510(k) Report

510(k) · Cleared

Recalled · Class I · 2012

Horizon Light

K994375510(k) TraditionalCleared Mar 8, 200072-day reviewClass II21 CFR 880.5725Recalls 5 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FRN Pump, Infusion

Cleared on Mar 8, 2000 as substantially equivalent to K904518 (Kendall Mcgaw Laboratories, Inc., 1990) and 1 other predicate after a 72-day review. Its family tree reaches back 1 generation to 1990. 17 of the 68 devices in its family, counting itself, have a Class I recall. 65 later devices descend from it. It has been recalled 5 times, including a Class I recall.

Decided
Mar 8, 2000
Review
72 days
Class
Class II
Product code
FRN
Recalls
5 (1 Class I)
Predicates
2

Its longest line of descent runs back 1 generation to K904518 (1990). 2 ancestors in all. 65 later devices descend from it, 3 citing it directly. 17 of the 68 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1994: 1 other ancestor2001: 1 descendant cleared, 1 with a Class I recall12003: 1 descendant cleared2002: 1 descendant cleared2004: 3 descendants cleared, 2 with a Class I recall22009: 3 descendants cleared, 1 with a Class I recall12006: 2 descendants cleared, 1 with a Class I recall12011: 1 descendant cleared2005: 1 descendant cleared2013: 2 descendants cleared, 1 with a Class I recall12014: 7 descendants cleared, 2 with a Class I recall22012: 1 descendant cleared2018: 2 descendants cleared, 1 with a Class I recall12020: 5 descendants cleared, 2 with a Class I recall22015: 2 descendants cleared, 1 with a Class I recall12007: 1 descendant cleared2025: 6 descendants cleared2022: 7 descendants cleared2017: 3 descendants cleared2016: 4 descendants cleared2023: 4 descendants cleared, 2 with a Class I recall22021: 2 descendants cleared2024: 3 descendants cleared, 1 with a Class I recall12019: 2 descendants cleared, 1 with a Class I recall12026: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 7 descendants, 4 with a Class I recall43 generations forward: 13 descendants, 5 with a Class I recall54 generations forward: 16 descendants, 3 with a Class I recall35 generations forward: 15 descendants, 2 with a Class I recall26 generations forward: 5 descendants7 generations forward: 4 descendants, 1 with a Class I recall18 generations forward: 2 descendants K904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0+5
  1. Root · 1 generation backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. This deviceK994375 Horizon LightB.Braun Medical, Inc. · 2000
  3. Descendant · cites it directly · 49 of its ownK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Recalled (Class I)
  4. Descendant · 2 generations on · 28 of its ownK042121 Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Recalled (Class I)
  5. Descendant · 3 generations on · 18 of its ownK133801 Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Recalled (Class I)
  6. Descendant · cites it directly · 16 of its ownK023189 Vista Basic with FM SystemB.Braun Medical, Inc. · 2003
  7. Descendant · 2 generations on · 15 of its ownK062700 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Recalled (Class I)

Showing 5 of 65 descendants. See all descendants

Descendants cleared per year

2001: 12002: 12003: 12004: 32005: 12006: 22007: 12009: 32011: 12012: 12013: 22014: 72015: 22016: 42017: 32018: 22019: 22020: 52021: 22022: 72023: 42024: 32025: 62026: 1
2001high 7 (2014) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 65 · See all
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · cites it directly49descendants
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · 2 generations on28descendants
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · 3 generations on18descendants
  • K023189Vista Basic with FM SystemB.Braun Medical, Inc. · 2003 · cites it directly16descendants
  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
0 (this device)K994375Horizon LightB.Braun Medical, Inc.2000—
+1K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001Class I
+2K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl.2004Class I
+3K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation2014Class I
+1K023189Vista Basic with FM SystemB.Braun Medical, Inc.2003—
+2K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2006Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200111
200210
200310
200432
200510
200621
200710
200931
201110
201210
201321
201472
201521
201640
201730
201821
201921
202052
202120
202270
202342
202431
202560
202610
FIG. 1 — Lineage of K994375, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

