe System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence.
Jun 7, 2013Class IIZ-1771-2013Software designSpacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K003251Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690Blease Medical Equipment, Ltd. · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 5.4 years on median.
  • 4 of the 5 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K101850.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.

The device is intended for use only by a suitably qualified physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113051: Spacelabs Arkon Anesthesia Workstation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113051

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.