Ohmeda Medical Giraffe Omnibed

K020543510(k) TraditionalCleared May 9, 200279-day reviewClass II21 CFR 880.5400Recalls 10 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMZ Incubator, Neonatal

Cleared on May 9, 2002 as substantially equivalent to K993407 (Ohmeda Medical, 2000) and 2 other predicates after a 79-day review. Its family tree reaches back 1 generation to 1997. 3 of the 10 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 10 times, including 2 Class I recalls.

Decided
May 9, 2002
Review
79 days
Class
Class II
Product code
FMZ
Recalls
10 (2 Class I)
Predicates
3

Its longest line of descent runs back 1 generation to K954204 (1997). 3 ancestors in all. 6 later devices descend from it, 2 citing it directly. 3 of the 10 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors2007: 1 descendant cleared2016: 1 descendant cleared, 1 with a Class I recall12011: 1 descendant cleared2022: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2026: 1 descendant clearedK954204 · Drager Incubator 8000 IC · Drager, Inc. · 19971995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 1 descendant K954204 · Drager Incubator 8000 IC · Drager, Inc. · 19970
  1. Root · 1 generation backK954204 Drager Incubator 8000 ICDrager, Inc. · 1997
  2. This deviceK020543 Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002
  3. Descendant · cites it directly · 5 of its ownK071175 Ohmeda Medical Giraffe OmnibedGE Healthcare · 2007
  4. Descendant · cites it directly · 3 of its ownK152814 Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Recalled (Class I)
  5. Descendant · 2 generations on · 2 of its ownK213553 Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)
  6. Descendant · 3 generations on · 1 of its ownK251663 Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025
  7. Descendant · 4 generations on · 0 of its ownK260326 Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation…Datex-Ohmeda, Inc. · 2026

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2007: 12011: 12016: 12022: 12025: 12026: 1
2007high 1 (2007) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K071175Ohmeda Medical Giraffe OmnibedGE Healthcare · 2007 · cites it directly5descendants
  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · cites it directly3descendants
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on2descendants
  • K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025 · 3 generations on1descendant
  • K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K954204Drager Incubator 8000 ICDrager, Inc.1997—
0 (this device)K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical2002—
+1K071175Ohmeda Medical Giraffe OmnibedGE Healthcare2007—
+1K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company2016Class I
+2K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc.2022Class I
+3K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.2025—
+4K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201110
201611
202211
202510
202610
FIG. 1 — Lineage of K020543, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993407Ohmeda Medical Omnibed
Summary page 1
…vice Information The Giraffe OmniBed is substantially equivalent to the following Class II devices: Device Last 510(k) Number Ohmeda Medical Giraffe OmniBed (original) K993407 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe Om…
Ohmeda Medical20004
*K954204Drager Incubator 8000 IC
Summary page 1
…iBed is substantially equivalent to the following Class II devices: Device Last 510(k) Number Ohmeda Medical Giraffe OmniBed (original) K993407 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe OmniBed is an infant bed which prov…
Drager, Inc.1997—
*K001242Isolette Infant Incubator, Model C2HS
Summary page 1
…wing Class II devices: Device Last 510(k) Number Ohmeda Medical Giraffe OmniBed (original) K993407 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe OmniBed is an infant bed which provides thermal support for infants who are unab…
Hill-Rom Air-Shields2000—

Also cleared against K993407: 5 other devices →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071175Ohmeda Medical Giraffe OmnibedGE Healthcare20075
K152814Giraffe OmniBed Carestation CS1 Class IOhmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company20163

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2026Class IIZ-2475-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
May 1, 2026Class IIZ-2477-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
Sep 3, 2024Class IZ-0047-2025Use errorGE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors…
May 19, 2023Class IIZ-2101-2023Nonconforming Material/ComponentGE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation…
Apr 28, 2020Class IIZ-2235-2020Device DesignPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Oct 11, 2019Class IZ-0538-2020Device DesignGiraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Nov 18, 2016Class IIZ-1504-2017Under Investigation by firmCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where…
Apr 25, 2005Not classifiedZ-1047-05OtherNeonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.
Oct 17, 2003Not classifiedZ-0288-04OtherHeater vent door on neonatal incubator may malfunction.
Jun 13, 2003Not classifiedZ-1064-03OtherFirm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K071175 (GE Healthcare, 2007) has gone furthest, with 5 descendants of its own.
  • 5 other devices also cite its predicate K993407.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—

What is it indicated for?

The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user’s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperatures controlled environment and…

Read the full statement (138 words)

The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user’s demand. It cannot be operated in both modes at the same time.

Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperatures controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate’s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020543: Ohmeda Medical Giraffe Omnibed.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K020543

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026