510(k) · Cleared
ClearedAssureTech Panel Dip Tests, AssureTech Quick Cup Tests
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Assure Tech. (Hangzhou) Co, Ltd. · NFT Test, Amphetamine, Over The Counter
Cleared on Aug 20, 2018 as substantially equivalent to K122809 (Advin Biotech, 2013) after a 48-day review. Its family tree reaches back 6 generations to 1995. 12 later devices descend from it.
- Decided
- Aug 20, 2018
- Review
- 48 days
- Class
- Class II
- Product code
- NFT
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K945610 (1995). 27 ancestors in all. 12 later devices descend from it, 2 citing it directly.
- Root · 6 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
- 5 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
- 4 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
- 3 generations backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
- 2 generations backK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006 · Recalled
- 1 generation backK122809 Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013
- This deviceK181768 AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2018
- Descendant · cites it directly · 8 of its ownK201630 AssureTech DOA Dipstick Screen Panel Tests, AssureTech DOA…Assure Tech. (Hangzhou) Co, Ltd. · 2021
- Descendant · 2 generations on · 6 of its ownK232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek® Multi-Drug Urine Test…Guangzhou Decheng Biotechnology Co., Ltd. · 2023
- Descendant · 3 generations on · 3 of its ownK240686 Healgen Accurate Urine Drug Screen Dip Card; Healgen Accurate Home…Healgen Scientific,, LLC · 2024
- Descendant · cites it directly · 2 of its ownK243996 AssureTech Panel Dip Tests; AssureTech Quick Cup Tests; AssureTech…Assure Tech., LLC · 2025
- Descendant · 2 generations on · 1 of its ownK252259 AssureTech Quick Cup Tests; AssureTech Multi-drug Urine Test CupAssure Tech., LLC · 2025
Showing 5 of 12 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 12 · See all- K201630AssureTech DOA Dipstick Screen Panel Tests, AssureTech DOA Integrated Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2021 · cites it directly8descendants
- K232659Dochek® Multi-Drug Urine Test Cup Rx, Dochek® Multi-Drug Urine Test CupGuangzhou Decheng Biotechnology Co., Ltd. · 2023 · 2 generations on6descendants
- K240686Healgen Accurate Urine Drug Screen Dip Card; Healgen Accurate Home Urine Drug Test Dip…Healgen Scientific,, LLC · 2024 · 3 generations on3descendants
- K243996AssureTech Panel Dip Tests; AssureTech Quick Cup Tests; AssureTech Multi-drug Urine Test…Assure Tech., LLC · 2025 · cites it directly2descendants
- K252259AssureTech Quick Cup Tests; AssureTech Multi-drug Urine Test CupAssure Tech., LLC · 2025 · 2 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K945610 | Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp | Princeton BioMeditech Corp. | 1995 | — |
| −5 | K011673 | Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test… | ACON Laboratories, Inc. | 2001 | — |
| −4 | K020313 | Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug… | Acon Laboratories Co. | 2002 | — |
| −3 | K031759 | Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum… | ACON Laboratories, Inc. | 2003 | 1 |
| −2 | K061718 | Innovacon Spectrum II Test Card | Innovacon, Inc. | 2006 | 6 |
| −1 | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and Cup | Advin Biotech | 2013 | — |
| 0 (this device) | K181768 | AssureTech Panel Dip Tests, AssureTech Quick Cup Tests | Assure Tech. (Hangzhou) Co, Ltd. | 2018 | — |
| +1 | K201630 | AssureTech DOA Dipstick Screen Panel Tests, AssureTech DOA Integrated Cup Tests | Assure Tech. (Hangzhou) Co, Ltd. | 2021 | — |
| +2 | K232659 | Dochek® Multi-Drug Urine Test Cup Rx, Dochek® Multi-Drug Urine Test Cup | Guangzhou Decheng Biotechnology Co., Ltd. | 2023 | — |
| +3 | K240686 | Healgen Accurate Urine Drug Screen Dip Card; Healgen Accurate Home Urine Drug… | Healgen Scientific,, LLC | 2024 | — |
| +1 | K243996 | AssureTech Panel Dip Tests; AssureTech Quick Cup Tests; AssureTech Multi-drug… | Assure Tech., LLC | 2025 | — |
| +2 | K252259 | AssureTech Quick Cup Tests; AssureTech Multi-drug Urine Test Cup | Assure Tech., LLC | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 7 | 0 |
| 2002 | 2 | 0 |
| 2003 | 2 | 0 |
| 2004 | 3 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2021 | 1 | 0 |
| 2023 | 2 | 0 |
| 2024 | 3 | 0 |
| 2025 | 5 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NFT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NFT device
- #1 K244043 AllTest Multi-Drug Rapid Test Cup ; AllTest…score 20.3 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and CupSummary page 6…II 21 CFR, 862.3700 Propoxyphene Toxicology (91) Propoxyphene Test System 5. Predicate Devices: K122809 The Advin Multi-Drug Screen Tests 6. Indications for Use AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are co… | 2013 |
What descends from it? (12)
See all 122 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 5.2 years on median.
- 2 of the 27 ancestors have a recall of any class.
- Of its descendants, K201630 (Assure Tech. (Hangzhou) Co, Ltd., 2021) has gone furthest, with 8 descendants of its own.
- 11 other devices also cite its predicate K122809.
How does its category compare?
All NFT devicesNFT Test, Amphetamine, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.
FDA decisions per year · NFT
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NFT
37in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NFT
No recall names NFT since 2007.
Adverse event reports per year · NFT
No MAUDE report names NFT since 2007.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 1 | — | 0 | 0 |
| 2018 | 6 | 56 | 0 | 0 |
| 2019 | 1 | — | 0 | 0 |
| 2020 | 2 | — | 0 | 0 |
| 2021 | 1 | — | 0 | 0 |
| 2022 | 3 | 63 | 0 | 0 |
| 2023 | 1 | — | 0 | 0 |
| 2024 | 4 | 31 | 0 | 0 |
| 2025 | 8 | 37 | 0 | 0 |
| 2026 | 5 | 35 | 0 | — |
What is it indicated for?
AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of: Drug(Identifier) Cut-off level…
Read the full statement (216 words)
AssureTech Panel Dip Tests and AssureTech Quick Cup Tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, EDDP, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:
Drug(Identifier) Cut-off level
Amphetamine 500 ng/mL
Oxazepam 300 ng/mL
Cocaine 150 ng/mL
Marijuana 50 ng/mL
Methamphetamine 500 ng/mL
Morphine 300 ng/mL
Oxycodone 100 ng/mL
Secobarbital 300 ng/mL
Buprenorphine 10 ng/mL
Methylenedioxy-methamphetamine 500 ng/mL
Phencyclidine 25 ng/mL
Methadone 300 ng/mL
EDDP 300 ng/mL
Nortriptyline 1000 ng/mL
d-Propoxyphene 300 ng/mL
Configuration of the AssureTech Panel Dip Tests and the AssureTech Quick Cup Tests can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
For in vitro diagnostic use only.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K181768 · printed Nov 6, 2026
Cite this page
510(k) Report. “K181768: AssureTech Panel Dip Tests, AssureTech Quick Cup Tests.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K181768
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026