510(k) · Cleared
Recalled · Class II · 2016Aquilion Prime, V6.00
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Toshibamedical Systems Corporation · JAK System, X-Ray, Tomography, Computed
Cleared on Nov 26, 2014 as substantially equivalent to K130645 (Toshibamedical Systems Corporation, 2013) after a 152-day review. Its family tree reaches back 2 generations to 2012. 18 later devices descend from it. It has been recalled 2 times.
- Decided
- Nov 26, 2014
- Review
- 152 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 2 generations to K120710 (2012). 18 later devices descend from it, 4 citing it directly.
- Root · 2 generations backK120710 Aquilion PrimeToshibamedical Systems Corporation · 2012 · Recalled
- 1 generation backK130645 Aquilion Prime, V5.00Toshibamedical Systems Corporation · 2013
- This deviceK141741 Aquilion Prime, V6.00Toshibamedical Systems Corporation · 2014
- Descendant · cites it directly · 9 of its ownK170019 Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation · 2017 · Recalled
- Descendant · 2 generations on · 4 of its ownK201836 Aquilion Lightning (TSX-036A/7) V10.2 with AiCE-iCanon Medical Systems Corporation · 2021
- Descendant · cites it directly · 4 of its ownK173468 Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0Toshibamedical Systems Corporation · 2018 · Recalled
- Descendant · 3 generations on · 3 of its ownK222819 Aquilion Serve (TSX-307A/1) V1.2 with AiCE-iCanon Medical Systems Corporation · 2023
- Descendant · 2 generations on · 3 of its ownK182901 Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation · 2019
Showing 5 of 18 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 18 · See all- K170019Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation · 2017 · cites it directly9descendants
- K201836Aquilion Lightning (TSX-036A/7) V10.2 with AiCE-iCanon Medical Systems Corporation · 2021 · 2 generations on4descendants
- K173468Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0Toshibamedical Systems Corporation · 2018 · cites it directly4descendants
- K222819Aquilion Serve (TSX-307A/1) V1.2 with AiCE-iCanon Medical Systems Corporation · 2023 · 3 generations on3descendants
- K182901Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation · 2019 · 2 generations on3descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K120710 | Aquilion Prime | Toshibamedical Systems Corporation | 2012 | 2 |
| −1 | K130645 | Aquilion Prime, V5.00 | Toshibamedical Systems Corporation | 2013 | — |
| 0 (this device) | K141741 | Aquilion Prime, V6.00 | Toshibamedical Systems Corporation | 2014 | — |
| +1 | K170019 | Aquilion Lightning, TSX-036A/1, V8.4 | Toshibamedical Systems Corporation | 2017 | 1 |
| +2 | K201836 | Aquilion Lightning (TSX-036A/7) V10.2 with AiCE-i | Canon Medical Systems Corporation | 2021 | — |
| +1 | K173468 | Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0 | Toshibamedical Systems Corporation | 2018 | 1 |
| +3 | K222819 | Aquilion Serve (TSX-307A/1) V1.2 with AiCE-i | Canon Medical Systems Corporation | 2023 | — |
| +2 | K182901 | Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCE | Canon Medical Systems Corporation | 2019 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2015 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 4 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K130645 | Aquilion Prime, V5.00Summary page 5…510(k) Number Clearance Date Aquilion PRIME, Toshiba America K130645 June 6, 2013 TSX-303A/2 and 303A/6, v5.00 Medical Systems 12. REASON FOR SUBMISSION: Modification… | 2013 |
What descends from it? (18)
See all 184 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 25, 2016 | Class II | Z-0622-2017 | Device Design | Toshiba American Medical Systems (TAMS) is recalling the Aquilion CT system because it has been reported that the cooling oil in the x-ray tube heat exchanger may leak during system operation. |
| May 8, 2015 | Class II | Z-0018-2016 | Radiation Control for Health and Safety Act | If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 2 years and passes through 1 company; a new generation was cleared every 1.5 years on median.
- 1 of the 2 ancestors has a recall of any class.
- Of its descendants, K170019 (Toshibamedical Systems Corporation, 2017) has gone furthest, with 9 descendants of its own.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.
The Aquilion Prime has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K141741 · printed Nov 6, 2026
Cite this page
510(k) Report. “K141741: Aquilion Prime, V6.00.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141741
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026