Philips Ultrasound, Inc. · LLZ System, Image Processing, Radiological
Cleared on Apr 17, 2014 as substantially equivalent to K061995 (Philips Medical Systems North America Co., 2006) and 1 other predicate after a 16-day review. Its family tree reaches back 16 generations to 1995. 6 of the 136 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times.
- Decided
- Apr 17, 2014
- Review
- 16 days
- Class
- Class II
- Product code
- LLZ
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 16 generations to K935595 (1995). 135 ancestors in all, the oldest K882383 (1988). No later device cites it as a predicate yet. 6 of the 136 devices in this family, counting this one, have a Class I recall.
- Root · 16 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
- 12 generations between
- 3 generations backK112255 AixplorerSupersonic Imagine · 2012
- 2 generations backK130499 IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- 1 generation backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- This deviceK140808 Q-StationPhilips Ultrasound, Inc. · 2014
Show all 16 generations
- Root · 16 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
- 15 generations backK973767 Harmonic Imaging with ContrastAcuson Corp. · 1997
- 14 generations backK022567 Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp. · 2002
- 13 generations backK032114 Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2003
- 12 generations backK041319 Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004
- 11 generations backK042770 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 10 generations backK050240 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
- 9 generations backK052894 Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
- 8 generations backK063138 Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2006
- 7 generations backK063803 Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · Recalled
- 6 generations backK072786 Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2007
- 5 generations backK091970 Aixplorer Ultrasound SystemSupersonic Imagine · 2009
- 4 generations backK102041 AixplorerSupersonic Imagine · 2010
- 3 generations backK112255 AixplorerSupersonic Imagine · 2012
- 2 generations backK130499 IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- 1 generation backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- This deviceK140808 Q-StationPhilips Ultrasound, Inc. · 2014
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −16 | K935595 | Acuson Model 3001 | Acuson Corp. | 1995 | — |
| −15 | K973767 | Harmonic Imaging with Contrast | Acuson Corp. | 1997 | — |
| −14 | K022567 | Sequoia Signature II Diagnostic Ultrasound System | Acuson Corp. | 2002 | — |
| −13 | K032114 | Sequoia 8.0 Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc. | 2003 | — |
| −12 | K041319 | Sequoia Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc. | 2004 | — |
| −11 | K042770 | Modification to Acuson CV70 Cardiovascular System | Siemens Medical Solutions USA, Inc. | 2004 | 1 |
| −10 | K050240 | Modification to Acuson CV70 Cardiovascular System | Siemens Medical Solutions USA, Inc. | 2005 | 1 |
| −9 | K052894 | Sonoline G60 S Ultrasound System | Siemens Medical Solutions USA, Inc. | 2005 | 1 |
| −8 | K063138 | Modification to Acuson Antares Ultrasound System | Siemens Medical Solutions USA, Inc. | 2006 | — |
| −7 | K063803 | Acuson Antares Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc., Ultrasound DI | 2007 | 9 |
| −6 | K072786 | Acuson S2000 Ultrasound System | Siemens Medical Solutions USA, Inc. | 2007 | — |
| −5 | K091970 | Aixplorer Ultrasound System | Supersonic Imagine | 2009 | — |
| −4 | K102041 | Aixplorer | Supersonic Imagine | 2010 | — |
| −3 | K112255 | Aixplorer | Supersonic Imagine | 2012 | — |
| −2 | K130499 | IU22 Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2013 | 2 |
| −1 | K132304 | Epiq Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2013 | 36 |
| 0 (this device) | K140808 | Q-Station | Philips Ultrasound, Inc. | 2014 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1989 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 5 | 0 |
| 1995 | 6 | 0 |
| 1996 | 5 | 1 |
| 1997 | 7 | 0 |
| 1998 | 10 | 3 |
| 1999 | 11 | 0 |
| 2000 | 13 | 1 |
| 2001 | 11 | 0 |
| 2002 | 15 | 0 |
| 2003 | 8 | 0 |
| 2004 | 13 | 1 |
| 2005 | 6 | 0 |
| 2006 | 3 | 0 |
| 2010 | 1 | 0 |
| 2012 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LLZ device
- #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K061995 | XceleraSummary page 1…lips Ultrasound believes that modifications introduced with Q-Station 3.0 are substantially equivalent to other commercially available devices including Philips Xcelera (K061995) and Philips EPIQ (K132304). The Q-Station analysis packages and multi-modality viewing are the same as those found on Philips ul… | 2006 |
| * | K132304 | Epiq Diagnostic Ultrasound SystemSummary page 1…at modifications introduced with Q-Station 3.0 are substantially equivalent to other commercially available devices including Philips Xcelera (K061995) and Philips EPIQ (K132304). The Q-Station analysis packages and multi-modality viewing are the same as those found on Philips ultrasound systems, including… | 2013 |
| † | K132165 | Qlab Quantification Software | 2013 |
| † | K103815 | Q-Station | 2011 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (6 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 15, 2015 | Class II | Z-1807-2015 | Software design | Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients. |
| Nov 5, 2014 | Class II | Z-0818-2015 | Software design | When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
- K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K960330InstatrakVisualization Technology, Inc. · 1996 · 4 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
Nearest 5 of 6; all are on the web page.
What does this lineage say?
- The line of descent spans 19 years and passes through 5 companies; a new generation was cleared every 11 months on median.
- 42 of the 135 ancestors have a recall of any class; 6 have a Class I recall.
- 6 other devices also cite its predicate K061995.
How does its category compare?
All LLZ devicesLLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · LLZ
58in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LLZ
173in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LLZ
11in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LLZ
116in 2025
MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 76 | 36 | 20 | 152 |
| 2008 | 94 | 52 | 32 | 154 |
| 2009 | 70 | 62 | 21 | 145 |
| 2010 | 51 | 98 | 17 | 145 |
| 2011 | 106 | 79 | 30 | 454 |
| 2012 | 82 | 87 | 12 | 266 |
| 2013 | 95 | 84 | 64 | 1,130 |
| 2014 | 78 | 89 | 6 | 15,631 |
| 2015 | 64 | 90 | 16 | 211 |
| 2016 | 82 | 90 | 12 | 290 |
| 2017 | 78 | 107 | 5 | 104 |
| 2018 | 104 | 105 | 7 | 72 |
| 2019 | 74 | 107 | 2 | 88 |
| 2020 | 81 | 115 | 6 | 82 |
| 2021 | 83 | 116 | 3 | 70 |
| 2022 | 65 | 137 | 4 | 28 |
| 2023 | 87 | 157 | 8 | 26 |
| 2024 | 62 | 131 | 6 | 42 |
| 2025 | 58 | 173 | 11 | 116 |
| 2026 | 45 | 110 | 14 | — |
What is it indicated for?
Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provides integrated tools that allow users to manually assess and score cardiac wall motion and export images and/or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NMV, CR, MG, XA, PET, RT, and X-Ray modalities for reference viewing. It supports connectivity to ultrasound systems, PACS and other DICOM storage repositories.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K140808 · printed Nov 6, 2026
Cite this page
510(k) Report. “K140808: Q-Station.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140808