510(k) · Cleared
ClearedSYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
STERIS Corporation · MED Sterilant, Medical Devices
Cleared on May 12, 2021 as substantially equivalent to K192929 (STERIS Corporation, 2019) after a 62-day review. Its family tree reaches back 6 generations to 2016. 7 later devices descend from it.
- Decided
- May 12, 2021
- Review
- 62 days
- Class
- Class II
- Product code
- MED
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K161683 (2016). 8 ancestors in all. 7 later devices descend from it, 1 citing it directly.
- Root · 6 generations backK161683 SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2016
- 5 generations backK170956 SYSTEM 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2017
- 4 generations backK173256 System 1 endo Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2018
- 3 generations backK182827 SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model 6900STERIS Corporation · 2019
- 2 generations backK191343 SYSTEM 1 endo Liquid Chemical Sterilant Processing System,Model P6900STERIS Corporation · 2019
- 1 generation backK192929 SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo…STERIS Corporation · 2019
- This deviceK210737 SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo…STERIS Corporation · 2021
- Descendant · cites it directly · 6 of its ownK211607 SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo…STERIS Corporation · 2021
- Descendant · 2 generations on · 3 of its ownK222615 SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo…STERIS Corporation · 2022
- Descendant · 2 generations on · 1 of its ownK232914 SYSTEM 1E Liquid Chemical Sterilant Processing System (P6500)…STERIS Corporation · 2023
- Descendant · 2 generations on · 1 of its ownK230930 enspire 3000 Cleaning and Liquid Chemical Sterile Processing SystemSTERIS Corporation · 2023
- Descendant · 3 generations on · 0 of its ownK251666 SYSTEM 1E Liquid Chemical Sterilant Processing System, Model P6700…STERIS Corporation · 2026
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 7 · See all- K211607SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical…STERIS Corporation · 2021 · cites it directly6descendants
- K222615SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical…STERIS Corporation · 2022 · 2 generations on3descendants
- K232914SYSTEM 1E Liquid Chemical Sterilant Processing System (P6500); SYSTEM 1 endo Liquid…STERIS Corporation · 2023 · 2 generations on1descendant
- K230930enspire 3000 Cleaning and Liquid Chemical Sterile Processing SystemSTERIS Corporation · 2023 · 2 generations on1descendant
- K251666SYSTEM 1E Liquid Chemical Sterilant Processing System, Model P6700; SYSTEM 1 endo Liquid…STERIS Corporation · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K161683 | SYSTEM 1E Liquid Chemical Sterilant Processing System | STERIS Corporation | 2016 | — |
| −5 | K170956 | SYSTEM 1E Liquid Chemical Sterilant Processing System | STERIS Corporation | 2017 | — |
| −4 | K173256 | System 1 endo Liquid Chemical Sterilant Processing System | STERIS Corporation | 2018 | — |
| −3 | K182827 | SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model 6900 | STERIS Corporation | 2019 | — |
| −2 | K191343 | SYSTEM 1 endo Liquid Chemical Sterilant Processing System,Model P6900 | STERIS Corporation | 2019 | — |
| −1 | K192929 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid… | STERIS Corporation | 2019 | — |
| 0 (this device) | K210737 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid… | STERIS Corporation | 2021 | — |
| +1 | K211607 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid… | STERIS Corporation | 2021 | — |
| +2 | K222615 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid… | STERIS Corporation | 2022 | — |
| +2 | K232914 | SYSTEM 1E Liquid Chemical Sterilant Processing System (P6500); SYSTEM 1 endo… | STERIS Corporation | 2023 | — |
| +2 | K230930 | enspire 3000 Cleaning and Liquid Chemical Sterile Processing System | STERIS Corporation | 2023 | — |
| +3 | K251666 | SYSTEM 1E Liquid Chemical Sterilant Processing System, Model P6700; SYSTEM 1… | STERIS Corporation | 2026 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2021 | 1 | 0 |
| 2022 | 1 | 0 |
| 2023 | 3 | 0 |
| 2024 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MED device
- #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K192929 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical…Summary page 7…sification Number: 21 CFR 880.6885 Product Code: MED 2. Predicate Device SYSTEM 1E Liquid Chemical Sterilant Processing System, K192929 3. Description of Device The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilizat… | 2019 |
What descends from it? (7)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K211607 | SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical… | 2021 | 6 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 11 months on median.
- None of the 8 ancestors has a recall.
- Of its descendants, K211607 (STERIS Corporation, 2021) has gone furthest, with 6 descendants of its own.
How does its category compare?
All MED devicesMED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.
FDA decisions per year · MED
2in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · MED
84in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MED
2in 2011
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MED
9in 2025
MAUDE reports naming product code MED, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 3 | 34 |
| 2008 | 1 | — | 2 | 79 |
| 2009 | 0 | — | 0 | 158 |
| 2010 | 6 | 281 | 0 | 89 |
| 2011 | 0 | — | 2 | 106 |
| 2012 | 3 | 111 | 0 | 337 |
| 2013 | 1 | — | 0 | 140 |
| 2014 | 1 | — | 0 | 105 |
| 2015 | 0 | — | 0 | 110 |
| 2016 | 1 | — | 0 | 135 |
| 2017 | 1 | — | 0 | 74 |
| 2018 | 2 | — | 0 | 42 |
| 2019 | 4 | 84 | 0 | 35 |
| 2020 | 0 | — | 0 | 32 |
| 2021 | 2 | — | 0 | 31 |
| 2022 | 1 | — | 0 | 16 |
| 2023 | 2 | — | 0 | 9 |
| 2024 | 0 | — | 0 | 8 |
| 2025 | 0 | — | 0 | 9 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60° C, and rinses the load with…
Read the full statement (413 words)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60° C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
* The extensive treatment of EPA potable water consists of:
1. Pre-filtration through two pre-filters:
• Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
• Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
2. UV Irradiation:
• During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
3. 0.1 micron filtration:
• The water prepared by pre-filtration and UV irradiation is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical heat-sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K210737 · printed Nov 4, 2026
Cite this page
510(k) Report. “K210737: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K210737
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026