510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Concentrated Platelet Admin. Set

K840212510(k) TraditionalCleared Mar 6, 198447-day reviewClass II21 CFR 880.5440Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

American Mcgaw · FPA Set, Administration, Intravascular

Cleared on Mar 6, 1984 after a 47-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 6 of the 18 devices in its family, counting itself, have a Class I recall. 17 later devices descend from it. It has been recalled 8 times.

Decided
Mar 6, 1984
Review
47 days
Class
Class II
Product code
FPA
Recalls
8
Predicates
0

17 later devices descend from it, 1 citing it directly. 6 of the 18 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1997: 1 descendant cleared2006: 1 descendant cleared, 1 with a Class I recall12014: 2 descendants cleared, 1 with a Class I recall12020: 3 descendants cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12009: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12022: 3 descendants cleared2016: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant cleared2021: 1 descendant cleared19801990200020102020
  1. This deviceK840212 Concentrated Platelet Admin. SetAmerican Mcgaw · 1984
  2. Descendant · cites it directly · 16 of its ownK972678 Millennium CRT Infusion Pump and Millennium CRT Infusion Pump…Mcgaw, Inc. · 1997
  3. Descendant · 2 generations on · 15 of its ownK062700 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Recalled (Class I)
  4. Descendant · 3 generations on · 6 of its ownK140838 IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014
  5. Descendant · 3 generations on · 4 of its ownK083689 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · Recalled (Class I)
  6. Descendant · 4 generations on · 3 of its ownK170595 IV Administration SetsB.Braun Medical, Inc. · 2017 · Recalled

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

1997: 12006: 12009: 12013: 12014: 22015: 12016: 12017: 12020: 32021: 12022: 32025: 1
1997high 3 (2020) · low 0 (1998)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 17 · See all
  • K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration SetsMcgaw, Inc. · 1997 · cites it directly16descendants
  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on15descendants
  • K140838IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014 · 3 generations on6descendants
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · 3 generations on4descendants
  • K170595IV Administration SetsB.Braun Medical, Inc. · 2017 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K840212Concentrated Platelet Admin. SetAmerican Mcgaw1984—
+1K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration…Mcgaw, Inc.1997—
+2K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2006Class I
+3K140838IV Administration Set with Hand PumpB.Braun Medical, Inc.2014—
+3K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2009Class I
+4K170595IV Administration SetsB.Braun Medical, Inc.201719

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
200611
200911
201311
201421
201511
201610
201710
202031
202110
202230
202510
FIG. 1 — Lineage of K840212, 1984–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (17)

See all 17

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration SetsMcgaw, Inc.199716

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2025Class IIZ-0616-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0617-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0619-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0620-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Nov 20, 2019Class IIZ-0936-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing
Nov 20, 2019Class IIZ-0937-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing
Nov 20, 2019Class IIZ-0938-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing
Nov 20, 2019Class IIZ-0940-2020Nonconforming Material/ComponentPotential for leakage at the joint between the blood filters and tubing

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K142036Infusomat Space Volumetric Infusion Pump Administration SetsB.Braun Medical, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · 4 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K972678 (Mcgaw, Inc., 1997) has gone furthest, with 16 descendants of its own.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—
Cite this page

510(k) Report. “K840212: Concentrated Platelet Admin. Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K840212

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.