510(k) Report

510(k) · Cleared

Recalled · 2010

Bleas Esiriuc Anaesthsia System, Models 3000, 2000

K051629510(k) SpecialCleared Sep 23, 200595-day reviewClass II21 CFR 868.5160Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Blease USA, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Sep 23, 2005 as substantially equivalent to K041430 (Blease Medical Equipment, Ltd., 2004) after a 95-day review. Its family tree reaches back 2 generations to 1999. 3 of the 7 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 3 times.

Decided
Sep 23, 2005
Review
95 days
Class
Class II
Product code
BSZ
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K982132 (1999). 3 ancestors in all. 3 later devices descend from it, 1 citing it directly. 3 of the 7 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor, 1 with a Class I recall12011: 1 descendant cleared, 1 with a Class I recall12012: 2 descendants cleared, 1 with a Class I recall1K982132 · Blease 6200 Anaesthesia Ventilator · Blease Medical Equipment, Ltd. · 1999 · recalledK041430 · Blease Sirius Anesthesia System, Models 3000 and 2000 · Blease Medical Equipment, Ltd. · 20041995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor, 1 with a Class I recall1 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall1 K041430 · Blease Sirius Anesthesia System, Models 3000 and 2000 · Blease Medical Equipment, Ltd. · 2004K982132 · Blease 6200 Anaesthesia Ventilator · Blease Medical Equipment, Ltd. · 1999 · recalled0
  1. Root · 2 generations backK982132 Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd. · 1999 · Recalled
  2. 1 generation backK041430 Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd. · 2004
  3. This deviceK051629 Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc. · 2005
  4. Descendant · cites it directly · 2 of its ownK101850 Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · Recalled (Class I)
  5. Descendant · 2 generations on · 0 of its ownK112729 BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · Recalled
  6. Descendant · 2 generations on · 0 of its ownK113051 Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · cites it directly2descendants
  • K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · 2 generations on0descendants
  • K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K982132Blease 6200 Anaesthesia VentilatorBlease Medical Equipment, Ltd.19991
−1K041430Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd.2004—
0 (this device)K051629Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc.2005—
+1K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd.2011Class I
+2K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare20121
+2K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare2012Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200111

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201111
201221
FIG. 1 — Lineage of K051629, 1999–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041430Blease Sirius Anesthesia System, Models 3000 and 2000
Summary page 1
…ntline Sirius 2000, 3000 Classification Code: BSZ Classification Name: Anaesthesia Gas Machine Blease Frontline Sirius K041430 Predicate Device: Device Description/ Comparison: The Frontline Sirius with 8700 ventilator is…
Blease Medical Equipment, Ltd.2004—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K101850Spacelabs Bleasesirius Anesthesia Workstation Class ISpacelabs Healthcare, Ltd.20112

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 24, 2010Not classifiedZ-2113-2010OtherPotential for the touch screen on the ventilator portion of the device to go blank when touched. Possibility that this failure may also cause mechanical ventilation to stop.
Mar 26, 2010Not classifiedZ-1408-2010OtherPotential for delivery of a lower than set concentration of anaesthetic to a patient due to a sticking valve. The lower shuttle may become jammed, not permitting flow of anaesthesia gas through the vapouriser. The vapouriser back bar…
Jun 25, 2009Not classifiedZ-0813-2011OtherUsers may experience difficulty turning the control knob to an active vapour delivery setting, or may not be able to turn on the Left Hand mounted vapouriser on the Blease Sirius Anaesthesia Systems.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003251Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690Blease Medical Equipment, Ltd. · 2001 · 2 generations back · ◆ in the 2001 ancestors bar of FIG. 12 Class I recalls
  • K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 5.5 years on median.
  • 2 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K101850 (Spacelabs Healthcare, Ltd., 2011) has gone furthest, with 2 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Blease Frontline Sirius Range, Anaesthesia Machines are intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air, and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of a mechanical ventilator, and may be used for the delivery of anaesthetic vapour by use of a dismountable vaporizer.

The device is intended for use only by a suitably qualified physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051629: Bleas Esiriuc Anaesthsia System, Models 3000, 2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051629

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026