Model IV 100B-DA Infant Ventilator

K833982510(k) TraditionalCleared Mar 5, 1984109-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sechrist Industries, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Mar 5, 1984 after a 109-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 9 of the 26 devices in its family, counting itself, have a Class I recall. 25 later devices descend from it. It has been recalled once.

Decided
Mar 5, 1984
Review
109 days
Class
Class II
Product code
CBK
Recalls
1
Predicates
0

25 later devices descend from it, 2 citing it directly. 9 of the 26 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 1 descendant cleared2004: 1 descendant cleared2006: 1 descendant cleared2012: 7 descendants cleared, 2 with a Class I recall22008: 2 descendants cleared, 1 with a Class I recall12007: 1 descendant cleared2013: 1 descendant cleared, 1 with a Class I recall12009: 1 descendant cleared2011: 1 descendant cleared2015: 1 descendant cleared2017: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared2026: 1 descendant cleared19801990200020102020
  1. This deviceK833982 Model IV 100B-DA Infant VentilatorSechrist Industries, Inc. · 1984
  2. Descendant · cites it directly · 22 of its ownK031745 Infant Flow Plus Infant Cpap System with Sipap FunctionSensor Medics Corp. · 2004
  3. Descendant · 2 generations on · 21 of its ownK062093 Avea VentilatorViasys Respiratory Care, Inc. · 2006 · Recalled
  4. Descendant · 3 generations on · 16 of its ownK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)
  5. Descendant · 4 generations on · 14 of its ownK102775 Hamilton-C2Hamilton Medical AG · 2011
  6. Descendant · 5 generations on · 13 of its ownK112006 Hamilton-TC1Hamilton Medical AG · 2012

Showing 5 of 25 descendants. See all descendants

Descendants cleared per year

1999: 12004: 12006: 12007: 12008: 22009: 12011: 12012: 72013: 12015: 12016: 12017: 22019: 12020: 12021: 12024: 12026: 1
1999high 7 (2012) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 25 · See all
  • K031745Infant Flow Plus Infant Cpap System with Sipap FunctionSensor Medics Corp. · 2004 · cites it directly22descendants
  • K062093Avea VentilatorViasys Respiratory Care, Inc. · 2006 · 2 generations on21descendants
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 3 generations on16descendants
  • K102775Hamilton-C2Hamilton Medical AG · 2011 · 4 generations on14descendants
  • K112006Hamilton-TC1Hamilton Medical AG · 2012 · 5 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K833982Model IV 100B-DA Infant VentilatorSechrist Industries, Inc.1984—
+1K031745Infant Flow Plus Infant Cpap System with Sipap FunctionSensor Medics Corp.2004—
+2K062093Avea VentilatorViasys Respiratory Care, Inc.20065
+3K081521Modification to Hamilton-G5Hamilton Medical AG2008Class I
+4K102775Hamilton-C2Hamilton Medical AG2011—
+5K112006Hamilton-TC1Hamilton Medical AG2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200410
200610
200710
200821
200910
201110
201272
201311
201510
201611
201721
201911
202011
202111
202410
202610
FIG. 1 — Lineage of K833982, 1984–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (25)

See all 25

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031745Infant Flow Plus Infant Cpap System with Sipap FunctionSensor Medics Corp.200422
K993167Millennium Infant VentilatorSechrist Industries, Inc.19991

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 20, 2004Not classifiedZ-1032-04OtherElectronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 3 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K130178Inspiration Ventilator System 5I/7IeVent Medical, Ltd. · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 5 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 6 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · 7 generations on · ◆ in the 2019 descendants bar of FIG. 15 Class I recalls
  • K153046Hamilton-MR1Hamilton Medical AG · 2016 · 7 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K201306Hamilton-C3Hamilton Medical AG · 2021 · 8 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K201658Hamilton-C6Hamilton Medical AG · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • Of its descendants, K031745 (Sensor Medics Corp., 2004) has gone furthest, with 22 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—
Cite this page

510(k) Report. “K833982: Model IV 100B-DA Infant Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K833982

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026