510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Oec Brivo 865 Advance, Oec Brivo 785 Essential, and Oec Brivo 715 Prime

K123603510(k) TraditionalCleared May 17, 2013177-day reviewClass II21 CFR 892.1650Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Healthcare Surgery GE Oec Medical Systems, Inc. · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on May 17, 2013 as substantially equivalent to K111551 (Ge Oec Medical Systems, Inc., 2011) after a 177-day review. Its family tree reaches back 5 generations to 2002. 2 of the 11 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 5 times.

Decided
May 17, 2013
Review
177 days
Class
Class II
Product code
OXO
Recalls
5
Predicates
1

Its longest line of descent runs back 5 generations to K021049 (2002). 6 ancestors in all. 4 later devices descend from it, 1 citing it directly. 2 of the 11 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor, 1 with a Class I recall12017: 1 descendant cleared2018: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant clearedK021049 · Oec 9800 Plus Digital Mobile Imaging System · Ge Oec Medical Systems · 2002 · Class I recallK041932 · Oec Olympus Fluoroscopic Imaging System · Ge Oec Medical Systems, Inc. · 2004 · recalledK073543 · Oec 9900 Elite · Ge Oec Medical Systems · 2008 · recalledK082781 · Oec 9900 Elite · Ge Healthcare Surgery · 2009 · recalledK111551 · Oec (R) 9800, Oec (R) 9800 Plus · Ge Oec Medical Systems, Inc. · 2011 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor, 1 with a Class I recall1 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K111551 · Oec (R) 9800, Oec (R) 9800 Plus · Ge Oec Medical Systems, Inc. · 2011 · recalledK082781 · Oec 9900 Elite · Ge Healthcare Surgery · 2009 · recalledK073543 · Oec 9900 Elite · Ge Oec Medical Systems · 2008 · recalledK041932 · Oec Olympus Fluoroscopic Imaging System · Ge Oec Medical Systems, Inc. · 2004 · recalledK021049 · Oec 9800 Plus Digital Mobile Imaging System · Ge Oec Medical Systems · 2002 · Class I recall−50
  1. Root · 5 generations backK021049 Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · Recalled (Class I)
  2. 4 generations backK041932 Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · Recalled
  3. 3 generations backK073543 Oec 9900 EliteGe Oec Medical Systems · 2008 · Recalled
  4. 2 generations backK082781 Oec 9900 EliteGe Healthcare Surgery · 2009 · Recalled
  5. 1 generation backK111551 Oec (R) 9800, Oec (R) 9800 PlusGe Oec Medical Systems, Inc. · 2011 · Recalled
  6. This deviceK123603 Oec Brivo 865 Advance, Oec Brivo 785 Essential, and Oec Brivo 715…Ge Healthcare Surgery GE Oec Medical Systems, Inc. · 2013
  7. Descendant · cites it directly · 3 of its ownK172700 OEC OneGe Hualun Medical Systems Co., Ltd. · 2017
  8. Descendant · 2 generations on · 2 of its ownK182626 OEC OneGe Hualun Medical Systems Co., Ltd. · 2018
  9. Descendant · 3 generations on · 1 of its ownK240828 OEC One ASDGe Hualun Medical Systems Co., Ltd. · 2024
  10. Descendant · 4 generations on · 0 of its ownK253269 OEC One CFDGe Hualun Medical Systems Co., Ltd. · 2025

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K172700OEC OneGe Hualun Medical Systems Co., Ltd. · 2017 · cites it directly3descendants
  • K182626OEC OneGe Hualun Medical Systems Co., Ltd. · 2018 · 2 generations on2descendants
  • K240828OEC One ASDGe Hualun Medical Systems Co., Ltd. · 2024 · 3 generations on1descendant
  • K253269OEC One CFDGe Hualun Medical Systems Co., Ltd. · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems2002Class I
−4K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.20047
−3K073543Oec 9900 EliteGe Oec Medical Systems20085
−2K082781Oec 9900 EliteGe Healthcare Surgery20094
−1K111551Oec (R) 9800, Oec (R) 9800 PlusGe Oec Medical Systems, Inc.20119
0 (this device)K123603Oec Brivo 865 Advance, Oec Brivo 785 Essential, and Oec Brivo 715 PrimeGe Healthcare Surgery GE Oec Medical Systems, Inc.2013—
+1K172700OEC OneGe Hualun Medical Systems Co., Ltd.2017—
+2K182626OEC OneGe Hualun Medical Systems Co., Ltd.2018—
+3K240828OEC One ASDGe Hualun Medical Systems Co., Ltd.2024—
+4K253269OEC One CFDGe Hualun Medical Systems Co., Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201810
202410
202510
FIG. 1 — Lineage of K123603, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K253296 Digital C-ARM with 3Dscore 15.1 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111551Oec (R) 9800, Oec (R) 9800 Plus
Summary page 1
…stem Classification Names: 21 CFR 892.1650 Image-intensified fluoroscopic x-ray system Product Code: 900X0, 9OJAA Predicate Device(s): K 111551 OEC® 9800 / OEC® 9800 Plus Device Description: The OEC®' Brivo®o is a system used to assist trained…
Ge Oec Medical Systems, Inc.20119

Also cleared against K111551: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172700OEC OneGe Hualun Medical Systems Co., Ltd.20173

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 23, 2016Class IIZ-0071-2017Nonconforming Material/ComponentGE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm…
Sep 15, 2014Class IIZ-0840-2015Radiation Control for Health and Safety ActmA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in…
Sep 15, 2014Class IIZ-0841-2015Radiation Control for Health and Safety ActmA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in…
May 22, 2014Class IIZ-1638-2015Radiation Control for Health and Safety ActLaser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.
Sep 18, 2013Class IIZ-0066-2014Device DesignGE is recalling certain lots of GE Brivo OEC715/785/865 C-Arm due to the potential for the wires in the high voltage cable bundle to fail, resulting in loss of system functionality.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 6 of the 6 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K172700 (Ge Hualun Medical Systems Co., Ltd., 2017) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K111551.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

The OEC® Brivo® Mobile C-Arm X-Ray Products are designed to provide digital spot-film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localization/s or needle positions for aspirations, injections or biopsy. The OEC® Brivo® is not indicated for interventional use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123603: Oec Brivo 865 Advance, Oec Brivo 785 Essential, and Oec Brivo 715 Prime.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K123603

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.