510(k) · Cleared
Recalled · Class II · 2022Wondfo T-Cup Multi-Drug Urine Test Cup
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Guangzhou Wondfo Biotech Co., Ltd. · NFT Test, Amphetamine, Over The Counter
Cleared on Dec 3, 2018 as substantially equivalent to K122809 (Advin Biotech, 2013) after a 67-day review. Its family tree reaches back 6 generations to 1995. 13 later devices descend from it. It has been recalled once.
- Decided
- Dec 3, 2018
- Review
- 67 days
- Class
- Class II
- Product code
- NFT
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 6 generations to K945610 (1995). 27 ancestors in all. 13 later devices descend from it, 7 citing it directly.
- Root · 6 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
- 5 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
- 4 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
- 3 generations backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
- 2 generations backK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006 · Recalled
- 1 generation backK122809 Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013
- This deviceK182701 Wondfo T-Cup Multi-Drug Urine Test CupGuangzhou Wondfo Biotech Co., Ltd. · 2018
- Descendant · cites it directly · 4 of its ownK202567 Wondfo T-Dip Multi-Drug Urine Test Panel, Wondfo T-Dip Multi-Drug…Guangzhou Wondfo Biotech Co., Ltd. · 2020
- Descendant · 2 generations on · 1 of its ownK233062 BioSieveTM Multi-Drug Urine Test Panel; BioSieveTM Multi-Drug Urine…Vivachek Biotech (Hangzhou) Co., Ltd. · 2023
- Descendant · cites it directly · 1 of its ownK222667 Wisdiag Multi-Drug Urine Test Cup, Wisdiag Multi-Drug Urine Test Cup…Vivachek Biotech (Hangzhou) Co., Ltd. · 2022
- Descendant · cites it directly · 1 of its ownK191841 Accurate Multi Panel Drug Urine Test CupHealstone Biotech, Inc. · 2019
- Descendant · 3 generations on · 0 of its ownK261280 BioSieve Multi-Drug Urine Test Panel;BioSieve Multi-Drug Urine Home…Vivachek Biotech (Hangzhou) Co., Ltd. · 2026
Showing 5 of 13 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 13 · See all- K202567Wondfo T-Dip Multi-Drug Urine Test Panel, Wondfo T-Dip Multi-Drug Urine Test Panel RxGuangzhou Wondfo Biotech Co., Ltd. · 2020 · cites it directly4descendants
- K233062BioSieveTM Multi-Drug Urine Test Panel; BioSieveTM Multi-Drug Urine Test Panel RxVivachek Biotech (Hangzhou) Co., Ltd. · 2023 · 2 generations on1descendant
- K222667Wisdiag Multi-Drug Urine Test Cup, Wisdiag Multi-Drug Urine Test Cup RxVivachek Biotech (Hangzhou) Co., Ltd. · 2022 · cites it directly1descendant
- K191841Accurate Multi Panel Drug Urine Test CupHealstone Biotech, Inc. · 2019 · cites it directly1descendant
- K261280BioSieve Multi-Drug Urine Test Panel;BioSieve Multi-Drug Urine Home Test PanelVivachek Biotech (Hangzhou) Co., Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K945610 | Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp | Princeton BioMeditech Corp. | 1995 | — |
| −5 | K011673 | Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test… | ACON Laboratories, Inc. | 2001 | — |
| −4 | K020313 | Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug… | Acon Laboratories Co. | 2002 | — |
| −3 | K031759 | Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum… | ACON Laboratories, Inc. | 2003 | 1 |
| −2 | K061718 | Innovacon Spectrum II Test Card | Innovacon, Inc. | 2006 | 6 |
| −1 | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and Cup | Advin Biotech | 2013 | — |
| 0 (this device) | K182701 | Wondfo T-Cup Multi-Drug Urine Test Cup | Guangzhou Wondfo Biotech Co., Ltd. | 2018 | — |
| +1 | K202567 | Wondfo T-Dip Multi-Drug Urine Test Panel, Wondfo T-Dip Multi-Drug Urine Test… | Guangzhou Wondfo Biotech Co., Ltd. | 2020 | — |
| +2 | K233062 | BioSieveTM Multi-Drug Urine Test Panel; BioSieveTM Multi-Drug Urine Test Panel… | Vivachek Biotech (Hangzhou) Co., Ltd. | 2023 | — |
| +1 | K222667 | Wisdiag Multi-Drug Urine Test Cup, Wisdiag Multi-Drug Urine Test Cup Rx | Vivachek Biotech (Hangzhou) Co., Ltd. | 2022 | — |
| +1 | K191841 | Accurate Multi Panel Drug Urine Test Cup | Healstone Biotech, Inc. | 2019 | — |