17 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K904518Volumetric Infusion Pump
Summary page 1
…B. Braun Medical Inc. is claiming substantial equivalence' to is the Horizon® Modular Infusion System, currently marketed by B. Braun Medical Inc. under cleared 510(k) K904518. The Barcode Reader that B. Braun Medical Inc. is claiming substantial equivalence to is part of the Verifuse® Multi-Mode Ambul…
Kendall Mcgaw Laboratories, Inc.199037
*K934551Verifuse Ambulatory Pump
Summary page 1
…. Braun Medical Inc. is claiming substantial equivalence to is part of the Verifuse® Multi-Mode Ambulatory Infusion Pump, marketed by Block Medical under cleared 510(k) K934551. ! The term "substantially equivalent" as used herein is intended to be a determination of substantial equivalence from an FDA -…
Block Medical, Inc.1994—

Also cleared against K904518: 1 other device →

What descends from it? (65)

See all 65

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K011975Horizon Outlook with Dosecom Class IB.Braun Medical, Inc.200149
K023189Vista Basic with FM SystemB.Braun Medical, Inc.200316
K022575Perfusor Compact with FM SystemB.Braun Medical, Inc.20025

Has anything in its family been recalled? (17 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 5, 2012Class IIZ-2175-2013Component design/selectionB. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC…
Nov 14, 2012Class IZ-2188-2013Software designThis notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in which the Outlook ES Safety Infusion System may halt infusion but the "RUN" Light Emitting Diodes…
Sep 7, 2011Not classifiedZ-0051-2012Software designB Braun Outlook ES Safety Infusion System "Run" Light Emitting Diodes (LEDs) pumps stop infusing and a backup alarm sounds, but the "Run" LEDs advance as if the pumps were infusing.
Aug 24, 2010Not classifiedZ-0837-2011Component design/selectionPossible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal…
Jul 20, 2010Not classifiedZ-0007-2012Software designThe Outlook ES Dose Guard Care Area can inadvertently be exited when the pump enters the KVO (Keep Vein Open) state. The same condition can occur when a sequence of HOLD-HOLD key strokes are performed on the pump key panel.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Cites it directly · ◆ in the 2001 descendants bar of FIG. 11 Class I recall
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 13 Class I recalls
  • K041241Medley System with Bar Code Module, Model 8600Alaris Medical Systems, Inc. · 2004 · 2 generations on · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · 2 generations on · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 15 Class I recalls
  • K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
  • K142036Infusomat Space Volumetric Infusion Pump Administration SetsB.Braun Medical, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K192860Sapphire Infusion Pump, Sapphire administration setsQ Core Medical, Ltd. · 2020 · 4 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K230022Baxter SIGMA Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 16 shown.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K011975 (B.Braun Medical, Inc., 2001) has gone furthest, with 49 descendants of its own.
  • 1 other device also cites its predicate K904518.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Horizon® Lite is an electrical, external volumetric infusion pump that provides infusions of parenteral fluids. The pump may be used to infuse all intravenous medications, blood, and blood components. Adjustable occlusion pressure settings allow arterial, epidural and intrathecal infusions. The system created by using dedicated cassettes is intended to provide accurate and continuous flow of these fluids to the patient. The pump is software…

Read the full statement (181 words)

The Horizon® Lite is an electrical, external volumetric infusion pump that provides infusions of parenteral fluids. The pump may be used to infuse all intravenous medications, blood, and blood components. Adjustable occlusion pressure settings allow arterial, epidural and intrathecal infusions. The system created by using dedicated cassettes is intended to provide accurate and continuous flow of these fluids to the patient. The pump is software controlled and operates using volumetric displacement with a stepper-motor mechanism.

The new incorporation of a Barcode Reading System is intended to decrease the incidence of medication errors. The Barcode Reader provides a method to automate the implementation of an infusion by verifying nurse and patient identifications and transferring infusion parameters to the pump.

The Horizon® Lite is intended for but not limited to use in the hospital, home care and/or nursing home (extended care) settings. The Horizon® Lite is intended for use by trained healthcare providers in accordance with the instructions provided in the Operation Manual. All data entry and validation of the Horizon® Lite is performed by the trained healthcare provider per a physician’s order.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K994375: Horizon Light.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K994375

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026