510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Hemotherm Model 400CE Cooler/Heater System

K122813510(k) TraditionalCleared Dec 18, 201296-day reviewClass II21 CFR 870.4250Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cincinnati Sub-Zero Products, Inc. · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on Dec 18, 2012 as substantially equivalent to K811742 (Cincinnati Sub-Zero, 1981) and 2 other predicates after a 96-day review. Its family tree reaches back 4 generations to 2000. It has been recalled 6 times.

Decided
Dec 18, 2012
Review
96 days
Class
Class II
Product code
DWC
Recalls
6
Predicates
3

Its longest line of descent runs back 4 generations to K003349 (2000). 10 ancestors in all, the oldest K811742 (1981). No later device cites it as a predicate yet.

  1. Root · 4 generations backK003349 Allon 2001 ModifiedMtre Advanced Technology, Ltd. · 2000
  2. 3 generations backK024128 Allon 2001 Version 5Mtre Advanced Technology, Ltd. · 2003 · Recalled
  3. 2 generations backK083662 CriticoolMtre Advanced Technology, Ltd. · 2009 · Recalled
  4. 1 generation backK101589 Blanketrol Iii, Model 233 and Coolblue, Model 25-01…Cincinnati Sub-Zero Products, Inc. · 2010 · Recalled
  5. This deviceK122813 Hemotherm Model 400CE Cooler/Heater SystemCincinnati Sub-Zero Products, Inc. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K003349Allon 2001 ModifiedMtre Advanced Technology, Ltd.2000—
−3K024128Allon 2001 Version 5Mtre Advanced Technology, Ltd.20031
−2K083662CriticoolMtre Advanced Technology, Ltd.20091
−1K101589Blanketrol Iii, Model 233 and Coolblue, Model 25-01 Hyper-Hypothermia SystemsCincinnati Sub-Zero Products, Inc.20106
0 (this device)K122813Hemotherm Model 400CE Cooler/Heater SystemCincinnati Sub-Zero Products, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198810
199110
200110
200310
200710
FIG. 1 — Lineage of K122813, 2000–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K811742Hemotherm Model 400
Summary page 1
…DWJ when used with a hypo-hyperthermia blanket 9. PANEL: Cardiovascular 10. DEVICE CLASSIFICATION: Class 11 11. IDENTIFICATION OF PREDICATES: *K81 1742 Cincinnati Sub-Zero Hemnotherm Model 400 Cooler/Heater System *K03 1544 Jostra HCU 30 Heater-Cooler Dev…
Cincinnati Sub-Zero19812
*K031544Heater Cooler Unit, Model Hcu 30
Summary page 1
…IFICATION: Class 11 11. IDENTIFICATION OF PREDICATES: *K81 1742 Cincinnati Sub-Zero Hemnotherm Model 400 Cooler/Heater System *K03 1544 Jostra HCU 30 Heater-Cooler Device *K101589 Cincinnati Sub-Zero Blanketrol III Model 233 and blankets DEVICE DESCRIP…
Jostra AB20034
*K101589Blanketrol Iii, Model 233 and Coolblue, Model 25-01 Hyper-Hypothermia Systems
Summary page 1
…: *K81 1742 Cincinnati Sub-Zero Hemnotherm Model 400 Cooler/Heater System *K03 1544 Jostra HCU 30 Heater-Cooler Device *K101589 Cincinnati Sub-Zero Blanketrol III Model 233 and blankets DEVICE DESCRIPTION The Hemotherm® Model 400CE Cooler/Heater System is…
Cincinnati Sub-Zero Products, Inc.20106

Also cleared against K811742: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 17, 2021Class IIZ-2092-2021Process controlThere is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler…
Dec 19, 2018Class IIZ-0825-2019Device DesignEvidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration…
Sep 4, 2018Class IIZ-0063-2019Device DesignPotential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.
Nov 27, 2017Class IIZ-0629-2018Device DesignDue to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.
Mar 9, 2017Class IIZ-1825-2017Component design/selectionImproperly rated fuses were used in the manufacture of the device.
Nov 7, 2016Class IIZ-0654-2017Material/Component ContaminationThe device may not maintain water flow or temperature control.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 2.2 years on median.
  • 9 of the 10 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K811742.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The Hemotherm® Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator/Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper/hypothermia blanket under the patient to provide warming through conductive heat transfer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122813: Hemotherm Model 400CE Cooler/Heater System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122813

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026