510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Q-Station

K140808510(k) TraditionalCleared Apr 17, 201416-day reviewClass II21 CFR 892.2050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Ultrasound, Inc. · LLZ System, Image Processing, Radiological

Cleared on Apr 17, 2014 as substantially equivalent to K061995 (Philips Medical Systems North America Co., 2006) and 1 other predicate after a 16-day review. Its family tree reaches back 16 generations to 1995. 6 of the 136 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times.

Decided
Apr 17, 2014
Review
16 days
Class
Class II
Product code
LLZ
Recalls
2
Predicates
2

Its longest line of descent runs back 16 generations to K935595 (1995). 135 ancestors in all, the oldest K882383 (1988). No later device cites it as a predicate yet. 6 of the 136 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2006: 3 other ancestors2012: 1 other ancestor2010: 1 other ancestor2004: 13 other ancestors, 1 with a Class I recall12003: 8 other ancestors2002: 15 other ancestors2000: 13 other ancestors, 1 with a Class I recall11999: 11 other ancestors1997: 7 other ancestors1995: 6 other ancestors2005: 6 other ancestors2001: 11 other ancestors1998: 10 other ancestors, 3 with a Class I recall31996: 5 other ancestors, 1 with a Class I recall11994: 5 other ancestors1993: 1 other ancestor1988: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestorReference device: K132165 · Qlab Quantification Software · Philips Ultrasound, Inc. · 2013 · recalledReference device: K103815 · Q-Station · Philips Ultrasound, Inc. · 2011K935595 · Acuson Model 3001 · Acuson Corp. · 1995K973767 · Harmonic Imaging with Contrast · Acuson Corp. · 1997K022567 · Sequoia Signature II Diagnostic Ultrasound System · Acuson Corp. · 2002K032114 · Sequoia 8.0 Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2003K041319 · Sequoia Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2004K042770 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK050240 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2005 · recalledK052894 · Sonoline G60 S Ultrasound System · Siemens Medical Solutions USA, Inc. · 2005 · recalled×2: 2 generations cleared close together, K063138 (2006), K063803 (2007) K072786 · Acuson S2000 Ultrasound System · Siemens Medical Solutions USA, Inc. · 2007K091970 · Aixplorer Ultrasound System · Supersonic Imagine · 2009K102041 · Aixplorer · Supersonic Imagine · 2010K112255 · Aixplorer · Supersonic Imagine · 2012×2: 2 generations cleared close together, K130499 (2013), K132304 (2013) 1985199019952000200520102015
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 6 other ancestors3 generations back: 12 other ancestors4 generations back: 45 other ancestors, 6 with a Class I recall65 generations back: 12 other ancestors6 generations back: 10 other ancestors7 generations back: 7 other ancestors8 generations back: 8 other ancestors9 generations back: 9 other ancestors10 generations back: 9 other ancestors K132304 · Epiq Diagnostic Ultrasound System · Philips Ultrasound, Inc. · 2013 · recalledK130499 · IU22 Diagnostic Ultrasound System · Philips Ultrasound, Inc. · 2013 · recalledK112255 · Aixplorer · Supersonic Imagine · 2012K102041 · Aixplorer · Supersonic Imagine · 2010K091970 · Aixplorer Ultrasound System · Supersonic Imagine · 2009K072786 · Acuson S2000 Ultrasound System · Siemens Medical Solutions USA, Inc. · 2007K063803 · Acuson Antares Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · recalledK063138 · Modification to Acuson Antares Ultrasound System · Siemens Medical Solutions USA, Inc. · 2006K052894 · Sonoline G60 S Ultrasound System · Siemens Medical Solutions USA, Inc. · 2005 · recalledK050240 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2005 · recalledK042770 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK041319 · Sequoia Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2004K032114 · Sequoia 8.0 Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2003K022567 · Sequoia Signature II Diagnostic Ultrasound System · Acuson Corp. · 2002K973767 · Harmonic Imaging with Contrast · Acuson Corp. · 1997K935595 · Acuson Model 3001 · Acuson Corp. · 1995−15−10−50
  1. Root · 16 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
  2. 12 generations between
  3. 3 generations backK112255 AixplorerSupersonic Imagine · 2012
  4. 2 generations backK130499 IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
  5. 1 generation backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
  6. This deviceK140808 Q-StationPhilips Ultrasound, Inc. · 2014
Show all 16 generations
  1. Root · 16 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
  2. 15 generations backK973767 Harmonic Imaging with ContrastAcuson Corp. · 1997
  3. 14 generations backK022567 Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp. · 2002
  4. 13 generations backK032114 Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2003
  5. 12 generations backK041319 Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004
  6. 11 generations backK042770 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  7. 10 generations backK050240 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
  8. 9 generations backK052894 Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
  9. 8 generations backK063138 Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2006
  10. 7 generations backK063803 Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · Recalled
  11. 6 generations backK072786 Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2007
  12. 5 generations backK091970 Aixplorer Ultrasound SystemSupersonic Imagine · 2009
  13. 4 generations backK102041 AixplorerSupersonic Imagine · 2010
  14. 3 generations backK112255 AixplorerSupersonic Imagine · 2012
  15. 2 generations backK130499 IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
  16. 1 generation backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
  17. This deviceK140808 Q-StationPhilips Ultrasound, Inc. · 2014

