510(k) Report

510(k) · Cleared

Cleared

Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap

K171533510(k) TraditionalCleared Jun 29, 201735-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ahlstrom Nonwovens, LLC · FRG Wrap, Sterilization

Cleared on Jun 29, 2017 as substantially equivalent to K160755 (Ahlstrom Nonwovens, LLC, 2016) after a 35-day review. Its family tree reaches back 6 generations to 1988. 2 later devices descend from it.

Decided
Jun 29, 2017
Review
35 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K881471 (1988). 10 ancestors in all. 2 later devices descend from it, 2 citing it directly.

  1. Root · 6 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 5 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 4 generations backK092167 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 3 generations backK113806 Kimguard One Step Sterilization WrapKimberly-Clark Health Care · 2013
  5. 2 generations backK141612 Kimguard One-Step Sterilization WrapKimberly-Clark Health Care · 2014
  6. 1 generation backK160755 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016
  7. This deviceK171533 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2017
  8. Descendant · cites it directly · 0 of its ownK240438 CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ®…Comercializadora Brever · 2024
  9. Descendant · cites it directly · 0 of its ownK230137 Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC · 2023

Most influential descendants

Showing 2 of 2 · See all
  • K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap (44 gsm)Comercializadora Brever · 2024 · cites it directly0descendants
  • K230137Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−5K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−4K092167Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp.2009—
−3K113806Kimguard One Step Sterilization WrapKimberly-Clark Health Care2013—
−2K141612Kimguard One-Step Sterilization WrapKimberly-Clark Health Care2014—
−1K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2016—
0 (this device)K171533Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2017—
+1K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap…Comercializadora Brever2024—
+1K230137Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200920
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
202410
FIG. 1 — Lineage of K171533, 1988–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap
Summary page 6
…Regulation Number: 880.6850 III. PREDICATE DEVICE The Ahlstrom Nonwovens LLC Reliance® Tandem and Reliance® Solo Sterilization Wrap, cleared under K160755, is the predicate device for this submission. IV. DEVICE DESCRIPTION The Reliance® Solo and Reliance® Tandem Steriliz…
Ahlstrom Nonwovens, LLC2016—

Also cleared against K160755: 2 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K240438CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap (44 gsm)Comercializadora Brever20240
K230137Reliance® Fusion Sterilization WrapsAhlstrom-Munksjo Nonwovens, LLC20230

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 3 companies; a new generation was cleared every 2.0 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • Of its descendants, K240438 (Comercializadora Brever, 2024) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K160755.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

Ahlstrom Reliance® Tandem and Solo Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following: Pre-vacuum Steam 270°F/132°C for 4 minutes Gravity Steam 250°F/121°C for 30 minutes 100% Ethylene Oxide (EO) with a concentration of 725-735 mg/L @ 131°F/55°C and 40% - 80% relative humidity for 60 minutes Advanced Sterilization Products (ASP) STERRAD® 100S Advanced…

Read the full statement (668 words)

Ahlstrom Reliance® Tandem and Solo Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:

Pre-vacuum Steam 270°F/132°C for 4 minutes

Gravity Steam 250°F/121°C for 30 minutes

100% Ethylene Oxide (EO) with a concentration of 725-735 mg/L @ 131°F/55°C and 40% - 80% relative humidity for 60 minutes

Advanced Sterilization Products (ASP) STERRAD® 100S

Advanced Sterilization Products (ASP) STERRAD® 100NX (Standard, Express and Flex cycles)

STERIS Amsco® V-PRO 1, STERIS Amsco® V-PRO 1 Plus, STERIS Amsco® V-PRO maX Low Temperature Sterilization Systems (Lumen, Non Lumen and Flexible Cycles)

The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used.

Pre-Vacuum Steam Sterilization:

Validated for dry time of 20 minutes for Models T100/S100, T200/S200, T300/S300.

Validated for dry time of 30 minutes for Models T400/S400, T500/S500, T600/S600.

All models of Reliance® Tandem and Solo validated for pre-vacuum steam sterilization with stainless steel lumens as part of the load with a maximum weight of 25lbs dependent on the model.

Gravity Steam Sterilization:

Validated for dry times of 20 minutes for Models T100/S100, T200/S200, T300/S300.

Validated for dry times of 30 minutes for Models T400/S400.

Models T100/S100, T200/S200, T300/S300, T400/S400 of Reliance® Tandem and Solo validated for gravity steam sterilization with stainless steel lumens as part of the load with a maximum weight of 25lbs dependent on the model.

Models T500/S500 and T600/S600 are not validated for use for gravity steam sterilization.

Ethylene Oxide Sterilization:

Validated for aeration time of 8 hours at 55oC

All models of Reliance® Tandem and Solo validated for ethylene oxide sterilization with stainless steel lumens as part of the load with a maximum weight of 25lbs dependent on the model.

Advanced Sterilization Products (ASP) STERRAD® 100S Sterilization

All models of Reliance® Tandem and Solo validated with ASP STERRAD® 100S sterilization with stainless steel lumens as part of the load with a maximum weight of 10.7lbs dependent on the model.

Advanced Sterilization Products (ASP) STERRAD® 100NX Sterilization

All models of Reliance® Tandem and Solo validated with ASP STERRAD® 100NX sterilization cycles detailed in Table 1.

Table 1: Validated Advanced Sterilization Products (ASP) STERRAD® 100NX Cycles

Advanced Sterilization Products (ASP) STERRAD® System and Cycle — Intended Load

100NX Standard Cycle: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: An inside diameter of 0.7mm or larger and a length of 500mm or shorter of single-channel stainless steel lumens.

100NX Flex Cycle: One or two single-channel flexible endoscopes with or without a silicone mat. Flexible endoscope may contain: A single channel Teflon/PE/PTFE lumen with an inside diameter of 1 mm or larger and a length of 850mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle)

100NX Express Cycle: Non-lumened reusable metal and non-metal medical devices requiring surface sterilization, or sterilization of mated stainless steel and titanium surfaces, and rigid or semi-rigid endoscopes without lumens.

STERIS Amsco® V-PRO 1, STERIS Amsco® V-PRO 1 Plus, STERIS Amsco® V-PRO maX Low Temperature Sterilization Systems

All models of Reliance® Tandem and Solo Sterilization Wraps have been validated for use with STERIS V-PRO® cycles detailed in Table 2.

Reliance® Tandem and Solo Sterilization Wraps were validated to be effectively aerated during the pre-programmed STERIS Amsco V-PRO® Sterilization Cycles.

Table 2: Validated STERIS Amsco® V-PRO Cycles

Lumen Cycle: Reusable metal and non-metal medical devices, including up to 20 stainless steel lumens with dimensions of 3.0mm diameter or larger and a length of 400mm or shorter.

Non Lumen Cycle: Non lumened reusable metal and non-metal medical devices.

Flexible Cycle: Single lumen surgical flexible endoscopes and bronchoscopes in the following load configuration: 1.) One tray containing a flexible endoscope with a light cord (if not integral to endoscope) and mat. The flexible endoscope having an inside diameter of 1 mm or larger and a length of 850 mm or shorter. 2.) An additional tray containing non-lumened medical devices.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171533: Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171533

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026