Thornhill Research, Inc. · BTL Ventilator, Emergency, Powered (Resuscitator)
Cleared on Mar 26, 2010 as substantially equivalent to K013931 (Sequal Technologies, Inc., 2002) and 7 other predicates after a 158-day review. Its family tree reaches back 7 generations to 1991. 4 of the 71 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled 2 times, including 2 Class I recalls.
- Decided
- Mar 26, 2010
- Review
- 158 days
- Class
- Class II
- Product code
- BTL
- Recalls
- 2 (2 Class I)
- Predicates
- 8
Its longest line of descent runs back 7 generations to K902859 (1991). 63 ancestors in all, the oldest K852650 (1985). 7 later devices descend from it, 2 citing it directly. 4 of the 71 devices in this family, counting this one, have a Class I recall.
- Root · 7 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
- 6 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
- 5 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
- 4 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
- 3 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
- 2 generations backK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
- 1 generation backK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
- This deviceK093261 MOVESThornhill Research, Inc. · 2010
- Descendant · cites it directly · 4 of its ownK141460 Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2015
- Descendant · 2 generations on · 3 of its ownK160967 Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2017
- Descendant · 2 generations on · 2 of its ownK173916 Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2018
- Descendant · cites it directly · 1 of its ownK140049 Moves SLCThornhill Research, Inc. · 2014
- Descendant · 3 generations on · 0 of its ownK223332 Task Force CORECnsystems Medizintechnik GmbH · 2023
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 7 · See all- K141460Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2015 · cites it directly4descendants
- K160967Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2017 · 2 generations on3descendants
- K173916Finapres Nova Noninvasive Hemodynamic MonitorFinapres Medical Systems B.V. · 2018 · 2 generations on2descendants
- K140049Moves SLCThornhill Research, Inc. · 2014 · cites it directly1descendant
- K223332Task Force CORECnsystems Medizintechnik GmbH · 2023 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K902859 | Servo Ventilator 300 | Siemens Life Support Systems | 1991 | 3 |
| −6 | K960010 | Servo Ventilator 300 Computer Interface | Siemens Elema AB | 1996 | — |
| −5 | K970839 | Servo Ventilator 300A | Siemens Elema AB | 1997 | 4 |
| −4 | K992788 | Bear Cub 750 PSV | Bear Medical Systems, Inc. | 1999 | — |
| −3 | K013642 | AVEA | Bird Products Corp. | 2002 | Class I |
| −2 | K032451 | Bird Tbird Vela Ventilator | Bird Products Corp. | 2004 | 1 |
| −1 | K052554 | Modification to: Ivent 201 Portable Ventilator | Versamed Medical Systems, Inc. | 2005 | — |
| 0 (this device) | K093261 | MOVES | Thornhill Research, Inc. | 2010 | — |
| +1 | K141460 | Finapres Nova Noninvasive Hemodynamic Monitor | Finapres Medical Systems B.V. | 2015 | — |
| +2 | K160967 | Finapres Nova Noninvasive Hemodynamic Monitor | Finapres Medical Systems B.V. | 2017 | — |
| +2 | K173916 | Finapres Nova Noninvasive Hemodynamic Monitor | Finapres Medical Systems B.V. | 2018 | — |
| +1 | K140049 | Moves SLC | Thornhill Research, Inc. | 2014 | — |
| +3 | K223332 | Task Force CORE | Cnsystems Medizintechnik GmbH | 2023 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1985 | 1 | 0 |
| 1988 | 1 | 0 |
| 1990 | 3 | 0 |
| 1991 | 4 | 0 |
| 1992 | 2 | 0 |
| 1993 | 1 | 0 |
| 1994 | 4 | 0 |
| 1995 | 7 | 0 |
| 1996 | 6 | 0 |
| 1997 | 1 | 0 |
| 1998 | 7 | 2 |
| 1999 | 6 | 0 |
| 2000 | 1 | 0 |
| 2001 | 4 | 0 |
| 2002 | 2 | 0 |
| 2003 | 2 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2023 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BTL device
- #1 K242769 VentStar Resus heated (N) (MP17030); VentStar…score 11.2 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (8)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K013931 | Omni Oxygen System, Model 1000Summary page 3…devices: Device 510(k) t Sequal Eclipse 2 K013931 VersaMed iWent 201 K052554 Univent Eagle 73x… | 2002 |
| * | K052554 | Modification to: Ivent 201 Portable VentilatorSummary page 3…510(k) t Sequal Eclipse 2 K013931 VersaMed iWent 201 K052554 Univent Eagle 73x K051476 Umivent Eagle 754… | 2005 |
| * | K051476 | Uni-Vent Model 73XSummary page 3…K013931 VersaMed iWent 201 K052554 Univent Eagle 73x K051476 Umivent Eagle 754 K931473 Hi,-O-x80… | 2005 |
| * | K931473 | Impact Model 753Summary page 3…K052554 Univent Eagle 73x K051476 Umivent Eagle 754 K931473 Hi,-O-x80 K030943 Medela Vario 8 & 18… | 1994 |
