510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Comprehensive Shoulder System

K193038510(k) TraditionalCleared Oct 28, 2020363-day reviewClass II21 CFR 888.3670Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet Manufacturing Corp · MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented

Cleared on Oct 28, 2020 as substantially equivalent to K060692 (Biomet Manufacturing Corp, 2006) after a 363-day review. Its family tree reaches back 5 generations to 1980. 8 later devices descend from it. It has been recalled once.

Decided
Oct 28, 2020
Review
363 days
Class
Class II
Product code
MBF
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K800494 (1980). 25 ancestors in all. 8 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1997: 1 other ancestor2001: 1 other ancestor1999: 3 other ancestors1998: 2 other ancestors1996: 1 other ancestor1995: 2 other ancestors1994: 1 other ancestor1992: 2 other ancestors1988: 1 other ancestor1987: 3 other ancestors1991: 1 other ancestor1980: 1 other ancestor2022: 1 descendant cleared2023: 1 descendant cleared2021: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant cleared2026: 2 descendants clearedReference device: K033506 · Comprehensive Humeral Fracture Positioning Sleeve · Biomet, Inc. · 2004Reference device: K040610 · Versa-Dial Humeral Head · Biomet Manufacturing, Inc. · 2004Reference device: K060694 · Modular Hybrid Glenoid · Biomet Manufacturing Corp · 2006 · recalledReference device: K060716 · Versa-Dial Humeral Head Prosthesis · Biomet Manufacturing Corp · 2006Reference device: K140390 · Titanium Versa-Dial Humeral Head Prosthesis · Biomet Manufacturing Corp · 2014Reference device: K140652 · Porous Coated Comprehensive Fracture Stems · Biomet Manufacturing Corp · 2014K800494 · New Jersey Hemiarthroplasty Sholder Pros · Depuy, Inc. · 1980K991585 · Bipolar Shoulder Prosthesis · Biomet, Inc. · 1999K002998 · Biomet BI-Polar Shoulder System · Biomet, Inc. · 2000 · recalledK030710 · Bio-Modular Shoulder System · Biomet, Inc. · 2003 · recalledK060692 · Comprehensive Primary Shoulder Stems · Biomet Manufacturing Corp · 2006 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 15 other ancestors4 generations back: 2 other ancestors5 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K060692 · Comprehensive Primary Shoulder Stems · Biomet Manufacturing Corp · 2006 · recalledK030710 · Bio-Modular Shoulder System · Biomet, Inc. · 2003 · recalledK002998 · Biomet BI-Polar Shoulder System · Biomet, Inc. · 2000 · recalledK991585 · Bipolar Shoulder Prosthesis · Biomet, Inc. · 1999K800494 · New Jersey Hemiarthroplasty Sholder Pros · Depuy, Inc. · 1980−50
  1. Root · 5 generations backK800494 New Jersey Hemiarthroplasty Sholder ProsDepuy, Inc. · 1980
  2. 4 generations backK991585 Bipolar Shoulder ProsthesisBiomet, Inc. · 1999
  3. 3 generations backK002998 Biomet BI-Polar Shoulder SystemBiomet, Inc. · 2000 · Recalled
  4. 2 generations backK030710 Bio-Modular Shoulder SystemBiomet, Inc. · 2003 · Recalled
  5. 1 generation backK060692 Comprehensive Primary Shoulder StemsBiomet Manufacturing Corp · 2006 · Recalled
  6. This deviceK193038 Comprehensive Shoulder SystemBiomet Manufacturing Corp · 2020
  7. Descendant · cites it directly · 4 of its ownK213856 Identity Shoulder SystemZimmer, Inc. · 2022
  8. Descendant · 2 generations on · 1 of its ownK240876 Identity Shoulder SystemZimmer, Inc. · 2024
  9. Descendant · cites it directly · 1 of its ownK214001 Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder SystemBiomet Manufacturing Corp · 2023
  10. Descendant · 2 generations on · 0 of its ownK252404 Comprehensive Reverse Shoulder - HA Glenosphere BaseplatesBiomet Orthopedics · 2026
  11. Descendant · 3 generations on · 0 of its ownK253592 Zimmer Biomet Reverse Shoulder Prosthesis (Alliance Humeral…Zimmer, Inc. · 2026

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2021: 12022: 12023: 12024: 12025: 22026: 2
2021high 2 (2025) · low 1 (2021)2026

