510(k) Report

510(k) · Cleared

Cleared

Kimguard One-Step Sterilization Wrap

K141612510(k) SpecialCleared Aug 7, 201452-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Kimberly-Clark Health Care · FRG Wrap, Sterilization

Cleared on Aug 7, 2014 as substantially equivalent to K113806 (Kimberly-Clark Health Care, 2013) after a 52-day review. Its family tree reaches back 4 generations to 1988. 7 later devices descend from it.

Decided
Aug 7, 2014
Review
52 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K881471 (1988). 7 later devices descend from it, 1 citing it directly.

  1. Root · 4 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 3 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 2 generations backK092167 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 1 generation backK113806 Kimguard One Step Sterilization WrapKimberly-Clark Health Care · 2013
  5. This deviceK141612 Kimguard One-Step Sterilization WrapKimberly-Clark Health Care · 2014
  6. Descendant · cites it directly · 6 of its ownK160755 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016
  7. Descendant · 2 generations on · 2 of its ownK171533 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2017
  8. Descendant · 2 generations on · 1 of its ownK182656 JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd. · 2019
  9. Descendant · 3 generations on · 0 of its ownK240438 CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ®…Comercializadora Brever · 2024
  10. Descendant · 3 generations on · 0 of its ownK230137 Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC · 2023

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2016: 12017: 12019: 12022: 12023: 22024: 1
2016high 2 (2023) · low 0 (2018)2026

Most influential descendants

Showing 5 of 7 · See all
  • K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016 · cites it directly6descendants
  • K171533Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2017 · 2 generations on2descendants
  • K182656JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd. · 2019 · 2 generations on1descendant
  • K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap (44 gsm)Comercializadora Brever · 2024 · 3 generations on0descendants
  • K230137Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC · 2023 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−3K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−2K092167Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp.2009—
−1K113806Kimguard One Step Sterilization WrapKimberly-Clark Health Care2013—
0 (this device)K141612Kimguard One-Step Sterilization WrapKimberly-Clark Health Care2014—
+1K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2016—
+2K171533Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2017—
+2K182656JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd.2019—
+3K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap…Comercializadora Brever2024—
+3K230137Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201710
201910
202210
202320
202410
FIG. 1 — Lineage of K141612, 1988–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K113806Kimguard One Step Sterilization Wrap
Summary page 7
…ICES KIMGUARD® ONE-STEP® Sterilization Wrap, which is currently manufactured and distributed by Kimberly-Clark Corporation [510(k) Notification K113806, cleared August 20, 2013] Special 510(k): Device Modification…
Kimberly-Clark Health Care2013—

Also cleared against K113806: 2 other devices →

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC20166

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 3.8 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K160755 (Ahlstrom Nonwovens, LLC, 2016) has gone furthest, with 6 descendants of its own.
  • 2 other devices also cite its predicate K113806.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

KIMGUARD* ONE-STEP* Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using: Advanced Sterilization Products STERRAD® Sterilization Systems that include: STERRAD® 50, 100S, and 200 STERRAD® NX® [Standard Cycle, Advanced Cycle] STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle] KIMGUARD* ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed…

Read the full statement (173 words)

KIMGUARD* ONE-STEP* Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:

Advanced Sterilization Products STERRAD® Sterilization Systems that include:

STERRAD® 50, 100S, and 200

STERRAD® NX® [Standard Cycle, Advanced Cycle]

STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]

KIMGUARD* ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.

Test results validated that KIMGUARD* ONE-STEP* Sterilization Wraps (KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® [Standard Cycle and Advanced Cycle] and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]).

The KIMGUARD* ONE-STEP* Sterilization Wrap (Models KC300, KC400, K500, KC600) have been validated for use with the STERRAD® 50, STERRAD® 100S, STERRAD® 200, STERRAD® NX® and STERRAD® 100NX® cycles in Table 1.

KIMGUARD* ONE-STEP* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in Table 2.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141612: Kimguard One-Step Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141612

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026