Hospira Mednet Medication Management Suite

K141102510(k) SpecialCleared Jul 8, 201470-day reviewClass II21 CFR 880.5725Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · FRN Pump, Infusion

Cleared on Jul 8, 2014 as substantially equivalent to K042609 (Hospira, Inc., 2004) after a 70-day review. Its family tree reaches back 4 generations to 1990. 4 of the 15 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 2 times.

Decided
Jul 8, 2014
Review
70 days
Class
Class II
Product code
FRN
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K904518 (1990). 11 ancestors in all. 3 later devices descend from it, 1 citing it directly. 4 of the 15 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor, 1 with a Class I recall12003: 2 other ancestors2002: 2 other ancestors2001: 1 other ancestor, 1 with a Class I recall11994: 1 other ancestor2023: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK042609 · Hospira Mednet Medication Management Suite · Hospira, Inc. · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors, 1 with a Class I recall13 generations back: 2 other ancestors4 generations back: 3 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K042609 · Hospira Mednet Medication Management Suite · Hospira, Inc. · 2004 · recalledK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0
  1. Root · 4 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 3 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 2 generations backK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Recalled (Class I)
  4. 1 generation backK042609 Hospira Mednet Medication Management SuiteHospira, Inc. · 2004 · Recalled
  5. This deviceK141102 Hospira Mednet Medication Management SuiteHospira, Inc. · 2014
  6. Descendant · cites it directly · 2 of its ownK223606 LifeShield™ Infusion Safety Software SuiteIcu Medical, Inc. · 2023 · Recalled
  7. Descendant · 2 generations on · 1 of its ownK242117 LifeShield Infusion Safety Software SuiteIcu Medical, Inc. · 2025 · Recalled
  8. Descendant · 3 generations on · 0 of its ownK252130 LifeShield Infusion Safety Software SuiteIcu Medical, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K223606LifeShield™ Infusion Safety Software SuiteIcu Medical, Inc. · 2023 · cites it directly2descendants
  • K242117LifeShield Infusion Safety Software SuiteIcu Medical, Inc. · 2025 · 2 generations on1descendant
  • K252130LifeShield Infusion Safety Software SuiteIcu Medical, Inc. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−3K994375Horizon LightB.Braun Medical, Inc.2000Class I
−2K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001Class I
−1K042609Hospira Mednet Medication Management SuiteHospira, Inc.20041
0 (this device)K141102Hospira Mednet Medication Management SuiteHospira, Inc.2014—
+1K223606LifeShield™ Infusion Safety Software SuiteIcu Medical, Inc.20231
+2K242117LifeShield Infusion Safety Software SuiteIcu Medical, Inc.20252
+3K252130LifeShield Infusion Safety Software SuiteIcu Medical, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
200111
200220
200320
200411

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202510
202610
FIG. 1 — Lineage of K141102, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042609Hospira Mednet Medication Management Suite
Summary page 1
…725 Product Code(s) FRN Legally marketed Hospira MedNetm" Medication Management Suite (MMS), cleared device(s) to which under K042609. equivalence Is claimed Reason for 510(k) Software Changes submission Device description The Hospira M…
Hospira, Inc.20041

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K223606LifeShield™ Infusion Safety Software SuiteIcu Medical, Inc.20232

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 30, 2017Class IIZ-0195-2018Under Investigation by firmIssue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b)…
Jul 28, 2016Class IIZ-2790-2016Software designHospira MedNet 6.1 software, in combination with the SapphirePlus 13.1x infusion pump, may result in incorrect bolus amount calculations for drugs whose concentrations are listed in Million Units.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · 2 generations back · ◆ in the 2004 ancestors bar of FIG. 14 Class I recalls
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 9.4 years on median.
  • 8 of the 11 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K223606 (Icu Medical, Inc., 2023) has gone furthest, with 2 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira Infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters.

The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physician's orders.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141102: Hospira Mednet Medication Management Suite.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141102

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026