510(k) · Cleared
Recalled · Class II · 2025Flash Ultrasound System 5100 Point of Care
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Ultrasound, LLC · IYN System, Imaging, Pulsed Doppler, Ultrasonic
Cleared on Apr 28, 2025 as substantially equivalent to K162329 (Philips Health Care, 2016) after a 112-day review. Its family tree reaches back 18 generations to 1995. 6 of the 139 devices in its family, counting itself, have a Class I recall. It has been recalled 5 times.
- Decided
- Apr 28, 2025
- Review
- 112 days
- Class
- Class II
- Product code
- IYN
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 18 generations to K935595 (1995). 138 ancestors in all, the oldest K882383 (1988). No later device cites it as a predicate yet. 6 of the 139 devices in this family, counting this one, have a Class I recall.
- Root · 18 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
- 14 generations between
- 3 generations backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- 2 generations backK160807 EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound…Philips Ultrasound, Inc. · 2016 · Recalled
- 1 generation backK162329 CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound SystemPhilips Health Care · 2016 · Recalled
- This deviceK250030 Flash Ultrasound System 5100 Point of CarePhilips Ultrasound, LLC · 2025
Show all 18 generations
- Root · 18 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
- 17 generations backK973767 Harmonic Imaging with ContrastAcuson Corp. · 1997
- 16 generations backK022567 Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp. · 2002
- 15 generations backK032114 Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2003
- 14 generations backK041319 Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004
- 13 generations backK042770 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 12 generations backK050240 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
- 11 generations backK052894 Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
- 10 generations backK063138 Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2006
- 9 generations backK063803 Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · Recalled
- 8 generations backK072786 Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2007
- 7 generations backK091970 Aixplorer Ultrasound SystemSupersonic Imagine · 2009
- 6 generations backK102041 AixplorerSupersonic Imagine · 2010
- 5 generations backK112255 AixplorerSupersonic Imagine · 2012
- 4 generations backK130499 IU22 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- 3 generations backK132304 Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2013 · Recalled
- 2 generations backK160807 EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound…Philips Ultrasound, Inc. · 2016 · Recalled
- 1 generation backK162329 CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound SystemPhilips Health Care · 2016 · Recalled
- This deviceK250030 Flash Ultrasound System 5100 Point of CarePhilips Ultrasound, LLC · 2025
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −18 | K935595 | Acuson Model 3001 | Acuson Corp. | 1995 | — |
| −17 | K973767 | Harmonic Imaging with Contrast | Acuson Corp. | 1997 | — |
| −16 | K022567 | Sequoia Signature II Diagnostic Ultrasound System | Acuson Corp. | 2002 | — |
| −15 | K032114 | Sequoia 8.0 Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc. | 2003 | — |
| −14 | K041319 | Sequoia Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc. | 2004 | — |
| −13 | K042770 | Modification to Acuson CV70 Cardiovascular System | Siemens Medical Solutions USA, Inc. | 2004 | 1 |
| −12 | K050240 | Modification to Acuson CV70 Cardiovascular System | Siemens Medical Solutions USA, Inc. | 2005 | 1 |
| −11 | K052894 | Sonoline G60 S Ultrasound System | Siemens Medical Solutions USA, Inc. | 2005 | 1 |
| −10 | K063138 | Modification to Acuson Antares Ultrasound System | Siemens Medical Solutions USA, Inc. | 2006 | — |
| −9 | K063803 | Acuson Antares Diagnostic Ultrasound System | Siemens Medical Solutions USA, Inc., Ultrasound DI | 2007 | 9 |
| −8 | K072786 | Acuson S2000 Ultrasound System | Siemens Medical Solutions USA, Inc. | 2007 | — |
| −7 | K091970 | Aixplorer Ultrasound System | Supersonic Imagine | 2009 | — |
| −6 | K102041 | Aixplorer | Supersonic Imagine | 2010 | — |
| −5 | K112255 | Aixplorer | Supersonic Imagine | 2012 | — |
| −4 | K130499 | IU22 Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2013 | 2 |
| −3 | K132304 | Epiq Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2013 | 36 |
| −2 | K160807 | EPIQ 5 Diagnostic Ultrasound System, EPIQ 7Diagnostic Ultrasound System… | Philips Ultrasound, Inc. | 2016 | 33 |
| −1 | K162329 | CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound System | Philips Health Care | 2016 | 3 |
| 0 (this device) | K250030 | Flash Ultrasound System 5100 Point of Care | Philips Ultrasound, LLC | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1989 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 5 | 0 |
| 1995 | 5 | 0 |
| 1996 | 3 | 1 |
| 1997 | 6 | 0 |
| 1998 | 10 | 3 |
| 1999 | 11 | 0 |
| 2000 | 13 | 1 |
| 2001 | 10 | 0 |
| 2002 | 15 | 0 |
| 2003 | 10 | 0 |
| 2004 | 11 | 1 |
| 2005 | 7 | 0 |
| 2006 | 4 | 0 |
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYN device
- #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K162329 | CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound SystemSummary page 5…trasonic, diagnostic 892.1570 ITX Device Class: Class II Review Panel: Radiology Predicate Device: K162329, CX50 and Sparq Ultrasound Diagnostic Systems Reference Device: K222648, 5000 Compact Series Ultrasound Systems… | 2016 |
| † | K222648 | 5000 Compact Series Ultrasound Systems (Ultrasound System 5500 G, Ultrasound System 5500… | 2022 |
| † | K231190 | EPIQ Series Diagnostic Ultrasound System | 2023 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (6 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 3, 2025 | Class III | Z-2414-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2386-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2413-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2369-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| May 28, 2025 | Class II | Z-2068-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 6 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
- K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 6 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
- K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
- K960330InstatrakVisualization Technology, Inc. · 1996 · 6 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
Nearest 5 of 6; all are on the web page.
