510(k) Report

510(k) · Cleared

Cleared

Modification to Datex-Ohmeda S/5 Avance Anesthesia System

K040743510(k) SpecialCleared Apr 22, 200430-day reviewClass II21 CFR 868.5160

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Apr 22, 2004 as substantially equivalent to K032803 (Datex-Ohmeda, 2004) and 1 other predicate after a 30-day review. Its family tree reaches back 2 generations to 1994. 12 of the 46 devices in its family, counting itself, have a Class I recall. 38 later devices descend from it.

Decided
Apr 22, 2004
Review
30 days
Class
Class II
Product code
BSZ
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K930017 (1994). 7 ancestors in all. 38 later devices descend from it, 3 citing it directly. 12 of the 46 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2000: 1 other ancestor1998: 2 other ancestors1996: 1 other ancestor2004: 1 descendant cleared2007: 1 descendant cleared2005: 2 descendants cleared2009: 2 descendants cleared2006: 1 descendant cleared2010: 1 descendant cleared2008: 2 descendants cleared2011: 4 descendants cleared, 3 with a Class I recall32012: 3 descendants cleared, 2 with a Class I recall22013: 5 descendants cleared, 1 with a Class I recall12017: 3 descendants cleared, 1 with a Class I recall12015: 3 descendants cleared2014: 2 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2016: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2019: 1 descendant cleared2026: 1 descendant clearedK930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 1994K023366 · Datex-Ohmeda 7900 Ventilator Enhancements · Datex-Ohmeda, Inc. · 2003 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors 1 generation forward: 3 descendants2 generations forward: 6 descendants3 generations forward: 5 descendants, 3 with a Class I recall34 generations forward: 8 descendants, 4 with a Class I recall45 generations forward: 6 descendants, 1 with a Class I recall16 generations forward: 4 descendants, 1 with a Class I recall17 generations forward: 3 descendants, 2 with a Class I recall28 generations forward: 2 descendants9 generations forward: 1 descendant K023366 · Datex-Ohmeda 7900 Ventilator Enhancements · Datex-Ohmeda, Inc. · 2003 · Class I recallK930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 19940+5
  1. Root · 2 generations backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. 1 generation backK023366 Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Recalled (Class I)
  3. This deviceK040743 Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc. · 2004
  4. Descendant · cites it directly · 36 of its ownK042154 GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc. · 2004 · Recalled
  5. Descendant · 2 generations on · 34 of its ownK061609 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2006 · Recalled
  6. Descendant · cites it directly · 33 of its ownK071142 GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda · 2007
  7. Descendant · 2 generations on · 30 of its ownK081844 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · Recalled
  8. Descendant · 3 generations on · 19 of its ownK090233 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · Recalled

Showing 5 of 38 descendants. See all descendants

Descendants cleared per year

2004: 12005: 22006: 12007: 12008: 22009: 22010: 12011: 42012: 32013: 52014: 22015: 32016: 12017: 32018: 22019: 12020: 12021: 12023: 12026: 1
2004high 5 (2013) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 38 · See all
  • K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc. · 2004 · cites it directly36descendants
  • K061609GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2006 · 2 generations on34descendants
  • K071142GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda · 2007 · cites it directly33descendants
  • K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · 2 generations on30descendants
  • K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · 3 generations on19descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K930017PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp.1994—
−1K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc.2003Class I
0 (this device)K040743Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc.2004—
+1K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc.20043
+2K061609GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20061
+1K071142GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda2007—
+2K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.20084
+3K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20092

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199820
200010
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200520
200610
200710
200820
200920
201010
201143
201232
201351
201421
201530
201611
201731
201820
201910
202011
202110
202311
202610
FIG. 1 — Lineage of K040743, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

12 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032803Datex-Ohmeda S/5 Avance Anesthesia System
Summary page 1
…d device: Datex-Ohmeda S/5 Avance Anesthesia System - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K032803 Datex-Ohmeda 7900 Ventilator Enhancements Class II - 21 CFR868.5160, which has been the subject of a clear…
Datex-Ohmeda2004—
*K023366Datex-Ohmeda 7900 Ventilator Enhancements
Summary page 1
…er K032803 Datex-Ohmeda 7900 Ventilator Enhancements Class II - 21 CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K023366 The Datex-Ohmeda S/5 Avance Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a…
Datex-Ohmeda, Inc.2003Class I

Also cleared against K032803: 1 other device →

What descends from it? (38)

See all 38

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc.200436
K071142GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda200733
K050626GE Datex-Ohmeda Aespire 7900 Anesthesia SystemDatex-Ohmeda, Inc.20054

Has anything in its family been recalled? (12 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K102471AS3700 Anesthsia Delivery SystemMindray DS USA, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K122445Aespire 7900, Aespire ViewDatex-Ohmeda · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K112114Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K123125GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc. · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · 5 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K151570Carestation 620/650/650CDatex-Ohmeda, Inc. · 2016 · 6 generations on · ◆ in the 2016 descendants bar of FIG. 19 Class I recalls
  • K213867Carestation 750/750cDatex-Ohmeda, Inc. · 2023 · 7 generations on · ◆ in the 2023 descendants bar of FIG. 14 Class I recalls
  • K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 9.0 years on median.
  • 4 of the 7 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K042154 (Datex-Ohmeda, Inc., 2004) has gone furthest, with 36 descendants of its own.
  • 1 other device also cites its predicate K032803.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The Datex-Ohmeda S/5 Avance Anesthesia System is intended to provide general

inhalation anesthesia and ventilatory support to a wide range of patients. The device is

intended for volume or pressure control ventilation. The S/5 Avance is not suitable for

use in a MRI environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040743: Modification to Datex-Ohmeda S/5 Avance Anesthesia System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040743

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026