510(k) Report

510(k) · Cleared

Cleared

Amsco V-Pro Max Low Temperature Sterilization System

K112760510(k) SpecialCleared Nov 22, 201161-day reviewClass II21 CFR 880.6860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MLR Sterilizer, Chemical

Cleared on Nov 22, 2011 as substantially equivalent to K102330 (STERIS Corporation, 2011) after a 61-day review. Its family tree reaches back 4 generations to 2003. 18 later devices descend from it.

Decided
Nov 22, 2011
Review
61 days
Class
Class II
Product code
MLR
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K020875 (2003). 6 ancestors in all. 18 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2005: 1 other ancestor2013: 1 descendant cleared2012: 1 descendant cleared2014: 2 descendants cleared2016: 3 descendants cleared2018: 4 descendants cleared2017: 1 descendant cleared2019: 4 descendants cleared2022: 2 descendants clearedK020875 · TSO3 Ozone Sterilizer, Model 125L · Technologies of Sterilization With Oxone, Tso3, Inc. · 2003K051595 · TSO3 Ozone Sterilizer, Model 125L · Tso3, Inc. · 2006K090636 · TSO3 Ozone Sterilizer, Model 125L · Tso3, Inc. · 2009K102330 · Amsco V-Pro Max Low Temperature Sterilization System · STERIS Corporation · 2011200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 4 descendants4 generations forward: 3 descendants5 generations forward: 2 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants K102330 · Amsco V-Pro Max Low Temperature Sterilization System · STERIS Corporation · 2011K090636 · TSO3 Ozone Sterilizer, Model 125L · Tso3, Inc. · 2009K051595 · TSO3 Ozone Sterilizer, Model 125L · Tso3, Inc. · 2006K020875 · TSO3 Ozone Sterilizer, Model 125L · Technologies of Sterilization With Oxone, Tso3, Inc. · 20030+5
  1. Root · 4 generations backK020875 TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc. · 2003
  2. 3 generations backK051595 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2006
  3. 2 generations backK090636 TSO3 Ozone Sterilizer, Model 125LTso3, Inc. · 2009
  4. 1 generation backK102330 Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2011
  5. This deviceK112760 Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2011
  6. Descendant · cites it directly · 17 of its ownK120632 Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation · 2012
  7. Descendant · cites it directly · 16 of its ownK131120 Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature…STERIS Corporation · 2013
  8. Descendant · 2 generations on · 8 of its ownK160433 V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation · 2016
  9. Descendant · 3 generations on · 7 of its ownK162413 AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1…STERIS Corporation · 2017 · Recalled
  10. Descendant · 4 generations on · 5 of its ownK172319 AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1…STERIS Corporation · 2018 · Recalled

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2012: 12013: 12014: 22016: 32017: 12018: 42019: 42022: 2
2012high 4 (2018) · low 0 (2015)2026

Most influential descendants

Showing 5 of 18 · See all
  • K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation · 2012 · cites it directly17descendants
  • K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation · 2013 · cites it directly16descendants
  • K160433V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation · 2016 · 2 generations on8descendants
  • K162413AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1 Plus Low Temperature…STERIS Corporation · 2017 · 3 generations on7descendants
  • K172319AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1 Plus Low Temperature…STERIS Corporation · 2018 · 4 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K020875TSO3 Ozone Sterilizer, Model 125LTechnologies of Sterilization With Oxone, Tso3, Inc.2003—
−3K051595TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2006—
−2K090636TSO3 Ozone Sterilizer, Model 125LTso3, Inc.2009—
−1K102330Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2011—
0 (this device)K112760Amsco V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2011—
+1K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation2012—
+1K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation2013—
+2K160433V-PRO® 1 Plus and V-PRO® maX Low TemperatureSTERIS Corporation2016—
+3K162413AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1 Plus Low…STERIS Corporation20171
+4K172319AMSCO V-PRO 1 Low Temperature Sterilization System, AMSCO V-PRO 1 Plus Low…STERIS Corporation20181

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201310
201420
201630
201710
201840
201940
202220
FIG. 1 — Lineage of K112760, 2003–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102330Amsco V-Pro Max Low Temperature Sterilization System
Summary page 2
…September 21, 2011 Page B-1 STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Modification of K102330 V-PRO maX Low Temperature Sterilization System 1. Device Namne Trade Name: Amsco V-PRO maX Low…
STERIS Corporation2011—

Also cleared against K102330: 4 other devices →

What descends from it? (18)

See all 18

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120632Amsco V-Pro 1 Low/Plus/Max Temperature Sterilization SystemSTERIS Corporation201217
K131120Amsco V-Pro 1 Low, V-Pro 1 Plus, V-Pro Max Low Temperature Sterilization SystemSTERIS Corporation201316

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • 2 of the 6 ancestors have a recall of any class.
  • Of its descendants, K120632 (STERIS Corporation, 2012) has gone furthest, with 17 descendants of its own.
  • 4 other devices also cite its predicate K102330.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The Amsco® V-PRO™ maX Low Temperature Sterilization System, with VAPROX™ HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for…

Read the full statement (588 words)

The Amsco® V-PRO™ maX Low Temperature Sterilization System, with VAPROX™ HC Sterilant, is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed and dried reusable metal and nonmetal medical devices used in healthcare facilities. The three pre-programmed sterilization cycles (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) operate at low pressure and low temperature and are thus suitable for processing medical devices sensitive to heat and moisture.

The Amsco V-PRO maX Low Temperature Sterilization System's Lumen Cycle, cleared under K062297, can sterilize:

* Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors

* Medical devices, including rigid endoscopes, with a single stainless steel lumen with:

o an inside diameter of 1 mm or larger and a length of 125 mm or shorter

o an inside diameter of 2 mm or larger and a length of 250 mm or shorter

o an inside diameter of 3 mm or larger and a length of 400 mm or shorter

The validation testing for all lumen sizes was conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

The Amsco V-PRO maX Low Temperature Sterilization System's Non Lumen Cycle, cleared under K083097, can sterilize:

Non-lumened instruments including non-lumened rigid endoscopes and non-lumened instruments with stainless steel diffusion-restricted areas such as the hinged portion of forceps or scissors.

The validation studies were conducted using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs.

The Amsco V-PRO maX Low Temperature Sterilization System's Flexible Cycle, the subject of this submission, can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of two load configurations:

1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load

The flexible endoscopes may contain either:

*a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter

*or two lumens with:

*one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter

*and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter

The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).

2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.

The flexible endoscope can contain either:

*a single lumen with an inside diameter of 1 mm or larger and a length of 1050 mm or shorter

*or two lumens with:

*one lumen having an inside diameter of 1 mm or larger and a length of 998 mm or shorter

*and the other lumen having an inside diameter of 1 mm or larger and a length of 850 mm or shorter

The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112760: Amsco V-Pro Max Low Temperature Sterilization System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K112760

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026