Recalled · Class I · 2013Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Blease Medical Equipment, Ltd. · BSZ Gas-Machine, Anesthesia
Cleared on May 23, 2001 after a 218-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 of the 6 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled 3 times, including 2 Class I recalls.
- Decided
- May 23, 2001
- Review
- 218 days
- Class
- Class II
- Product code
- BSZ
- Recalls
- 3 (2 Class I)
- Predicates
- 0
5 later devices descend from it, 1 citing it directly. 3 of the 6 devices in this family, counting this one, have a Class I recall.
- This deviceK003251 Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and…Blease Medical Equipment, Ltd. · 2001
- Descendant · cites it directly · 4 of its ownK041430 Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd. · 2004
- Descendant · 2 generations on · 3 of its ownK051629 Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc. · 2005 · Recalled
- Descendant · 3 generations on · 2 of its ownK101850 Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · Recalled (Class I)
- Descendant · 4 generations on · 0 of its ownK112729 BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · Recalled
- Descendant · 4 generations on · 0 of its ownK113051 Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · Recalled (Class I)
Class I recalls in this family
Most influential descendants
Showing 5 of 5 · See all- K041430Blease Sirius Anesthesia System, Models 3000 and 2000Blease Medical Equipment, Ltd. · 2004 · cites it directly4descendants
- K051629Bleas Esiriuc Anaesthsia System, Models 3000, 2000Blease USA, Inc. · 2005 · 2 generations on3descendants
- K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · 3 generations on2descendants
- K112729BLEASE700/900 Series VentilatorsSpacelabs Healthcare · 2012 · 4 generations on0descendants
- K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · 4 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K003251 | Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690 | Blease Medical Equipment, Ltd. | 2001 | — |
| +1 | K041430 | Blease Sirius Anesthesia System, Models 3000 and 2000 | Blease Medical Equipment, Ltd. | 2004 | — |
| +2 | K051629 | Bleas Esiriuc Anaesthsia System, Models 3000, 2000 | Blease USA, Inc. | 2005 | 3 |
| +3 | K101850 | Spacelabs Bleasesirius Anesthesia Workstation | Spacelabs Healthcare, Ltd. | 2011 | Class I |
| +4 | K112729 | BLEASE700/900 Series Ventilators | Spacelabs Healthcare | 2012 | 1 |
| +4 | K113051 | Spacelabs Arkon Anesthesia Workstation | Spacelabs Healthcare | 2012 | Class I |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2011 | 1 | 1 |
| 2012 | 2 | 1 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BSZ device
- #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (5)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K041430 | Blease Sirius Anesthesia System, Models 3000 and 2000 | 2004 | 4 |
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 17, 2013 | Class I | Z-0352-2014 | Process control | Spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber. |
| Jan 15, 2013 | Class I | Z-0968-2013 | Process control | A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient. |
| Nov 2, 2012 | Class II | Z-0565-2013 | Labeling design | The BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators were shipped with manuals that describe a secondary O2 Supply Failure alarm which has not been incorporated into the system. The manuals were prematurely updated for a… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K101850Spacelabs Bleasesirius Anesthesia WorkstationSpacelabs Healthcare, Ltd. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
- K113051Spacelabs Arkon Anesthesia WorkstationSpacelabs Healthcare · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls
What does this lineage say?
- Of its descendants, K041430 (Blease Medical Equipment, Ltd., 2004) has gone furthest, with 4 descendants of its own.
How does its category compare?
All BSZ devicesBSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · BSZ
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · BSZ
145in 2017
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BSZ
5in 2025
Class I (38 since 2007) Class II/III not classified
Adverse event reports per year · BSZ
9,198in 2025
MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 175 | 1 | 132 |
| 2008 | 8 | 99 | 8 | 104 |
| 2009 | 3 | 34 | 3 | 85 |
| 2010 | 1 | — | 14 | 97 |
| 2011 | 6 | 218 | 1 | 87 |
| 2012 | 5 | 134 | 7 (6) | 89 |
| 2013 | 5 | 125 | 5 (5) | 308 |
| 2014 | 4 | 235 | 0 | 279 |
| 2015 | 4 | 134 | 7 (2) | 572 |
| 2016 | 1 | — | 1 | 2,135 |
| 2017 | 3 | 145 | 5 (1) | 2,355 |
| 2018 | 2 | — | 3 | 2,589 |
| 2019 | 0 | — | 5 (2) | 1,957 |
| 2020 | 1 | — | 7 | 2,687 |
| 2021 | 1 | — | 2 (1) | 3,123 |
| 2022 | 0 | — | 4 (3) | 4,352 |
| 2023 | 2 | — | 6 | 5,595 |
| 2024 | 0 | — | 13 (13) | 6,067 |
| 2025 | 1 | — | 5 (3) | 9,198 |
| 2026 | 2 | — | 7 (2) | — |
https://510kreport.com/device/K003251 · printed Nov 6, 2026
Cite this page
510(k) Report. “K003251: Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K003251
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026