510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

OEC 3D

K233669510(k) TraditionalCleared Mar 28, 2024134-day reviewClass II21 CFR 892.1650Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Oec Medical Systems, Inc. · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on Mar 28, 2024 as substantially equivalent to K203346 (Ge Oec Medical Systems, Inc., 2021) after a 134-day review. Its family tree reaches back 11 generations to 2002. 2 of the 13 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times.

Decided
Mar 28, 2024
Review
134 days
Class
Class II
Product code
OXO
Recalls
2
Predicates
1

Its longest line of descent runs back 11 generations to K021049 (2002). 12 ancestors in all. No later device cites it as a predicate yet. 2 of the 13 devices in this family, counting this one, have a Class I recall.

  1. Root · 11 generations backK021049 Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · Recalled (Class I)
  2. 7 generations between
  3. 3 generations backK172550 OEC EliteGe Oec Medical Systems, Inc. · 2017 · Recalled
  4. 2 generations backK192819 OEC EliteGe Oec Medical Systems, Inc. · 2019 · Recalled
  5. 1 generation backK203346 OEC 3DGe Oec Medical Systems, Inc. · 2021 · Recalled
  6. This deviceK233669 OEC 3DGe Oec Medical Systems, Inc. · 2024
Show all 11 generations
  1. Root · 11 generations backK021049 Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · Recalled (Class I)
  2. 10 generations backK041932 Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · Recalled
  3. 9 generations backK073543 Oec 9900 EliteGe Oec Medical Systems · 2008 · Recalled
  4. 8 generations backK082781 Oec 9900 EliteGe Healthcare Surgery · 2009 · Recalled
  5. 7 generations backK120613 Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · Recalled
  6. 6 generations backK122234 Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · Recalled
  7. 5 generations backK170752 OEC EliteOec Medical Systems, Inc. · 2017 · Recalled
  8. 4 generations backK171565 OEC EliteGe Oec Medical Systems, Inc. · 2017 · Recalled
  9. 3 generations backK172550 OEC EliteGe Oec Medical Systems, Inc. · 2017 · Recalled
  10. 2 generations backK192819 OEC EliteGe Oec Medical Systems, Inc. · 2019 · Recalled
  11. 1 generation backK203346 OEC 3DGe Oec Medical Systems, Inc. · 2021 · Recalled
  12. This deviceK233669 OEC 3DGe Oec Medical Systems, Inc. · 2024

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems2002Class I
−10K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.20047
−9K073543Oec 9900 EliteGe Oec Medical Systems20085
−8K082781Oec 9900 EliteGe Healthcare Surgery20094
−7K120613Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc.20127
−6K122234Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc.20126
−5K170752OEC EliteOec Medical Systems, Inc.20174
−4K171565OEC EliteGe Oec Medical Systems, Inc.20174
−3K172550OEC EliteGe Oec Medical Systems, Inc.20174
−2K192819OEC EliteGe Oec Medical Systems, Inc.20194
−1K203346OEC 3DGe Oec Medical Systems, Inc.20213
0 (this device)K233669OEC 3DGe Oec Medical Systems, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200311
FIG. 1 — Lineage of K233669, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K253296 Digital C-ARM with 3Dscore 15.1 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K203346OEC 3D
Summary page 4
…Salt Lake City, Utah 84116 Predicate Device: Device Name: OEC 3D 510(k) number: K203346 Manufacturer: GE OEC Medical Systems, Inc Page 1 of…
Ge Oec Medical Systems, Inc.20213

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 21, 2025Class IIZ-1820-2025Process controlOEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Nov 21, 2024Class IIZ-0703-2025Process controlDue to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 10 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 10 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 12 of the 12 ancestors have a recall of any class; 2 have a Class I recall.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K233669: OEC 3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K233669

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026