510(k) Report

510(k) · Cleared

Cleared

INOmax DSIR Plus

K200389510(k) SpecialCleared Jun 17, 2020120-day reviewClass II21 CFR 868.5165

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mallinckrodt Manufacturing, LLC · MRN Apparatus, Nitric Oxide Delivery

Cleared on Jun 17, 2020 as substantially equivalent to K131686 (Ino Therapeutics/Ikaria, 2013) after a 120-day review. Its family tree reaches back 12 generations to 2001. 3 later devices descend from it.

Decided
Jun 17, 2020
Review
120 days
Class
Class II
Product code
MRN
Recalls
None
Predicates
1

Its longest line of descent runs back 12 generations to K003665 (2001). 20 ancestors in all, the oldest K000186 (2000). 3 later devices descend from it, 2 citing it directly.

  1. Root · 12 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 8 generations between
  3. 3 generations backK121021 Inomax DS (Delivery System)Ino Therapeutics · 2012
  4. 2 generations backK130605 Inomax Dsir (Delivery System)Ino Therapeutics · 2013
  5. 1 generation backK131686 Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria · 2013 · Recalled
  6. This deviceK200389 INOmax DSIR PlusMallinckrodt Manufacturing, LLC · 2020
  7. Descendant · cites it directly · 1 of its ownK212409 Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics · 2023
  8. Descendant · 2 generations on · 0 of its ownK242374 ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC · 2025
  9. Descendant · cites it directly · 0 of its ownK211153 Inomax DSIRMallinckrodt Manufacturing, LLC · 2023
Show all 12 generations
  1. Root · 12 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 11 generations backK011874 Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  3. 10 generations backK052663 Ino Therapeutics Inoblender, Model 10004Ino Therapeutics · 2005
  4. 9 generations backK061901 Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2006 · Recalled
  5. 8 generations backK070867 Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC · 2007
  6. 7 generations backK071516 Inomax DSIno Therapeutics, LLC · 2007
  7. 6 generations backK080484 Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2008
  8. 5 generations backK110635 Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria · 2011
  9. 4 generations backK113272 Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria · 2012
  10. 3 generations backK121021 Inomax DS (Delivery System)Ino Therapeutics · 2012
  11. 2 generations backK130605 Inomax Dsir (Delivery System)Ino Therapeutics · 2013
  12. 1 generation backK131686 Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria · 2013 · Recalled
  13. This deviceK200389 INOmax DSIR PlusMallinckrodt Manufacturing, LLC · 2020

Most influential descendants

Showing 3 of 3 · See all
  • K212409Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics · 2023 · cites it directly1descendant
  • K242374ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC · 2025 · 2 generations on0descendants
  • K211153Inomax DSIRMallinckrodt Manufacturing, LLC · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K003665Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−11K011874Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−10K052663Ino Therapeutics Inoblender, Model 10004Ino Therapeutics2005—
−9K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics20063
−8K070867Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC2007—
−7K071516Inomax DSIno Therapeutics, LLC2007—
−6K080484Inomax DS (Delivery System), Model 10003Ino Therapeutics2008—
−5K110635Inomax DS (IR) (Delivery System)Ino Therapeutics, LLC/Dba Ikaria2011—
−4K113272Inomax DS (Delivery SystemsIno Therapeutics, LLC/Dba Ikaria2012—
−3K121021Inomax DS (Delivery System)Ino Therapeutics2012—
−2K130605Inomax Dsir (Delivery System)Ino Therapeutics2013—
−1K131686Inomax Dsir (Delivery System)Ino Therapeutics/Ikaria20134
0 (this device)K200389INOmax DSIR PlusMallinckrodt Manufacturing, LLC2020—
+1K212409Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics2023—
+2K242374ULSPIRA TS™ Nitric Oxide Therapy SystemAirgas Therapeutics, LLC2025—
+1K211153Inomax DSIRMallinckrodt Manufacturing, LLC2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200310
200810
200910
201020
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202320
202510
FIG. 1 — Lineage of K200389, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRN device

  1. #1 K242374 ULSPIRA TS™ Nitric Oxide Therapy Systemscore 12.6 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131686Inomax Dsir (Delivery System)
Summary page 4
…kup Delivery Device Classification: Class II – 21 CFR 868.5165 Product Code: MRN (Primary), MRQ, MRP Predicate Device(s) K131686 Description of Device The INOmax DSIR® Plus uses a "dual-channel" design to ensure the safe delivery…
Ino Therapeutics/Ikaria20134

Also cleared against K131686: 2 other devices →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K212409Ulspira TS Nitric Oxide Therapy SystemAirgas Therapeutics20231
K211153Inomax DSIRMallinckrodt Manufacturing, LLC20230

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 8 months on median.
  • 4 of the 20 ancestors have a recall of any class.
  • Of its descendants, K212409 (Airgas Therapeutics, 2023) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K131686.

How does its category compare?

All MRN devices

MRN Apparatus, Nitric Oxide Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MRN

2in 2025

2007: 22008: 22009: 12010: 22011: 22012: 32013: 22015: 12018: 12020: 32022: 12023: 42024: 22025: 2
2007high 4 (2023) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRN

566in 2023

2012: 117 days median2020: 84 days median2023: 566 days median
2007high 566 (2023) · low 84 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MRN

2in 2025

2010: 22014: 12023: 12025: 2, 2 Class I2026 (partial year): 2
2007high 2 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MRN

155in 2025

2007: 32008: 92009: 102010: 1552011: 552012: 242013: 322014: 792015: 1562016: 942017: 1062018: 1382019: 1442020: 1992021: 2212022: 1152023: 992024: 1382025: 155
2007high 221 (2021) · low 3 (2007)2025

MAUDE reports naming product code MRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—03
20082—09
20091—010
20102—2155
20112—055
20123117024
20132—032
20140—179
20151—0156
20160—094
20170—0106
20181—0138
20190—0144
20203840199
20210—0221
20221—0115
20234566199
20242—0138
20252—2 (2)155
20260—2—

What is it indicated for?

The INOmax® DSIR Plus delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It…

Read the full statement (210 words)

The INOmax® DSIR Plus delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200389: INOmax DSIR Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K200389

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026