510(k) Report

510(k) · Cleared

Cleared

Sterilization Wrap

K223600510(k) TraditionalCleared Aug 25, 2023266-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Wuhan Zonsen Medical Products Co., Ltd. · FRG Wrap, Sterilization

Cleared on Aug 25, 2023 as substantially equivalent to K160755 (Ahlstrom Nonwovens, LLC, 2016) after a 266-day review. Its family tree reaches back 6 generations to 1988.

Decided
Aug 25, 2023
Review
266 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K881471 (1988). 10 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 5 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 4 generations backK092167 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 3 generations backK113806 Kimguard One Step Sterilization WrapKimberly-Clark Health Care · 2013
  5. 2 generations backK141612 Kimguard One-Step Sterilization WrapKimberly-Clark Health Care · 2014
  6. 1 generation backK160755 Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC · 2016
  7. This deviceK223600 Sterilization WrapWuhan Zonsen Medical Products Co., Ltd. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−5K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−4K092167Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp.2009—
−3K113806Kimguard One Step Sterilization WrapKimberly-Clark Health Care2013—
−2K141612Kimguard One-Step Sterilization WrapKimberly-Clark Health Care2014—
−1K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization WrapAhlstrom Nonwovens, LLC2016—
0 (this device)K223600Sterilization WrapWuhan Zonsen Medical Products Co., Ltd.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200920
201210
201410
FIG. 1 — Lineage of K223600, 1988–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160755Reliance Solo Sterilization Wrap, Reliance Tandem Sterilization Wrap
Summary page 4
…cess Device Class: Class II Classification Panel: General Hospital Product Code: FRG Regulation Number: 21 CFR 880.6850 4. Pr edicate Devices Primary Predicate Device: K160755 Product Name: Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap Submitter: Ahlstrom Nonwovens LLC…
Ahlstrom Nonwovens, LLC2016—

Also cleared against K160755: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 35 years and passes through 4 companies; a new generation was cleared every 3.8 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K160755.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following: A) Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes dry time. B) Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360-363 mg/L @ 54°C and 40% relative humidity for 480 minutes, 24 hours aeration time at 42 ºC Model: ZSW1610, ZSW 1611. Material: 45gms SMS, 54gms SMS. Color of wrap…

Read the full statement (228 words)

Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider

via the following:

A) Pre-vacuum Steam: 134°C for 4 minutes, 30 minutes dry time.

B) Ethylene Oxide Sterilization: 100% Ethylene Oxide (EO) with a concentration of 360-363 mg/L @ 54°C and 40%

relative humidity for 480 minutes, 24 hours aeration time at 42 ºC

Model: ZSW1610, ZSW 1611.

Material: 45gms SMS, 54gms SMS.

Color of wrap: Pink/Blue (double layers).

Size (inches): 12*12, 15*15, 18*18, 20*20, 24*24, 30*30, 36*36, 40*40, 45*45, 48*48, 54*54, 72*54.

The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain

sterility of the enclosed device(s) up to 90 days post sterilization.

Types of medical devices to be sterilized in the Pre-Vacuum Steam Sterilization and Ethylene Oxide Sterilization:

General purpose non-lumened reusable metal and nonmetal devices including devices with stainless steel diffusionrestricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments etc.

Models validated for both sterilization methods with stainless steel as part of the load, which with a maximum weight of

each types, range from 2lbs to 9lbs, which dependent on each model's size.

For size2*12, 15*15, 18*18, maximum weight is 2lb, for 20*20, 24*24, 30*30, 36*36, 40*40, 45*45, 48*48, maximum

weight is 6lb, and for 54*54, 72*54, maximum weight is 9lb.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K223600: Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K223600

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026