Horizon Outlook with Dosecom

K011975510(k) TraditionalCleared Sep 19, 200186-day reviewClass II21 CFR 880.5725Recalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FRN Pump, Infusion

Cleared on Sep 19, 2001 as substantially equivalent to K994375 (B.Braun Medical, Inc., 2000) after an 86-day review. Its family tree reaches back 2 generations to 1990. 14 of the 53 devices in its family, counting itself, have a Class I recall. 49 later devices descend from it. It has been recalled 4 times, including a Class I recall.

Decided
Sep 19, 2001
Review
86 days
Class
Class II
Product code
FRN
Recalls
4 (1 Class I)
Predicates
1

Its longest line of descent runs back 2 generations to K904518 (1990). 3 ancestors in all. 49 later devices descend from it, 5 citing it directly. 14 of the 53 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1994: 1 other ancestor2004: 3 descendants cleared, 2 with a Class I recall22009: 3 descendants cleared, 1 with a Class I recall12011: 1 descendant cleared2005: 1 descendant cleared2013: 2 descendants cleared, 1 with a Class I recall12014: 5 descendants cleared, 1 with a Class I recall12015: 2 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared, 1 with a Class I recall12007: 1 descendant cleared2025: 5 descendants cleared2022: 6 descendants cleared2020: 2 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared2016: 3 descendants cleared2023: 4 descendants cleared, 2 with a Class I recall22024: 3 descendants cleared, 1 with a Class I recall12019: 2 descendants cleared, 1 with a Class I recall12021: 1 descendant cleared2026: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 5 descendants, 3 with a Class I recall32 generations forward: 9 descendants, 3 with a Class I recall33 generations forward: 13 descendants, 3 with a Class I recall34 generations forward: 11 descendants, 2 with a Class I recall25 generations forward: 5 descendants6 generations forward: 4 descendants, 1 with a Class I recall17 generations forward: 2 descendants K994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0+5
  1. Root · 2 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 1 generation backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. This deviceK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001
  4. Descendant · cites it directly · 28 of its ownK042121 Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Recalled (Class I)
  5. Descendant · 2 generations on · 18 of its ownK133801 Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Recalled (Class I)
  6. Descendant · 3 generations on · 9 of its ownK173084 Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · Recalled (Class I)
  7. Descendant · 2 generations on · 7 of its ownK123049 Sapphire Infusion PumpQ Core Medical, Ltd. · 2013
  8. Descendant · cites it directly · 7 of its ownK041241 Medley System with Bar Code Module, Model 8600Alaris Medical Systems, Inc. · 2004 · Recalled (Class I)

Showing 5 of 49 descendants. See all descendants

Descendants cleared per year

2004: 32005: 12007: 12009: 32011: 12013: 22014: 52015: 22016: 32017: 22018: 22019: 22020: 22021: 12022: 62023: 42024: 32025: 52026: 1
2004high 6 (2022) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 49 · See all
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · cites it directly28descendants
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · 2 generations on18descendants
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · 3 generations on9descendants
  • K123049Sapphire Infusion PumpQ Core Medical, Ltd. · 2013 · 2 generations on7descendants
  • K041241Medley System with Bar Code Module, Model 8600Alaris Medical Systems, Inc. · 2004 · cites it directly7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−1K994375Horizon LightB.Braun Medical, Inc.2000Class I
0 (this device)K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001—
+1K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl.2004Class I
+2K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation2014Class I
+3K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation2018Class I
+2K123049Sapphire Infusion PumpQ Core Medical, Ltd.2013—
+1K041241Medley System with Bar Code Module, Model 8600Alaris Medical Systems, Inc.2004Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200432
200510
200710
200931
201110
201321
201451
201521
201630
201720
201821
201921
202021
202110
202260
202342
202431
202550
202610
FIG. 1 — Lineage of K011975, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

14 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K994375Horizon Light
Summary page 1
…ce that B. Braun Medical Inc. 1s claiming substantial equivalence' to is the Horizon Outlook™ with DoseScan™, (formerly named Horizon® Lite) marketed under cleared 510(k)994375 by B. Braun Medical Inc. Horizon Outlook™ with DoseScan™ is an electrical, external, volumetric infusion pump. There are no new i…
B.Braun Medical, Inc.2000Class I

Also cleared against K994375: 2 other devices →

What descends from it? (49)

See all 49

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042121Spectrum and Spectrum with Master Drug Library Class ISigma Intl.200428
K041241Medley System with Bar Code Module, Model 8600 Class IAlaris Medical Systems, Inc.20047
K083689Infusomat Space Volumetric Infusion Pump System Class IB.Braun Medical, Inc.20094
K042609Hospira Mednet Medication Management SuiteHospira, Inc.20044
K092313Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc.20091

Has anything in its family been recalled? (14 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 14, 2012Class IZ-2188-2013Software designThis notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in which the Outlook ES Safety Infusion System may halt infusion but the "RUN" Light Emitting Diodes…
Sep 7, 2011Not classifiedZ-0051-2012Software designB Braun Outlook ES Safety Infusion System "Run" Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the "Run" LEDs advance as if the pumps were infusing.
Aug 24, 2010Not classifiedZ-0837-2011Component design/selectionPossible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal…
Jul 20, 2010Not classifiedZ-0007-2012Software designThe Outlook ES Dose Guard Care Area can inadvertently be exited when the pump enters the KVO (Keep Vein Open) state. The same condition can occur when a sequence of HOLD-HOLD key strokes are performed on the pump key panel.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 13 Class I recalls
  • K041241Medley System with Bar Code Module, Model 8600Alaris Medical Systems, Inc. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 15 Class I recalls
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K192860Sapphire Infusion Pump, Sapphire administration setsQ Core Medical, Ltd. · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K230022Baxter SIGMA Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · 3 generations on · ◆ in the 2018 descendants bar of FIG. 14 Class I recalls
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls
  • K211122Novum IQ Large Volume PumpBaxter Healthcare Corporation · 2024 · 4 generations on · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls

Nearest 12 of 13 shown.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 9.4 years on median.
  • 2 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K042121 (Sigma Intl., 2004) has gone furthest, with 28 descendants of its own.
  • 2 other devices also cite its predicate K994375.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Horizon Outlook™ with DoseCom™ is an electrical, external volumetric infusion pump intended to provide infusions of parenteral fluids. Parenteral fluids may include all standard fluids and/or medications indicated for infusion as well as blood and blood products. Adjustable occlusion pressure settings allow arterial, epidural and intrathecal infusions. The system created by using dedicated cassettes is intended to provide accurate and continuous flow…

Read the full statement (164 words)

The Horizon Outlook™ with DoseCom™ is an electrical, external volumetric infusion pump intended to provide infusions of parenteral fluids. Parenteral fluids may include all standard fluids and/or medications indicated for infusion as well as blood and blood products. Adjustable occlusion pressure settings allow arterial, epidural and intrathecal infusions. The system created by using dedicated cassettes is intended to provide accurate and continuous flow of these fluids to the patient.

The Horizon Outlook™ with DoseCom™ is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician’s order.

The Horizon Outlook™ with DoseCom™ is intended for but not limited to use in the hospital, and/or nursing home (extended care) settings. The Horizon Outlook™ with DoseCom™ is intended for use by trained healthcare providers in accordance with the instructions provided in the Operation Manual.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K011975: Horizon Outlook with Dosecom.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K011975

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.