510(k) Report

510(k) · Cleared

Cleared

StarGuide

K210173510(k) TraditionalCleared Mar 29, 202166-day reviewClass II21 CFR 892.1200

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · KPS System, Tomography, Computed, Emission

Cleared on Mar 29, 2021 as substantially equivalent to K153402 (GE Healthcare, 2016) after a 66-day review. Its family tree reaches back 9 generations to 1995. 6 of the 34 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Mar 29, 2021
Review
66 days
Class
Class II
Product code
KPS
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K952674 (1995). 32 ancestors in all, the oldest K892358 (1989). 1 later device descends from it, 1 citing it directly. 6 of the 34 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2005: 3 other ancestors, 1 with a Class I recall12002: 2 other ancestors, 1 with a Class I recall12004: 2 other ancestors1997: 2 other ancestors1995: 2 other ancestors, 1 with a Class I recall12003: 2 other ancestors2001: 2 other ancestors2000: 4 other ancestors1999: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestor2025: 1 descendant clearedReference device: K173816 · NM/CT 850, NM/CT 860 · GE Healthcare · 2018 · Class I recallReference device: K182484 · VERITON CT whole body SPECT/CT system · Spectrum Dynamics Medical, Ltd. · 2018K952674 · Transact: Attenuation Correction System for the Apex Cardial Gamma Camera · Elscint, Inc. · 1995 · recalledK980959 · Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1998K991841 · Hawkeye Option for Dual-Head Variable Angle Gamma Camera · Ge Medical Systems F.I. Haifa · 1999 · Class I recallK023190 · E.Cam Computer / E.Soft Workstation · Siemens Medical Solutions USA, Inc. · 2002K063826 · Syngo MI Applications 2007A · Siemens Medical Solutions USA, Inc. · 2007K072567 · Symbia-E, Model 10275879 · Siemens Medical Solutions USA, Inc. · 2007 · recalledK082506 · Symbia, Version 4.0 · Siemens Medical Solutions USA, Inc. · 2008 · recalledK093514 · Discovery Model NM/CT 670 · Ge Medical Systems F.I. Haifa · 2009 · Class I recallK153402 · Discovery NM/CT 670 CZT · GE Healthcare · 2016 · Class I recall1990200020102020
  1. Root · 9 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 5 generations between
  3. 3 generations backK082506 Symbia, Version 4.0Siemens Medical Solutions USA, Inc. · 2008 · Recalled
  4. 2 generations backK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
  5. 1 generation backK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
  6. This deviceK210173 StarGuideGE Healthcare · 2021
  7. Descendant · cites it directly · 0 of its ownK241350 Clarify DLGe Medical Systems Israel, Functional Imaging · 2025
Show all 9 generations
  1. Root · 9 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
  2. 8 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
  3. 7 generations backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
  4. 6 generations backK023190 E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc. · 2002
  5. 5 generations backK063826 Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc. · 2007
  6. 4 generations backK072567 Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc. · 2007 · Recalled
  7. 3 generations backK082506 Symbia, Version 4.0Siemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 2 generations backK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
  9. 1 generation backK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
  10. This deviceK210173 StarGuideGE Healthcare · 2021

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K241350Clarify DLGe Medical Systems Israel, Functional Imaging · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K952674Transact: Attenuation Correction System for the Apex Cardial Gamma CameraElscint, Inc.19951
−8K980959Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma…Ge Medical Systems F.I. Haifa1998—
−7K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa1999Class I
−6K023190E.Cam Computer / E.Soft WorkstationSiemens Medical Solutions USA, Inc.2002—
−5K063826Syngo MI Applications 2007ASiemens Medical Solutions USA, Inc.2007—
−4K072567Symbia-E, Model 10275879Siemens Medical Solutions USA, Inc.20072
−3K082506Symbia, Version 4.0Siemens Medical Solutions USA, Inc.20089
−2K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa2009Class I
−1K153402Discovery NM/CT 670 CZTGE Healthcare2016Class I
0 (this device)K210173StarGuideGE Healthcare2021—
+1K241350Clarify DLGe Medical Systems Israel, Functional Imaging2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199110
199521
199720
199910
200040
200120
200221
200320
200420
200531
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K210173, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K153402Discovery NM/CT 670 CZT
Summary page 6
…mation Device Name: Discovery NM/CT 670 CZT Manufacturer: GE Medical Systems Israel, Functional Imaging 510(k) Number: K153402 Regulation Number/ 21CFR 892.1200 & 21CFR 892.1750 Product Code: 90 KPS & 90 JAK…
GE Healthcare2016Class I
†K173816NM/CT 850, NM/CT 860GE Healthcare2018Class I
†K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd.2018—

Also cleared against K153402: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K241350Clarify DLGe Medical Systems Israel, Functional Imaging20250

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K052434Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems · 2005 · 3 generations back · ◆ in the 2005 ancestors bar of FIG. 11 Class I recall
  • K022960GE Quasar Nuclear Medicine SystemGe Medical Systems F.I. Haifa · 2002 · 3 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K953801VaricamElscint, Inc. · 1995 · 4 generations back · ◆ in the 1995 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 2.7 years on median.
  • 22 of the 32 ancestors have a recall of any class; 6 have a Class I recall.
  • Of its descendants, K241350 (Ge Medical Systems Israel, Functional Imaging, 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K153402.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The GE StarGuide system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders, orthopedic disorders, and cancer. The system output can also be used by the physician for staging and restaging of tumors…

Read the full statement (341 words)

The GE StarGuide system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders, orthopedic disorders, and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy, including the nuclear medicine part of theragnostic procedures. The GE StarGuide system, combining a CZT-based, high energy and spatial resolution, Nuclear Medicine (NM) system and a Computed Tomography (CT) system, is intended to produce:

• NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types and by imaging features designed to enhance image quality.

• CT System: Cross sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained with or without contrast. The CT system is indicated for head, whole body, cardiac and vascular X-Ray Computed Tomography applications.

• NM + CT System: Combined, hybrid SPECT and CT protocols, for CT-based corrections of SPECT images as well as functional and anatomical mapping imaging (localization, registration and fusion).

The GE StarGuide system includes digital processing of data and images, signal analysis and display equipment, patient and equipment supports, components and accessories. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation. The system may be used for patients of all ages.

FAME: Functional-Anatomical Mutual Enhancement (FAME) technology is an image processing method intended for Computed Tomography (CT) based corrections of Nuclear Medicine (NM) bone scintigraphy images. FAME adjusts the radioisotope tracer distribution to correlate with the skeletal anatomical structures in the CT image. FAME may be used for patients of all ages.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K210173: StarGuide.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K210173

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026