510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

SpaceStation MRI

K191910510(k) TraditionalCleared Mar 12, 2020239-day reviewClass II21 CFR 880.5725Recalls 8 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B. Braun Melsugen AG · MRZ Accessories, Pump, Infusion

Cleared on Mar 12, 2020 as substantially equivalent to K062699 (B.Braun Medical, Inc., 2006) and 2 other predicates after a 239-day review. Its family tree reaches back 4 generations to 1990. 5 of the 31 devices in its family, counting itself, have a Class I recall. It has been recalled 8 times, including 3 Class I recalls.

Decided
Mar 12, 2020
Review
239 days
Class
Class II
Product code
MRZ
Recalls
8 (3 Class I)
Predicates
3

Its longest line of descent runs back 4 generations to K904518 (1990). 30 ancestors in all, the oldest K820133 (1982). No later device cites it as a predicate yet. 5 of the 31 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2006: 1 other ancestor, 1 with a Class I recall12003: 3 other ancestors2002: 1 other ancestor, 1 with a Class I recall11997: 1 other ancestor1995: 2 other ancestors, 1 with a Class I recall11999: 1 other ancestor1998: 1 other ancestor1994: 4 other ancestors1993: 1 other ancestor1992: 2 other ancestors1991: 1 other ancestor1989: 2 other ancestors1986: 1 other ancestor1984: 2 other ancestors1982: 1 other ancestor1996: 1 other ancestor1990: 1 other ancestorK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK022575 · Perfusor Compact with FM System · B.Braun Medical, Inc. · 2002K062699 · Perfusor Space Infusion Syringe Pump System · B.Braun Medical, Inc. · 200619801990200020102020
Other ancestors, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 5 other ancestors, 2 with a Class I recall23 generations back: 13 other ancestors4 generations back: 6 other ancestors K062699 · Perfusor Space Infusion Syringe Pump System · B.Braun Medical, Inc. · 2006K022575 · Perfusor Compact with FM System · B.Braun Medical, Inc. · 2002K994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0
  1. Root · 4 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 3 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 2 generations backK022575 Perfusor Compact with FM SystemB.Braun Medical, Inc. · 2002
  4. 1 generation backK062699 Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc. · 2006
  5. This deviceK191910 SpaceStation MRIB. Braun Melsugen AG · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−3K994375Horizon LightB.Braun Medical, Inc.2000Class I
−2K022575Perfusor Compact with FM SystemB.Braun Medical, Inc.2002—
−1K062699Perfusor Space Infusion Syringe Pump SystemB.Braun Medical, Inc.2006—
0 (this device)K191910SpaceStation MRIB. Braun Melsugen AG2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198420
198610
198920
199010
199110
199220
199310
199440
199521
199610
199710
199810
199910
200211
200330
200611
FIG. 1 — Lineage of K191910, 1990–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRZ device

  1. #1 K193654 BlueStar Rxscore 7.5 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062699Perfusor Space Infusion Syringe Pump System
Summary page 6
…where the summary is organized in 3 sub-sections which correspond to each device bundled within the subject 510(k): • Section A) MRI conditional labeling changes to K062699 Perfusor Space Infusion Syringe Pump (See pages 2 - 4) • Section B) MRI conditional labeling changes to K062700 Infus…
B.Braun Medical, Inc.2006—
*K062700Infusomat Space Volumetric Infusion Pump System
Summary page 6
…ection A) MRI conditional labeling changes to K062699 Perfusor Space Infusion Syringe Pump (See pages 2 - 4) • Section B) MRI conditional labeling changes to K062700 Infusomat Space Volumetric Infusion Pump (See pages 5 - 7) • Section C) Traditional 510(k) SpaceStation MRI (See page…
B.Braun Medical, Inc.2006Class I
*K030323Mri-Caddy
Summary page 13
…Classification Panel: General Hospital C02) Legally Marketed Predicate Device(s) MRI-Caddy (MIPM Mammendorfer Institut für Physik und Medizin GmbH) - K030323 C03) Device Description The SpaceStation MRI is a movable MRI Rack System for the operation of up to four B.Braun Sp…
Mammendorfer Institute For Physics & Medicine GmbH2003—

