510(k) Report

510(k) · Cleared

Cleared

Amsco Evolution Medium Steam Sterilizer

K173481510(k) TraditionalCleared Mar 9, 2018116-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · FLE Sterilizer, Steam

Cleared on Mar 9, 2018 as substantially equivalent to K091136 (STERIS Corporation, 2009) after a 116-day review. Its family tree reaches back 4 generations to 2001. 1 later device descends from it.

Decided
Mar 9, 2018
Review
116 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K010865 (2001). 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK010865 Amsco Century Medium Sterilizer, Model CenturySTERIS Corporation · 2001 · Recalled
  2. 3 generations backK020747 Amsco Century Medium Sterilizer 26 X 26STERIS Corporation · 2002
  3. 2 generations backK082435 Amsco Evolution Medium Steam SterilizerSTERIS Corporation · 2008 · Recalled
  4. 1 generation backK091136 Amsco Evolution Medium Steam Sterilizer, Model HC-900 and HC-1200STERIS Corporation · 2009 · Recalled
  5. This deviceK173481 Amsco Evolution Medium Steam SterilizerSTERIS Corporation · 2018
  6. Descendant · cites it directly · 0 of its ownK203061 Amsco Evolution Medium Steam Sterilizer (Models HC-900 and…STERIS Corporation · 2020

Most influential descendants

Showing 1 of 1 · See all
  • K203061Amsco Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200),Evolution Automated…STERIS Corporation · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K010865Amsco Century Medium Sterilizer, Model CenturySTERIS Corporation20011
−3K020747Amsco Century Medium Sterilizer 26 X 26STERIS Corporation2002—
−2K082435Amsco Evolution Medium Steam SterilizerSTERIS Corporation20084
−1K091136Amsco Evolution Medium Steam Sterilizer, Model HC-900 and HC-1200STERIS Corporation20093
0 (this device)K173481Amsco Evolution Medium Steam SterilizerSTERIS Corporation2018—
+1K203061Amsco Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200),Evolution…STERIS Corporation2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K173481, 2001–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091136Amsco Evolution Medium Steam Sterilizer, Model HC-900 and HC-1200
Summary page 8
…880.6880 Product Code: FLE 2. Predicate Device AMSCO Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200), K091136. 3. Description of Device The Amsco Evolution Medium Steam Sterilizer models HC-900 and HC-1200 are de…
STERIS Corporation20093

Also cleared against K091136: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203061Amsco Evolution Medium Steam Sterilizer (Models HC-900 and HC-1200),Evolution Automated…STERIS Corporation20200

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 1 company; a new generation was cleared every 6.6 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K203061 (STERIS Corporation, 2020) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K091136.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Amsco Evolution Medium Steam Sterilizer Models HC-900 and HC-1200 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations: * Prevacuum — is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. » Steam Flush Pressure-Pulse (SFPP) — is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles. The Amsco Evolution…

Read the full statement (728 words)

The Amsco Evolution Medium Steam Sterilizer Models HC-900 and HC-1200 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

* Prevacuum — is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

» Steam Flush Pressure-Pulse (SFPP) — is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

The Amsco Evolution Medium Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-1):

Table 5-1. Amsco Evolution Medium Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values

Cycles Sterilize Sterilize Dry Time Recommended Load

Temperature Time

Prevac 270°F (132°C) | 4 minutes 30 minutes | Double wrapped instrument

trays, maximum weight 25

lbs (11.3 kg) each and

Fabric Packs. Refer to

Table 5-3 for recommended

quantities.

Prevac 270°F (132°C) | 4 minutes 5 minutes Single Fabric Pack

Gravity 250°F (121°C) | 30 minutes | 15 minutes | Fabric Packs. Refer to

Table 5-3 for recommended

quantities.

Liquid® 250°F (121°C) | 45 minutes | N/A Refer to Table 5-4 for

recommended quantities.

Prevac 275°F (135°C) | 3 minutes 30 minutes | Double wrapped instrument

trays, maximum weight 25

lbs (11.3 kg) each. Refer to

Table 5-3 for recommended

quantities.

Gravity 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument

trays, maximum weight 25

lbs (11.3 kg) each. Refer to

Table 5-3 for recommended

quantities.

Gravity 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument

trays, maximum weight 25

lbs (11.3 kg) each. Refer to

Table 5-3 for recommended

quantities.

Gravity 270°F (132°C) | 25 minutes | 15 minutes | Fabric Packs. Refer to

Table 5-3 for recommended

quantities.

DART 270°F (132°C) | 3 minutes 1 minute N/A

Warm-Up

DART 270°F (132°C) | 3 ½ minutes | 1 minute Bowie-Dick Test Pack,

DART Test Pack

Leak Test | N/A N/A N/A N/A

* The liquid cycle is for non-patient contact use only.

The Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 5-2):

Table 5-2. Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer factory-programmed sterilization cycles and cycle

Cycles Sterilize Sterlize Dry Time Recommended Load

Temperature Time

SFPP 270°F (132°C) | 4 minutes 30 minutes | Double wrapped

instrument trays,

maximum weight 25 lbs

(11.3 kg) each and Fabric

Packs. Refer to Table 5-3

for recommended

quantities.

SFPP 275°F (135°C) | 3 minutes 30 minutes | Double wrapped

instrument trays,

maximum weight 25 lbs

(11.3 kg) each. Refer to

Table 5-3 for

recommended quantities.

Prevac 270°F (132°C) | 4 minutes 30 minutes | Double wrapped

instrument trays,

maximum weight 25 lbs

(11.3 kg) each and Fabric

Packs. Refer to Table 5-3

for recommended

quantities.

Gravity 250°F (121°C) | 30 minutes 15 minutes | Fabric Packs. Refer to

Table 5-3 for recommended quantities.

SFPP 270°F (132°C) | 4 minutes 5 minutes | Single Fabric Pack

Prevac 275°F (135°C) | 3 minutes 30 minutes | Double wrapped

instrument trays,

maximum weight 25 lbs

(11.3 kg) each. Refer to

Table 5-3 for recommended quantities.

Gravity 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped

instrument trays,

maximum weight 25 lbs

(11.3 kg) each. Refer to

Table 5-3 for recommended quantities.

Liquid’ 250°F (121°C) | 45 minutes | N/A Refer to Table 5-4 for

recommended quantities.

DART 270°F (132°C) | 3 minutes 1 minute N/A

Warm-Up

DART 270°F (132°C) | 3 ½ minutes | 1 minute Bowie-Dick Test Pack,

DART Test Pack

Leak Test N/A N/A N/A N/A

*The liquid cycle is for non-patient contact use only

The following table lists STERIS’s recommended loads by sterilizer size:

Table 5-3. Amsco Evolution Medium Steam Sterilizer recommended loads per sterilizer size

Sterilizer Size Wrapped Instrument Trays Fabric Packs

26” x 37 ½” x 42” 12 18

26” x 37 ½” x 54” 16 30

The following table is a guideline for liquid cycle processing:

Table 5-4. Amsco Evolution Medium Steam Sterilizer Liquid Cycle* Guideline

Number of Containers Volume of Liquid In One Container Minimum Recommended Sterilize Time at 250°F (121°C)

3 1000 ml 45 minutes

*The liquid cycle is for non-patient contact use only.

The Amsco Evolution Medium Steam Sterilizer is offered in the following chamber sizes:

• 26” x 37½” x 42” (Model HC — 900)

• 26” x 37½” x 54” (Model HC — 1200)

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173481: Amsco Evolution Medium Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173481

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.