510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

Oxylog 3000

K062267510(k) TraditionalCleared Oct 31, 200688-day reviewClass II21 CFR 868.5895Recalls 5 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Drager Medical B.V. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 31, 2006 as substantially equivalent to K984577 (Drager, Inc., 1999) and 4 other predicates after an 88-day review. Its family tree reaches back 5 generations to 1997. 11 of the 36 devices in its family, counting itself, have a Class I recall. 20 later devices descend from it. It has been recalled 5 times, including 3 Class I recalls.

Decided
Oct 31, 2006
Review
88 days
Class
Class II
Product code
CBK
Recalls
5 (3 Class I)
Predicates
5

Its longest line of descent runs back 5 generations to K962603 (1997). 15 ancestors in all, the oldest K943531 (1995). 20 later devices descend from it, 5 citing it directly. 11 of the 36 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2001: 2 other ancestors1999: 2 other ancestors1995: 2 other ancestors2000: 1 other ancestor1998: 1 other ancestor1996: 1 other ancestor2008: 1 descendant cleared, 1 with a Class I recall12011: 1 descendant cleared, 1 with a Class I recall12015: 2 descendants cleared, 1 with a Class I recall12012: 5 descendants cleared, 2 with a Class I recall22014: 1 descendant cleared2019: 2 descendants cleared, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2017: 2 descendants cleared, 1 with a Class I recall12024: 1 descendant cleared2021: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12026: 1 descendant clearedReference device: K042607 · Primus US · Draeger Medical, Inc. · 2004 · recalledK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K990966 · Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable Sensor, Masimo Lnop -PDT-Pediatric/Small Dis · Masimo Corp. · 1999K021981 · Ivent 201 Portable Ventilator · Versamed, Ltd. · 2003K042468 · Ivent 201 Portable Ventilator · Versamed Medical Systems, Inc. · 20041995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 3 other ancestors3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 5 descendants, 3 with a Class I recall32 generations forward: 6 descendants, 3 with a Class I recall33 generations forward: 3 descendants, 2 with a Class I recall24 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 2 descendants, 2 with a Class I recall27 generations forward: 1 descendant K042468 · Ivent 201 Portable Ventilator · Versamed Medical Systems, Inc. · 2004K021981 · Ivent 201 Portable Ventilator · Versamed, Ltd. · 2003K990966 · Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable Sensor, Masimo Lnop -PDT-Pediatric/Small Dis · Masimo Corp. · 1999K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−50+5
  1. Root · 5 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 4 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 3 generations backK990966 Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric…Masimo Corp. · 1999
  4. 2 generations backK021981 Ivent 201 Portable VentilatorVersamed, Ltd. · 2003
  5. 1 generation backK042468 Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2004
  6. This deviceK062267 Oxylog 3000Drager Medical B.V. · 2006
  7. Descendant · cites it directly · 13 of its ownK112006 Hamilton-TC1Hamilton Medical AG · 2012
  8. Descendant · 2 generations on · 11 of its ownK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  9. Descendant · 3 generations on · 8 of its ownK121225 Hamilton-C2Hamilton Medical AG · 2012 · Recalled (Class I)
  10. Descendant · 4 generations on · 4 of its ownK123637 Hamilton-C3Hamilton Medical AG · 2012
  11. Descendant · 5 generations on · 3 of its ownK161450 Hamilton-C3Hamilton Medical AG · 2017 · Recalled

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12012: 52014: 12015: 22016: 12017: 22019: 22020: 12021: 12023: 12024: 12026: 1
2008high 5 (2012) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 20 · See all
  • K112006Hamilton-TC1Hamilton Medical AG · 2012 · cites it directly13descendants
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 2 generations on11descendants
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 3 generations on8descendants
  • K123637Hamilton-C3Hamilton Medical AG · 2012 · 4 generations on4descendants
  • K161450Hamilton-C3Hamilton Medical AG · 2017 · 5 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−4K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−3K990966Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable…Masimo Corp.1999—
−2K021981Ivent 201 Portable VentilatorVersamed, Ltd.2003—
−1K042468Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2004—
0 (this device)K062267Oxylog 3000Drager Medical B.V.2006—
+1K112006Hamilton-TC1Hamilton Medical AG2012—
+2K120670Hamilton-T1Hamilton Medical AG2012Class I
+3K121225Hamilton-C2Hamilton Medical AG2012Class I
+4K123637Hamilton-C3Hamilton Medical AG2012—
+5K161450Hamilton-C3Hamilton Medical AG20172