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−16K935595Acuson Model 3001Acuson Corp.1995—
−15K973767Harmonic Imaging with ContrastAcuson Corp.1997—
−14K022567Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp.2002—
−13K032114Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2003—
−12K041319Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2004—
−11K042770Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20041
−10K050240Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20051
−9K052894Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc.20051
−8K063138Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc.2006—
−7K063803Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI20079
−6K072786Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc.2007—
−5K091970Aixplorer Ultrasound SystemSupersonic Imagine2009—
−4K102041AixplorerSupersonic Imagine2010—
−3K112255AixplorerSupersonic Imagine2012—
−2K130499IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc.20132
−1K132304Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc.201336
0 (this device)K140808Q-StationPhilips Ultrasound, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199210
199310
199450
199560
199651
199770
1998103
1999110
2000131
2001110
2002150
200380
2004131
200560
200630
201010
201210
FIG. 1 — Lineage of K140808, 1995–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061995Xcelera
Summary page 1
…lips Ultrasound believes that modifications introduced with Q-Station 3.0 are substantially equivalent to other commercially available devices including Philips Xcelera (K061995) and Philips EPIQ (K132304). The Q-Station analysis packages and multi-modality viewing are the same as those found on Philips ul…
Philips Medical Systems North America Co.2006—
*K132304Epiq Diagnostic Ultrasound System
Summary page 1
…at modifications introduced with Q-Station 3.0 are substantially equivalent to other commercially available devices including Philips Xcelera (K061995) and Philips EPIQ (K132304). The Q-Station analysis packages and multi-modality viewing are the same as those found on Philips ultrasound systems, including…
Philips Ultrasound, Inc.201336
†K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc.20131
†K103815Q-StationPhilips Ultrasound, Inc.2011—

Also cleared against K061995: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 15, 2015Class IIZ-1807-2015Software designPhilips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients.
Nov 5, 2014Class IIZ-0818-2015Software designWhen using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 4 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 5 companies; a new generation was cleared every 11 months on median.
  • 42 of the 135 ancestors have a recall of any class; 6 have a Class I recall.
  • 6 other devices also cite its predicate K061995.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provides integrated tools that allow users to manually assess and score cardiac wall motion and export images and/or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NMV, CR, MG, XA, PET, RT, and X-Ray modalities for reference viewing. It supports connectivity to ultrasound systems, PACS and other DICOM storage repositories.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140808: Q-Station.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140808

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026