510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic Device

K103625510(k) TraditionalCleared Oct 6, 2011300-day reviewClass II21 CFR 868.5895Recalls 4 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical Systems, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 6, 2011 as substantially equivalent to K984577 (Drager, Inc., 1999) and 2 other predicates after a 300-day review. Its family tree reaches back 7 generations to 1997. 3 of the 20 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 4 times, including 4 Class I recalls.

Decided
Oct 6, 2011
Review
300 days
Class
Class II
Product code
CBK
Recalls
4 (4 Class I)
Predicates
3

Its longest line of descent runs back 7 generations to K962603 (1997). 17 ancestors in all, the oldest K943531 (1995). 2 later devices descend from it, 1 citing it directly. 3 of the 20 devices in this family, counting this one, have a Class I recall.

  1. Root · 7 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 6 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 5 generations backK990966 Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric…Masimo Corp. · 1999
  4. 4 generations backK021981 Ivent 201 Portable VentilatorVersamed, Ltd. · 2003
  5. 3 generations backK042468 Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2004
  6. 2 generations backK062267 Oxylog 3000Drager Medical B.V. · 2006 · Recalled (Class I)
  7. 1 generation backK082600 Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic…Draeger Medical B.V. · 2008 · Recalled (Class I)
  8. This deviceK103625 Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic DeviceDraeger Medical Systems, Inc. · 2011
  9. Descendant · cites it directly · 1 of its ownK121886 Savina 300 Savina 300, Country VariantsDraeger Medical GmbH · 2014
  10. Descendant · 2 generations on · 0 of its ownK180779 Savina 300Dragerwerk AG & CO Kgaa · 2019

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K121886Savina 300 Savina 300, Country VariantsDraeger Medical GmbH · 2014 · cites it directly1descendant
  • K180779Savina 300Dragerwerk AG & CO Kgaa · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−6K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−5K990966Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable…Masimo Corp.1999—
−4K021981Ivent 201 Portable VentilatorVersamed, Ltd.2003—
−3K042468Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2004—
−2K062267Oxylog 3000Drager Medical B.V.2006Class I
−1K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model…Draeger Medical B.V.2008Class I
0 (this device)K103625Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic DeviceDraeger Medical Systems, Inc.2011—
+1K121886Savina 300 Savina 300, Country VariantsDraeger Medical GmbH2014—
+2K180779Savina 300Dragerwerk AG & CO Kgaa2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199520
199610
199810
199920
200010
200120
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201910
FIG. 1 — Lineage of K103625, 1997–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K984577Drager Oxylog 2000
Summary page 2
…alence: The new Oxylog 3000 plus has been compared to the following 3 predicative devices: the Oxylog 2000 plus (K082600), the Oxylog 3000 (K062267) and the Oxylog 2000 (K984577). All 4 devices are intended to be used for patients for emergency and transport. All devices contain the equivalent set of venti…
Drager, Inc.1999—
*K062267Oxylog 3000
Summary page 2
…the 510(k): Substantial Equivalence: The new Oxylog 3000 plus has been compared to the following 3 predicative devices: the Oxylog 2000 plus (K082600), the Oxylog 3000 (K062267) and the Oxylog 2000 (K984577). All 4 devices are intended to be used for patients for emergency and transport. All devices conta…
Drager Medical B.V.2006Class I
*K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model 5705082)
Summary page 2
…Brauer Reason for the 510(k): Substantial Equivalence: The new Oxylog 3000 plus has been compared to the following 3 predicative devices: the Oxylog 2000 plus (K082600), the Oxylog 3000 (K062267) and the Oxylog 2000 (K984577). All 4 devices are intended to be used for patients for emergency and t…
Draeger Medical B.V.2008Class I

Also cleared against K984577: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121886Savina 300 Savina 300, Country VariantsDraeger Medical GmbH20141

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2023Class IZ-2018-2023Device DesignDraeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
Dec 22, 2015Class IZ-0692-2016Device DesignDrager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Dec 22, 2015Class IZ-0691-2016Device DesignDrager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Mar 5, 2014Class IZ-1440-2014Process controlThe potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model 5705082)Draeger Medical B.V. · 2008 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K062267Oxylog 3000Drager Medical B.V. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 3 of the 17 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K121886 (Draeger Medical GmbH, 2014) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K984577.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Oxylog 3000 plus is a time-cycled, volume controlled and pressure controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K103625: Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103625

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026