Ohmeda Medical Omnibed

K993407510(k) TraditionalCleared Jan 5, 200089-day reviewClass II21 CFR 880.5400Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMZ Incubator, Neonatal

Cleared on Jan 5, 2000 after an 89-day review. The FDA summary names its predicate without a number. 6 of the 14 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 4 times.

Decided
Jan 5, 2000
Review
89 days
Class
Class II
Product code
FMZ
Recalls
4
Predicates
0

13 later devices descend from it, 6 citing it directly. 6 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 3 descendants cleared, 1 with a Class I recall12006: 1 descendant cleared2002: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared2016: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12009: 1 descendant cleared, 1 with a Class I recall12011: 1 descendant cleared2022: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK993407 Ohmeda Medical OmnibedOhmeda Medical · 2000
  2. Descendant · cites it directly · 6 of its ownK020543 Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002 · Recalled (Class I)
  3. Descendant · cites it directly · 5 of its ownK071175 Ohmeda Medical Giraffe OmnibedGE Healthcare · 2007
  4. Descendant · cites it directly · 3 of its ownK152814 Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Recalled (Class I)
  5. Descendant · cites it directly · 3 of its ownK070377 Giraffe and Panda WarmerOhmeda Medical · 2007 · Recalled (Class I)
  6. Descendant · 2 generations on · 2 of its ownK213553 Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2002: 12006: 12007: 32009: 12011: 12013: 12016: 12019: 12022: 12025: 12026: 1
2002high 3 (2007) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002 · cites it directly6descendants
  • K071175Ohmeda Medical Giraffe OmnibedGE Healthcare · 2007 · cites it directly5descendants
  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · cites it directly3descendants
  • K070377Giraffe and Panda WarmerOhmeda Medical · 2007 · cites it directly3descendants
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K993407Ohmeda Medical OmnibedOhmeda Medical2000—
+1K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical2002Class I
+1K071175Ohmeda Medical Giraffe OmnibedGE Healthcare2007—
+1K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company2016Class I
+1K070377Giraffe and Panda WarmerOhmeda Medical2007Class I
+2K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc.2022Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200211
200610
200731
200911
201110
201311
201611
201910
202211
202510
202610
FIG. 1 — Lineage of K993407, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Ohmeda - Ohio Care Plus Incubator (all models)
Named without a submission number
—
*—Ohmeda - Ohio Infant Warmer System (all models)
Named without a submission number
—
*—Ohmeda — Ohio Intensive Care Incubator
Named without a submission number
—

What descends from it? (13)

See all 13

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K020543Ohmeda Medical Giraffe Omnibed Class IOhmeda Medical20026
K071175Ohmeda Medical Giraffe OmnibedGE Healthcare20075
K152814Giraffe OmniBed Carestation CS1 Class IOhmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company20163
K070377Giraffe and Panda Warmer Class IOhmeda Medical20073
K190494Infant IncubatorNingbo David Medical Device Co., Ltd.20190
K053399Versalet 7700 Care CenterDraeger Medical Infant Care, Inc.20060

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 28, 2020Class IIZ-2236-2020Device DesignPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Oct 5, 2018Class IIZ-0041-2020Labeling designGE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source…
Nov 18, 2016Class IIZ-1509-2017Under Investigation by firmCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where…
Apr 25, 2005Not classifiedZ-1047-05OtherNeonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K070377Giraffe and Panda WarmerOhmeda Medical · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 12 Class I recalls
  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls
  • K123309Giraffe and Panda WarmersOhmeda Medical · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 13 Class I recalls
  • K090697Modification to: Giraffe and Panda WarmerOhmeda Medical · 2009 · 2 generations on · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K020543 (Ohmeda Medical, 2002) has gone furthest, with 6 descendants of its own.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—

What is it indicated for?

The OmniBed is a combination of an infant incubator (“incubator”) and an infant radiant warmer (“warmer”). Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate’s transition from the uterus to the external environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993407: Ohmeda Medical Omnibed.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K993407

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026