510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Oec 9800 Plus

K132027510(k) SpecialCleared May 1, 2014304-day reviewClass II21 CFR 892.1650Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Oec Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on May 1, 2014 as substantially equivalent to K111551 (Ge Oec Medical Systems, Inc., 2011) after a 304-day review. Its family tree reaches back 5 generations to 2002. 2 of the 13 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 8 times.

Decided
May 1, 2014
Review
304 days
Class
Class II
Product code
OWB
Recalls
8
Predicates
1

Its longest line of descent runs back 5 generations to K021049 (2002). 6 ancestors in all. 6 later devices descend from it, 2 citing it directly. 2 of the 13 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor, 1 with a Class I recall12025: 1 descendant cleared2019: 1 descendant cleared2020: 2 descendants cleared2021: 2 descendants clearedK021049 · Oec 9800 Plus Digital Mobile Imaging System · Ge Oec Medical Systems · 2002 · Class I recallK041932 · Oec Olympus Fluoroscopic Imaging System · Ge Oec Medical Systems, Inc. · 2004 · recalledK073543 · Oec 9900 Elite · Ge Oec Medical Systems · 2008 · recalledK082781 · Oec 9900 Elite · Ge Healthcare Surgery · 2009 · recalledK111551 · Oec (R) 9800, Oec (R) 9800 Plus · Ge Oec Medical Systems, Inc. · 2011 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor, 1 with a Class I recall1 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K111551 · Oec (R) 9800, Oec (R) 9800 Plus · Ge Oec Medical Systems, Inc. · 2011 · recalledK082781 · Oec 9900 Elite · Ge Healthcare Surgery · 2009 · recalledK073543 · Oec 9900 Elite · Ge Oec Medical Systems · 2008 · recalledK041932 · Oec Olympus Fluoroscopic Imaging System · Ge Oec Medical Systems, Inc. · 2004 · recalledK021049 · Oec 9800 Plus Digital Mobile Imaging System · Ge Oec Medical Systems · 2002 · Class I recall−50
  1. Root · 5 generations backK021049 Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · Recalled (Class I)
  2. 4 generations backK041932 Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · Recalled
  3. 3 generations backK073543 Oec 9900 EliteGe Oec Medical Systems · 2008 · Recalled
  4. 2 generations backK082781 Oec 9900 EliteGe Healthcare Surgery · 2009 · Recalled
  5. 1 generation backK111551 Oec (R) 9800, Oec (R) 9800 PlusGe Oec Medical Systems, Inc. · 2011 · Recalled
  6. This deviceK132027 Oec 9800 PlusGe Oec Medical Systems, Inc. · 2014
  7. Descendant · cites it directly · 4 of its ownK183109 ControlRad Trace Model 8Controlrad, Inc. · 2019
  8. Descendant · 2 generations on · 3 of its ownK200663 ControlRad Trace Model 9Controlrad, Inc. · 2020
  9. Descendant · 3 generations on · 2 of its ownK202431 ControlRad Select Model ZControlrad, Inc. · 2020
  10. Descendant · cites it directly · 0 of its ownK251650 Insight Enhanced™ DRF (EN-1002-01)Imaging Engineering, LLC · 2025
  11. Descendant · 4 generations on · 0 of its ownK213455 ControlRad Select Model ZControlrad, Inc. · 2021

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2019: 12020: 22021: 22025: 1
2019high 2 (2020) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K183109ControlRad Trace Model 8Controlrad, Inc. · 2019 · cites it directly4descendants
  • K200663ControlRad Trace Model 9Controlrad, Inc. · 2020 · 2 generations on3descendants
  • K202431ControlRad Select Model ZControlrad, Inc. · 2020 · 3 generations on2descendants
  • K251650Insight Enhanced™ DRF (EN-1002-01)Imaging Engineering, LLC · 2025 · cites it directly0descendants
  • K213455ControlRad Select Model ZControlrad, Inc. · 2021 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems2002Class I
−4K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.20047
−3K073543Oec 9900 EliteGe Oec Medical Systems20085
−2K082781Oec 9900 EliteGe Healthcare Surgery20094
−1K111551Oec (R) 9800, Oec (R) 9800 PlusGe Oec Medical Systems, Inc.20119
0 (this device)K132027Oec 9800 PlusGe Oec Medical Systems, Inc.2014—
+1K183109ControlRad Trace Model 8Controlrad, Inc.2019—
+2K200663ControlRad Trace Model 9Controlrad, Inc.2020—
+3K202431ControlRad Select Model ZControlrad, Inc.2020—
+1K251650Insight Enhanced™ DRF (EN-1002-01)Imaging Engineering, LLC2025—
+4K213455ControlRad Select Model ZControlrad, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202020
202120
202510
FIG. 1 — Lineage of K132027, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111551Oec (R) 9800, Oec (R) 9800 PlusGe Oec Medical Systems, Inc.20119

Also cleared against K111551: 1 other device →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183109ControlRad Trace Model 8Controlrad, Inc.20194
K251650Insight Enhanced™ DRF (EN-1002-01)Imaging Engineering, LLC20250

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 2, 2025Class IIZ-2341-2025Radiation Control for Health and Safety ActGE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will…
Nov 21, 2024Class IIZ-0706-2025Process controlDue to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
May 15, 2023Class IIZ-2279-2023Process controlImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries…
May 15, 2023Class IIZ-2276-2023Process controlImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries…
May 8, 2020Class IIZ-2872-2020Radiation Control for Health and Safety ActGEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live…
Dec 27, 2016Class IIZ-1059-2017Packaging change controlGE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement…
Dec 19, 2016Class IIZ-0973-2017Use errorGE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Oct 3, 2014Class IIZ-0195-2015Component design/selectionGE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 6 of the 6 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K183109 (Controlrad, Inc., 2019) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K111551.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The QEC 9800 Plus is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography. endoscopic. urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K132027: Oec 9800 Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K132027

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026