| * | K030943 | Modification to HioxSummary page 3…K051476 Umivent Eagle 754 K931473 Hi,-O-x80 K030943 Medela Vario 8 & 18 K06 1435 Criticare Poet Plus Patient Monitor (Model 8… | 2003 |
| * | K061435 | Medela Vario Suction PumpsSummary page 3…K931473 Hi,-O-x80 K030943 Medela Vario 8 & 18 K06 1435 Criticare Poet Plus Patient Monitor (Model 8 100) K012059 Welch Allyn Propag Encore… | 2006 |
| * | K012059 | 8100/8500 Vital Signs MonitorSummary page 3…K030943 Medela Vario 8 & 18 K06 1435 Criticare Poet Plus Patient Monitor (Model 8 100) K012059 Welch Allyn Propag Encore K012451 Comparison to Predicate Devices A Compariso… | 2001 |
| * | K012451 | Propaq Encore Models 202, 204, 206; Propaq CS Models 242, 244, 246Summary page 3…K06 1435 Criticare Poet Plus Patient Monitor (Model 8 100) K012059 Welch Allyn Propag Encore K012451 Comparison to Predicate Devices A Comparison of specifications demonstrates that the MOVES is substantially equivalent to the i… | 2001 |
| † | K081844 | GE Datex-Ohmeda Avance | 2008 |
What descends from it? (7)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 19, 2015 | Class I | Z-0586-2016 | Component design/selection | Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled. |
| Nov 19, 2015 | Class I | Z-0587-2016 | Component design/selection | Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K013642AVEABird Products Corp. · 2002 · 3 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
- K981072Esprit VentilatorRespironics, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 1.9 years on median.
- 13 of the 63 ancestors have a recall of any class; 3 have a Class I recall.
- Of its descendants, K141460 (Finapres Medical Systems B.V., 2015) has gone furthest, with 4 descendants of its own.
- 9 other devices also cite its predicate K013931.
How does its category compare?
All BTL devicesBTL Ventilator, Emergency, Powered (Resuscitator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.
FDA decisions per year · BTL
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · BTL
203in 2014
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BTL
2in 2024
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · BTL
165in 2025
MAUDE reports naming product code BTL, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 5 | 122 | 1 | 5 |
| 2008 | 4 | 198 | 0 | 37 |
| 2009 | 5 | 173 | 0 | 69 |
| 2010 | 3 | 140 | 0 | 96 |
| 2011 | 2 | — | 1 | 75 |
| 2012 | 0 | — | 2 (2) | 115 |
| 2013 | 1 | — | 3 (1) | 103 |
| 2014 | 7 | 203 | 0 | 70 |
| 2015 | 1 | — | 3 (2) | 79 |
| 2016 | 0 | — | 0 | 80 |
| 2017 | 2 | — | 0 | 101 |
| 2018 | 1 | — | 0 | 197 |
| 2019 | 1 | — | 3 (3) | 341 |
| 2020 | 1 | — | 0 | 454 |
| 2021 | 1 | — | 1 | 506 |
| 2022 | 0 | — | 0 | 755 |
| 2023 | 1 | — | 0 | 731 |
| 2024 | 0 | — | 2 (2) | 667 |
| 2025 | 1 | — | 0 | 165 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The MOVESTM is a portable computer controlled electrically powered transport ventilator intended to provide continuous or intermittent emergency ventilatory support for the care of adults who require mechanical ventilation. The MOVESTM is intended to deliver high inspired oxygen concentrations via the MOVESTM 02 mask to spontaneously breathing patients who require elevated inspired oxygen MOVESTM is intended to be used in a transport or emergency setting…
Read the full statement (196 words)
The MOVESTM is a portable computer controlled electrically powered transport ventilator intended to provide continuous or intermittent emergency ventilatory support for the care of adults who require mechanical ventilation.
The MOVESTM is intended to deliver high inspired oxygen concentrations via the MOVESTM 02 mask to spontaneously breathing patients who require elevated inspired oxygen
MOVESTM is intended to be used in a transport or emergency setting on adult patients who weigh between 40 and 120kg.
MOVESTM provides the following supplemental functions for patients that it is ventilating or to whom it is delivering elevated inspired oxygen:
a. Suction
The MOVESTM suction pump is intended for aspiration and removal of fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system.
b. Supplementary Oxygen
The MOVESTM is intended to provide supplemental oxygen enriched air to patients that require supplemental oxygen.
c. Patient Monitoring
The MOVESTm is intended to monitor physiological parameters of patients and provide these parameters to a health care provider for interpretation in the form of physiological data and system alarms. Physiological data and system alarms will be available to the care provider from the monitor.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K093261 · printed Nov 6, 2026
Cite this page
510(k) Report. “K093261: MOVES.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K093261
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026