510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Infinix INFX-8000V, V6.40

K170909510(k) TraditionalCleared Jun 16, 201780-day reviewClass II21 CFR 892.1650Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba Medical Systems Coroporation · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jun 16, 2017 as substantially equivalent to K162614 (Toshibamedical Systems Corporation, 2016) after an 80-day review. Its family tree reaches back 14 generations to 1989. It has been recalled 5 times.

Decided
Jun 16, 2017
Review
80 days
Class
Class II
Product code
OWB
Recalls
5
Predicates
1

Its longest line of descent runs back 14 generations to K890348 (1989). 31 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2001: 3 other ancestors2005: 1 other ancestor2003: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor1998: 3 other ancestors2002: 2 other ancestors1995: 3 other ancestors1994: 1 other ancestor1993: 1 other ancestorK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK031637 · Innova 3100 · GE Medical Systems · 2003 · recalledK050489 · Innova 2100 · Ge Medical Systems, LLC · 2005 · recalledK052412 · Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ · GE Healthcare · 2005 · recalledK060259 · Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121 · Ge Medical Systems, LLC · 2006 · recalledK081582 · Infx-8000V, Infinix-Cfi and Infinix Vfi · Toshiba America Medical Systems, In.C · 2008 · recalledK101868 · Infinix-I Interventional Angiography System Model Infx-8000V · Toshiba America Medical Systems, In.C · 2010K120073 · Infx-8000V with · Toshibamedical Systems Corporation · 2012 · recalledK133535 · Infx-8000V, V5.30 · Toshibamedical Systems Corporation · 2014 · recalledK143225 · INFX-8000V with Wireless Footswitch · Toshibamedical Systems Corporation · 2015 · recalledK152696 · Infx-8000V, V6.20 · Toshibamedical Systems Corporation · 2016 · recalledK162614 · Infinix, INFX-8000V, V6.35 · Toshibamedical Systems Corporation · 20161985199019952000200520102015
Other ancestors, by generation 7 generations back: 1 other ancestor8 generations back: 6 other ancestors9 generations back: 5 other ancestors10 generations back: 5 other ancestors K162614 · Infinix, INFX-8000V, V6.35 · Toshibamedical Systems Corporation · 2016K152696 · Infx-8000V, V6.20 · Toshibamedical Systems Corporation · 2016 · recalledK143225 · INFX-8000V with Wireless Footswitch · Toshibamedical Systems Corporation · 2015 · recalledK133535 · Infx-8000V, V5.30 · Toshibamedical Systems Corporation · 2014 · recalledK120073 · Infx-8000V with · Toshibamedical Systems Corporation · 2012 · recalledK101868 · Infinix-I Interventional Angiography System Model Infx-8000V · Toshiba America Medical Systems, In.C · 2010K081582 · Infx-8000V, Infinix-Cfi and Infinix Vfi · Toshiba America Medical Systems, In.C · 2008 · recalledK060259 · Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121 · Ge Medical Systems, LLC · 2006 · recalledK052412 · Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ · GE Healthcare · 2005 · recalledK050489 · Innova 2100 · Ge Medical Systems, LLC · 2005 · recalledK031637 · Innova 3100 · GE Medical Systems · 2003 · recalledK022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled−10−50
  1. Root · 14 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 10 generations between
  3. 3 generations backK143225 INFX-8000V with Wireless FootswitchToshibamedical Systems Corporation · 2015 · Recalled
  4. 2 generations backK152696 Infx-8000V, V6.20Toshibamedical Systems Corporation · 2016 · Recalled
  5. 1 generation backK162614 Infinix, INFX-8000V, V6.35Toshibamedical Systems Corporation · 2016
  6. This deviceK170909 Infinix INFX-8000V, V6.40Toshiba Medical Systems Coroporation · 2017
Show all 14 generations
  1. Root · 14 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 13 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 12 generations backK022322 Innova and Innova 2000 SGe Medical Systems, Inc. · 2002 · Recalled
  4. 11 generations backK031637 Innova 3100GE Medical Systems · 2003 · Recalled
  5. 10 generations backK050489 Innova 2100Ge Medical Systems, LLC · 2005 · Recalled
  6. 9 generations backK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
  7. 8 generations backK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
  8. 7 generations backK081582 Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · Recalled
  9. 6 generations backK101868 Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010
  10. 5 generations backK120073 Infx-8000V withToshibamedical Systems Corporation · 2012 · Recalled
  11. 4 generations backK133535 Infx-8000V, V5.30Toshibamedical Systems Corporation · 2014 · Recalled
  12. 3 generations backK143225 INFX-8000V with Wireless FootswitchToshibamedical Systems Corporation · 2015 · Recalled
  13. 2 generations backK152696 Infx-8000V, V6.20Toshibamedical Systems Corporation · 2016 · Recalled
  14. 1 generation backK162614 Infinix, INFX-8000V, V6.35Toshibamedical Systems Corporation · 2016
  15. This deviceK170909 Infinix INFX-8000V, V6.40Toshiba Medical Systems Coroporation · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−14K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−13K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−12K022322Innova and Innova 2000 SGe Medical Systems, Inc.20027
−11K031637Innova 3100GE Medical Systems200325
−10K050489Innova 2100Ge Medical Systems, LLC200527
−9K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200550
−8K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC20064
−7K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C20081
−6K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C2010—
−5K120073Infx-8000V withToshibamedical Systems Corporation20123
−4K133535Infx-8000V, V5.30Toshibamedical Systems Corporation20142
−3K143225INFX-8000V with Wireless FootswitchToshibamedical Systems Corporation20152
−2K152696Infx-8000V, V6.20Toshibamedical Systems Corporation20161
−1K162614Infinix, INFX-8000V, V6.35Toshibamedical Systems Corporation2016—
0 (this device)K170909Infinix INFX-8000V, V6.40Toshiba Medical Systems Coroporation2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199410
199530
199830
199910
200010
200130
200220
200310
200510
FIG. 1 — Lineage of K170909, 1989–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K162614Infinix, INFX-8000V, V6.35
Summary page 5
…e Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard]. 15. PREDICATE DEVICE INFX-8000V, V6.35 (K162614) Product Marketed by 510(k) Number Clearance Date INFX-8000V, V6.35 To…
Toshibamedical Systems Corporation2016—

Also cleared against K162614: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 31, 2022Class IIZ-1666-2022Software designThe fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.
Feb 5, 2021Class IIZ-1174-2021Software designThe x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.
Dec 8, 2020Class IIZ-1269-2021Component design/selectionSystem table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Mar 3, 2020Class IIZ-2325-2020Process controlBolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Dec 29, 2017Class IIZ-0844-2018Process controlDuring an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 8 companies; a new generation was cleared every 1.4 years on median.
  • 21 of the 31 ancestors have a recall of any class.
  • 1 other device also cites its predicate K162614.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

This device is a digital radiography/fluoroscopy system used in a diagnostic interventional angiography configuration.

The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen

and lower extremities.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K170909: Infinix INFX-8000V, V6.40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K170909

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026