What does this lineage say?
- The line of descent spans 30 years and passes through 7 companies; a new generation was cleared every 11 months on median.
- 49 of the 138 ancestors have a recall of any class; 6 have a Class I recall.
- 5 other devices also cite its predicate K162329.
How does its category compare?
All IYN devicesIYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · IYN
57in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYN
124in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYN
10in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · IYN
610in 2025
MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 26 | 16 | 3 |
| 2008 | 25 | 18 | 12 | 2 |
| 2009 | 18 | 29 | 8 | 9 |
| 2010 | 29 | 67 | 10 | 17 |
| 2011 | 39 | 46 | 8 | 18 |
| 2012 | 43 | 37 | 3 | 20 |
| 2013 | 51 | 60 | 8 | 18 |
| 2014 | 56 | 39 | 2 | 48 |
| 2015 | 54 | 55 | 3 | 116 |
| 2016 | 60 | 56 | 2 | 145 |
| 2017 | 64 | 83 | 0 | 146 |
| 2018 | 59 | 85 | 7 | 85 |
| 2019 | 47 | 79 | 0 | 59 |
| 2020 | 57 | 72 | 4 | 67 |
| 2021 | 46 | 105 | 8 | 213 |
| 2022 | 53 | 105 | 10 | 186 |
| 2023 | 44 | 115 | 4 | 229 |
| 2024 | 50 | 119 | 1 | 266 |
| 2025 | 57 | 124 | 10 | 610 |
| 2026 | 44 | 107 | 2 (1) | — |
What is it indicated for?
The Flash Ultrasound System 5100 Point of Care (5100 POC and 5100 POC Pro) is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal Adult Cephalic Cardiac Adult Cardiac Pediatric Cerebral Vascular Fetal/OB Lung Musculoskeletal (Conventional) Musculoskeletal (Superficial) Neonatal Cephalic Ophthalmic Other: Carotid Other: GYN Pediatric Peripheral Vessel Small Organ (Breast…
Read the full statement (198 words)
The Flash Ultrasound System 5100 Point of Care (5100 POC and 5100 POC Pro) is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal
Adult Cephalic
Cardiac Adult
Cardiac Pediatric
Cerebral Vascular
Fetal/OB
Lung
Musculoskeletal (Conventional)
Musculoskeletal (Superficial)
Neonatal Cephalic
Ophthalmic
Other: Carotid
Other: GYN
Pediatric
Peripheral Vessel
Small Organ (Breast, Thyroid, Testicle)
Trans-Esophageal (Cardiac)
Trans-Vaginal
Modes of operation include: B Mode (or 2D), Color Doppler, Color Power Angio Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-Mode (including anatomical M-Mode), Harmonics (Tissue), and Combined modes.
The clinical environments where the Flash Ultrasound System 5100 POC may be used are: hospitals, surgery centers, clinics, physicians’ offices, diagnostic centers, critical care and emergency room environments, and point-of-care clinical settings for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces the user’s responsibility for sound clinical judgment and best clinical procedure.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K250030 · printed Nov 6, 2026
Cite this page
510(k) Report. “K250030: Flash Ultrasound System 5100 Point of Care.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K250030
Report an extraction error