Also cleared against K062699: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 7, 2024Class IZ-2972-2024Component design/selectionExtension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms.…
Jun 14, 2024Class IIZ-2435-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Jun 14, 2024Class IIZ-2433-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Jun 14, 2024Class IIZ-2434-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Jun 14, 2024Class IIZ-2437-2024Device DesignPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility…
Sep 21, 2023Class IZ-0123-2024Component design/selectionUpstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent…
Sep 21, 2023Class IZ-0124-2024Component design/selectionUpstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent…
Feb 7, 2022Class IIZ-0819-2022Under Investigation by firmUnits not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy, overinfusion, or underinfusion.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Direct predicate · ◆ in the 2006 ancestors bar of FIG. 11 Class I recall
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 30 years and passes through 3 companies; a new generation was cleared every 9.4 years on median.
  • 14 of the 30 ancestors have a recall of any class; 4 have a Class I recall.
  • 3 other devices also cite its predicate K062699.

How does its category compare?

All MRZ devices

MRZ Accessories, Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MRZ

1in 2023

2007: 22008: 42009: 22010: 22011: 12012: 22013: 12014: 22015: 12016: 22017: 22018: 12019: 12020: 22021: 12022: 12023: 1
2007high 4 (2008) · low 0 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRZ

96in 2008

2008: 96 days median
2007high 96 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MRZ

5in 2019

2009: 12018: 12019: 5
2007high 5 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MRZ

35in 2025

2011: 22012: 12015: 1172016: 552017: 182018: 462019: 362020: 442021: 612022: 392023: 272024: 412025: 35
2007high 117 (2015) · low 0 (2007)2025

MAUDE reports naming product code MRZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
200849600
20092—10
20102—00
20111—02
20122—01
20131—00
20142—00
20151—0117
20162—055
20172—018
20181—146
20191—536
20202—044
20211—061
20221—039
20231—027
20240—041
20250—035
20260—0—

What is it indicated for?

The SpaceStation MRI is a MRI (Magnetic Resonance Imaging) System Rack for the operation of Space Infusion Pumps during MRI examinations (MRI procedures) of adult, pediatric or neonatal patients. The product is intended to be used by qualified healthcare professionals. The Perfusor® Space Infusion Syringe Pump System includes an external transportable electronic infusion syringe pump and pump accessories. The system is intended for use on adults…

Read the full statement (291 words)

The SpaceStation MRI is a MRI (Magnetic Resonance Imaging) System Rack for the operation of Space Infusion Pumps during MRI examinations (MRI procedures) of adult, pediatric or neonatal patients.

The product is intended to be used by qualified healthcare professionals.

The Perfusor® Space Infusion Syringe Pump System includes an external transportable electronic infusion syringe pump and pump accessories. The system is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral and enteral fluids through clinically accepted routes of administration. These routes include, but are not limited to intravenous, intra-arterial, subcutaneous, epidural, irrigation/ablation, and enteral. The system is used for the delivery of medications indicated for infusion therapy including but not limited to drugs like anesthetics, sedatives, analgesics, catecholamines, anticoagulants etc., blood and blood components, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Perfusor® Space Infusion Syringe Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments (only road ambulances).

The Infusomat® Space Volumetric Infusion Pump System includes an external transportable electronic volumetric infusion pump, dedicated administration sets, and pump accessories. The system is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral and enteral fluids through clinically accepted routes of administration. These routes include, but are not limited to intravenous, irrigation/ablation, and enteral. The system is used for the delivery of medications indicated for infusion therapy including but not limited to colloids and cristalloids, blood and blood components, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments. (only road ambulances).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191910: SpaceStation MRI.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191910

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026