Most influential descendants

Showing 5 of 8 · See all
  • K213856Identity Shoulder SystemZimmer, Inc. · 2022 · cites it directly4descendants
  • K240876Identity Shoulder SystemZimmer, Inc. · 2024 · 2 generations on1descendant
  • K214001Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder SystemBiomet Manufacturing Corp · 2023 · cites it directly1descendant
  • K252404Comprehensive Reverse Shoulder - HA Glenosphere BaseplatesBiomet Orthopedics · 2026 · 2 generations on0descendants
  • K253592Zimmer Biomet Reverse Shoulder Prosthesis (Alliance Humeral Bearings, Identity Humeral…Zimmer, Inc. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K800494New Jersey Hemiarthroplasty Sholder ProsDepuy, Inc.1980—
−4K991585Bipolar Shoulder ProsthesisBiomet, Inc.1999—
−3K002998Biomet BI-Polar Shoulder SystemBiomet, Inc.20001
−2K030710Bio-Modular Shoulder SystemBiomet, Inc.20031
−1K060692Comprehensive Primary Shoulder StemsBiomet Manufacturing Corp20066
0 (this device)K193038Comprehensive Shoulder SystemBiomet Manufacturing Corp2020—
+1K213856Identity Shoulder SystemZimmer, Inc.2022—
+2K240876Identity Shoulder SystemZimmer, Inc.2024—
+1K214001Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder SystemBiomet Manufacturing Corp2023—
+2K252404Comprehensive Reverse Shoulder - HA Glenosphere BaseplatesBiomet Orthopedics2026—
+3K253592Zimmer Biomet Reverse Shoulder Prosthesis (Alliance Humeral Bearings, Identity…Zimmer, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198730
198810
199110
199220
199410
199520
199610
199710
199820
199930
200110
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
202310
202410
202520
202620
FIG. 1 — Lineage of K193038, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBF device

  1. #1 K202716 Ignite Anatomic Shoulder Systemscore 20.3 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060692Comprehensive Primary Shoulder Stems
Summary page 4
…Primary Predicate 510(k) Number Comprehensive Primary Shoulder K060692 Reference Predicates 510(k) Number Comprehensive Humeral Fracture…
Biomet Manufacturing Corp20066
†K033506Comprehensive Humeral Fracture Positioning SleeveBiomet, Inc.2004—
†K040610Versa-Dial Humeral HeadBiomet Manufacturing, Inc.2004—
†K060694Modular Hybrid GlenoidBiomet Manufacturing Corp20064
†K060716Versa-Dial Humeral Head ProsthesisBiomet Manufacturing Corp2006—
†K140390Titanium Versa-Dial Humeral Head ProsthesisBiomet Manufacturing Corp2014—
†K140652Porous Coated Comprehensive Fracture StemsBiomet Manufacturing Corp2014—

Also cleared against K060692: 14 other devices →

What descends from it? (8)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213856Identity Shoulder SystemZimmer, Inc.20224
K214001Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder SystemBiomet Manufacturing Corp20231
K251098Identity Revision Humeral StemsZimmer, Inc.20250
K212435Comprehensive Humeral Fracture Positioning SleevesBiomet, Inc.20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 3, 2025Class IIZ-1403-2025Nonconforming Material/ComponentFive complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 40 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
  • 5 of the 25 ancestors have a recall of any class.
  • Of its descendants, K213856 (Zimmer, Inc., 2022) has gone furthest, with 4 descendants of its own.
  • 14 other devices also cite its predicate K060692.

How does its category compare?

All MBF devices

MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MBF

2in 2025

2007: 22008: 12009: 12012: 22013: 12014: 12017: 22018: 12019: 22020: 22021: 12022: 42023: 42024: 22025: 22026 (partial year): 1
2007high 4 (2022) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBF

144in 2023

2022: 164 days median2023: 144 days median
2007high 164 (2022) · low 144 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBF

1in 2023

2009: 12017: 12018: 12019: 22022: 82023: 1
2007high 8 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBF

255in 2025

2007: 22008: 72009: 162010: 242011: 322012: 302013: 402014: 282015: 492016: 852017: 1932018: 992019: 472020: 242021: 362022: 1142023: 1432024: 1642025: 255
2007high 255 (2025) · low 2 (2007)2025

MAUDE reports naming product code MBF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—02
20081—07
20091—116
20100—024
20110—032
20122—030
20131—040
20141—028
20150—049
20160—085
20172—1193
20181—199
20192—247
20202—024
20211—036
202241648114
202341441143
20242—0164
20252—0255
20261—0—

What is it indicated for?

1 Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis. 3. Correction of functional deformity. 4, Fractures of the proximal humerus, where other methods of treatment are deemed inadequate. 5. Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate. Optional use in revision in some medical conditions (e. g0…

Read the full statement (230 words)

1 Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.

2. Rheumatoid arthritis.

3. Correction of functional deformity.

4, Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.

5. Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.

Optional use in revision in some medical conditions (e. g0 revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure

Humeral components with a porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.

The Comprehensive Modular Hybrid Glenoid is intended to be implanted with bone cement. The optional porous titanium peg may be inserted without bone cement. The optional polyethylene peg should be inserted with bone cement.

The Comprehensive Hurneral Positioning Sleeves are for cemented use only and are intended for use with the Comprehensive Fracture Stem.

The Versa-Dial Humeral Head Prosthesis is intended for use only with the Comprehensive Shoulder Stems (Fracture, Primary and Revision), and the glenoid components of the Comprehensive Shoulder System.

The Titanium Versa-Dial Humeral Head Prosthesis is indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack suspected material sensitivity to cobalt alloy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K193038: Comprehensive Shoulder System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K193038

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026