| +3 | K261280 | BioSieve Multi-Drug Urine Test Panel;BioSieve Multi-Drug Urine Home Test Panel | Vivachek Biotech (Hangzhou) Co., Ltd. | 2026 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 7 | 0 |
| 2002 | 2 | 0 |
| 2003 | 2 | 0 |
| 2004 | 3 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 1 | 0 |
| 2022 | 3 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NFT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NFT device
- #1 K244043 AllTest Multi-Drug Rapid Test Cup ; AllTest…score 20.3 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and CupSummary page 6…NFW 862.3870, Cannabinoids Test Toxicology Cannabinoids (THC 50) System 6 Predicate K122809 Device Advin Biotech Multi-Drug Screen Tests 7. Intended Use Wondfo T-Cup® Multi-Drug Urine Test Cup… | 2013 |
What descends from it? (13)
See all 137 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K202567 | Wondfo T-Dip Multi-Drug Urine Test Panel, Wondfo T-Dip Multi-Drug Urine Test Panel Rx | 2020 | 4 |
| K222667 | Wisdiag Multi-Drug Urine Test Cup, Wisdiag Multi-Drug Urine Test Cup Rx | 2022 | 1 |
| K191841 | Accurate Multi Panel Drug Urine Test Cup | 2019 | 1 |
| K252550 | SAFElife T-Cup Multi-Drug Urine Test Cup; SAFElife T-Cup Multi-Drug Urine Test Cup Dx | 2025 | 0 |
| K242077 | PocguideTM Multi-Drug Test Cup OTC, PocguideTM Multi-Drug Test Cup | 2024 | 0 |
| K220685 | HIGHTOP Multi-Drug Urine Test Cup, HIGHTOP Multi-Drug Urine Test Cup Rx | 2022 | 0 |
| K210327 | First Sign Multi-Drug Test Dip Card, First Sign Multi-Drug Test Cup, First Sign… | 2021 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 25, 2022 | Class II | Z-1589-2022 | Environmental control | Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 5.5 years on median.
- 2 of the 27 ancestors have a recall of any class.
- Of its descendants, K202567 (Guangzhou Wondfo Biotech Co., Ltd., 2020) has gone furthest, with 4 descendants of its own.
- 11 other devices also cite its predicate K122809.
How does its category compare?
All NFT devicesNFT Test, Amphetamine, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.
FDA decisions per year · NFT
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NFT
37in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NFT
No recall names NFT since 2007.
Adverse event reports per year · NFT
No MAUDE report names NFT since 2007.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 1 | — | 0 | 0 |
| 2018 | 6 | 56 | 0 | 0 |
| 2019 | 1 | — | 0 | 0 |
| 2020 | 2 | — | 0 | 0 |
| 2021 | 1 | — | 0 | 0 |
| 2022 | 3 | 63 | 0 | 0 |
| 2023 | 1 | — | 0 | 0 |
| 2024 | 4 | 31 | 0 | 0 |
| 2025 | 8 | 37 | 0 | 0 |
| 2026 | 5 | 35 | 0 | — |
What is it indicated for?
Wondfo T-Cup® Multi-Drug Urine Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in human urine at the cutoff concentrations of…
Read the full statement (251 words)
Wondfo T-Cup® Multi-Drug Urine Test Cup tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Cannabinoids in human urine at the cutoff concentrations of:
Drug (Identifier) Cut-off level
Amphetamine (AMP) 1000 ng/mL or 500 ng/mL
Buprenorphine (BUP) 10 ng/mL
Secobarbital (BAR) 300 ng/mL
Oxazepam (BZO) 300 ng/mL
Cocaine (COC) 300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) 300 ng/mL
Methamphetamine (MET) 1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA) 500 ng/mL
Morphine (MOP 300/OPI 2000) 2000 ng/mL or 300 ng/mL
Methadone (MTD) 300 ng/mL
Oxycodone (OXY) 100 ng/mL
Phencyclidine (PCP) 25 ng/mL
Propoxyphene (PPX) 300 ng/mL
Nortriptyline (TCA) 1000 ng/mL
Cannabinoids (THC 50) 50 ng/mL
Wondfo T-Cup® Multi-Drug Urine Test Cup offers any combinations from 2 to 15 drugs of abuse tests but only one cutoff concentration under same drug condition will be included per device. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene, and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K182701 · printed Nov 6, 2026
Cite this page
510(k) Report. “K182701: Wondfo T-Cup Multi-Drug Urine Test Cup.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182701
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026