510(k) Report

510(k) · Cleared

Cleared

JAMBRO Single Core A Sterilization wrap

K182656510(k) TraditionalCleared Jun 11, 2019259-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jiangsu Zhande Medical Supplies Co., Ltd. · FRG Wrap, Sterilization

Cleared on Jun 11, 2019 as substantially equivalent to K160755 (Ahlstrom Nonwovens, LLC, 2016) after a 259-day review. Its family tree reaches back 6 generations to 1988. 1 later device descends from it.

Decided
Jun 11, 2019
Review
259 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K881471 (1988). 10 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 6 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 5 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 4 generations backK092167 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 3 generations backK113806 Kimguard One Step Sterilization WrapKimberly-Clark Health Care · 2013
  5. 2 generations backK141612 Kimguard One-Step Sterilization WrapKimberly-Clark Health Care · 2014
  6. 1 generation backK160755 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016
  7. This deviceK182656 JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd. · 2019
  8. Descendant · cites it directly · 0 of its ownK220133 Sterilization WrapsWeihai Hongyu Nonwoven Fabric Products Co., Ltd. · 2022

Most influential descendants

Showing 1 of 1 · See all
  • K220133Sterilization WrapsWeihai Hongyu Nonwoven Fabric Products Co., Ltd. · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−5K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−4K092167Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp.2009—
−3K113806Kimguard One Step Sterilization WrapKimberly-Clark Health Care2013—
−2K141612Kimguard One-Step Sterilization WrapKimberly-Clark Health Care2014—
−1K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2016—
0 (this device)K182656JAMBRO Single Core A Sterilization wrapJiangsu Zhande Medical Supplies Co., Ltd.2019—
+1K220133Sterilization WrapsWeihai Hongyu Nonwoven Fabric Products Co., Ltd.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200920
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K182656, 1988–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap
Summary page 5
…II Product Code: FRG III. PREDICATE DEVICE Ahlstrom Nonwovens LLC’s Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap, K160755 IV. DEVICE DESCRIPTION JAMBRO Single Core A is a 63gsm, latex-free, 3-layer (SMS) non-woven sterile wrap, m…
Ahlstrom Nonwovens, LLC2016—

Also cleared against K160755: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220133Sterilization WrapsWeihai Hongyu Nonwoven Fabric Products Co., Ltd.20220

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 4 companies; a new generation was cleared every 2.9 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • Of its descendants, K220133 (Weihai Hongyu Nonwoven Fabric Products Co., Ltd., 2022) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K160755.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

JAMBRO Single Core A is intended to be used to enclose another medical device that is to be sterilized by a health care provider using: • Gravity steam at 250°F/121°C for 30 minutes • Pre-vacuum steam at 270°F/132°C for 4 minutes Gravity steam and Pre-vacuum steam sterilization: Validated for dry times of 30 minutes for JAMBRO Single Core A Types of medical devices to be sterilized in the gravity cycle; • General purpose non-lumened reusable metal and…

Read the full statement (236 words)

JAMBRO Single Core A is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:

• Gravity steam at 250°F/121°C for 30 minutes

• Pre-vacuum steam at 270°F/132°C for 4 minutes

Gravity steam and Pre-vacuum steam sterilization:

Validated for dry times of 30 minutes for JAMBRO Single Core A

Types of medical devices to be sterilized in the gravity cycle;

• General purpose non-lumened reusable metal and nonmetal devices including devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.

Types of medical devices to be sterilized in the pre-vacuum cycle are;

• General purpose reusable metal and nonmetal devices including devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.

• Up to 2 single channel stainless steel lumened devices of the following dimensions; An inside diameter of 3 mm or larger and a length of 400 mm or shorter;

Color of wrap: Blue

Size of wrap: 48 in x 48 in

The maximum validated weight of load for JAMBRO Single Core A is 25 lbs.

The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until used.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182656: JAMBRO Single Core A Sterilization wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182656

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026