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199520
199610
199810
199920
200010
200120
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200811
201111
201252
201410
201521
201611
201721
201921
202011
202111
202310
202410
202610
FIG. 1 — Lineage of K062267, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

11 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K984577Drager Oxylog 2000
Summary page 2
…044 A1Drfgerand StefensCompany Predicate Devices: 510(k) Number Device Name K984577 Oxylog 2000 K040790 LTV 1000 K(003068 Savina K(942938 Crossvent-4 K…
Drager, Inc.1999—
*K040790Modification to LTV 1000 Ventilator/Breathing Circuits
Summary page 2
…A1Drfgerand StefensCompany Predicate Devices: 510(k) Number Device Name K984577 Oxylog 2000 K040790 LTV 1000 K(003068 Savina K(942938 Crossvent-4 K(042468 Vent 201 K(04…
Pulmonetic Systems, Inc.2004—
*K003068Savina
Summary page 2
…tures of the LTV 1000 (K040790). The Oxylog 3000 incorporates a Pressure Controlled Ventilation plus mode and a Pressure Support mode identical to the Savina ventilator (K003068). Furthermore, the Oxylog 3000 uses the same operating principle as the Savina. Specific features and selling ranges of the Oxylo…
Drager Medizintechnik GmbH2001—
*K942938Crossvent-4 Ventilator
Summary page 2
…es the same operating principle as the Savina. Specific features and selling ranges of the Oxylog 3000 are found similar to those features on the Crossvent-4 ventilator (K942938) and the Vent 201 ventilator (K042468). The accessories of the Oxylog 3000 are similar to the accessories of the LTV 1000 (K04079…
Bio-Med Devices, Inc.1995—
*K042468Ivent 201 Portable Ventilator
Summary page 2
…Savina. Specific features and selling ranges of the Oxylog 3000 are found similar to those features on the Crossvent-4 ventilator (K942938) and the Vent 201 ventilator (K042468). The accessories of the Oxylog 3000 are similar to the accessories of the LTV 1000 (K040790) and the accessories of the Primus U…
Versamed Medical Systems, Inc.2004—
†K042607Primus USDraeger Medical, Inc.20042

Also cleared against K984577: 2 other devices →

What descends from it? (20)

See all 20

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112006Hamilton-TC1Hamilton Medical AG201213
K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model 5705082) Class IDraeger Medical B.V.20083
K140939Hamilton-T1,Hamilton-C1Hamilton Medical AG20152
K103625Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic Device Class IDraeger Medical Systems, Inc.20112
K141595O-Two E700,O-Two E600,O-Two E500 Class IO-Two Medical Technologies, Inc.20151

Has anything in its family been recalled? (11 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 31, 2016Class IZ-2215-2016Device DesignIsolated episodes of leakage at the check valve were found
Dec 22, 2015Class IZ-0692-2016Device DesignDrager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Dec 22, 2015Class IZ-0691-2016Device DesignDrager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Aug 19, 2010Not classifiedZ-0018-2011Error in labelingPatients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume.
Oct 3, 2008Not classifiedZ-0202-2009Software designMay experience an interruption of ventilation for approximately 5 seconds

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model 5705082)Draeger Medical B.V. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
  • K103625Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic DeviceDraeger Medical Systems, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K141595O-Two E700,O-Two E600,O-Two E500O-Two Medical Technologies, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 13 Class I recalls
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 15 Class I recalls
  • K153046Hamilton-MR1Hamilton Medical AG · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K201306Hamilton-C3Hamilton Medical AG · 2021 · 6 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K201658Hamilton-C6Hamilton Medical AG · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 9 years and passes through 4 companies; a new generation was cleared every 1.4 years on median.
  • 1 of the 15 ancestors has a recall of any class.
  • Of its descendants, K112006 (Hamilton Medical AG, 2012) has gone furthest, with 13 descendants of its own.
  • 2 other devices also cite its predicate K984577.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

Oxylog 3000 is a time-cycled, volume-constant and pressure controlled emergency and transport ventilator for patients with a tidal volume from 50 mL upwards.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062267: Oxylog 3000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K062267

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026