BR-FHUS Viewer 1.0

K171709510(k) TraditionalCleared Oct 20, 2017134-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Taihao Medical, Inc. · LLZ System, Image Processing, Radiological

Cleared on Oct 20, 2017 as substantially equivalent to K123901 (Tractus Corporation, 2013) after a 134-day review. Its family tree reaches back 8 generations to 1991. 6 of the 72 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Oct 20, 2017
Review
134 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K913770 (1991). 70 ancestors in all, the oldest K882383 (1988). 1 later device descends from it, 1 citing it directly. 6 of the 72 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2006: 2 other ancestors2005: 3 other ancestors2004: 2 other ancestors, 1 with a Class I recall12003: 5 other ancestors2002: 7 other ancestors2001: 7 other ancestors2000: 7 other ancestors, 1 with a Class I recall11999: 5 other ancestors1998: 10 other ancestors, 3 with a Class I recall31997: 2 other ancestors1996: 5 other ancestors, 1 with a Class I recall11995: 2 other ancestors1994: 2 other ancestors1993: 1 other ancestor1988: 1 other ancestor2021: 1 descendant clearedK913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK033400 · Seno Advantage · Ge Medical Systems W-400 · 2003 · recalledK043216 · Cadstream Version 4.0 · Confirma, Inc. · 2005K070244 · AEGIS · Sentinelle Medical, Inc. · 2007K093672 · Aegis Navigation Anda Egis Pelvic Application · Sentinelle Medical, Inc. · 2009K123901 · Tractus Tissuemapper Reviewer Application · Tractus Corporation · 20131990200020102020
  1. Root · 8 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 7 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 6 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 5 generations backK033400 Seno AdvantageGe Medical Systems W-400 · 2003 · Recalled
  5. 4 generations backK043216 Cadstream Version 4.0Confirma, Inc. · 2005
  6. 3 generations backK070244 AEGISSentinelle Medical, Inc. · 2007
  7. 2 generations backK093672 Aegis Navigation Anda Egis Pelvic ApplicationSentinelle Medical, Inc. · 2009
  8. 1 generation backK123901 Tractus Tissuemapper Reviewer ApplicationTractus Corporation · 2013
  9. This deviceK171709 BR-FHUS Viewer 1.0Taihao Medical, Inc. · 2017
  10. Descendant · cites it directly · 0 of its ownK211098 Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc. · 2021

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K211098Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−7K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−6K020483Advantage Workstation 4.1GE Medical Systems20022
−5K033400Seno AdvantageGe Medical Systems W-40020036
−4K043216Cadstream Version 4.0Confirma, Inc.2005—
−3K070244AEGISSentinelle Medical, Inc.2007—
−2K093672Aegis Navigation Anda Egis Pelvic ApplicationSentinelle Medical, Inc.2009—
−1K123901Tractus Tissuemapper Reviewer ApplicationTractus Corporation2013—
0 (this device)K171709BR-FHUS Viewer 1.0Taihao Medical, Inc.2017—
+1K211098Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199310
199420
199520
199651
199720
1998103
199950
200071
200170
200270
200350
200421
200530
200620
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K171709, 1991–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K123901Tractus Tissuemapper Reviewer Application
Summary page 4
…Number: 21 CFR 892.2050 Classification Product Code: LLZ Classification: Class II Predicate Device: Tractus TissueMapper Reviewer Application (K123901) 5-1…
Tractus Corporation2013—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211098Breast Volume Navigator (BVN) Model G-2000Metritrack, Inc.20210

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 4 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 7 companies; a new generation was cleared every 3.1 years on median.
  • 23 of the 70 ancestors have a recall of any class; 6 have a Class I recall.
  • Of its descendants, K211098 (Metritrack, Inc., 2021) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

BR-FHUS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. BR-FHUS Viewer 1.0 provides the capability to visualize two-dimensional ultrasound images along with the scanning paths and position information of probe that stored in the DICOM file in advance.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171709: BR-FHUS Viewer 1.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171709

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026