510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Agilia Infusion System

K121613510(k) TraditionalCleared May 1, 2013334-day reviewClass II21 CFR 880.5725Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Kabi AG · FRN Pump, Infusion

Cleared on May 1, 2013 as substantially equivalent to K062700 (B.Braun Medical, Inc., 2006) and 1 other predicate after a 334-day review. Its family tree reaches back 5 generations to 1990. 6 of the 25 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
May 1, 2013
Review
334 days
Class
Class II
Product code
FRN
Recalls
2 (1 Class I)
Predicates
2

Its longest line of descent runs back 5 generations to K904518 (1990). 22 ancestors in all, the oldest K820133 (1982). 2 later devices descend from it, 2 citing it directly. 6 of the 25 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors2001: 1 other ancestor, 1 with a Class I recall11997: 1 other ancestor1995: 1 other ancestor, 1 with a Class I recall11994: 3 other ancestors1993: 1 other ancestor1992: 2 other ancestors1991: 1 other ancestor1989: 1 other ancestor1986: 1 other ancestor1984: 2 other ancestors1982: 1 other ancestor2022: 2 descendants clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK023189 · Vista Basic with FM System · B.Braun Medical, Inc. · 2003K062700 · Infusomat Space Volumetric Infusion Pump System · B.Braun Medical, Inc. · 2006 · Class I recallK083689 · Infusomat Space Volumetric Infusion Pump System · B.Braun Medical, Inc. · 2009 · Class I recall19801990200020102020
  1. Root · 5 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 4 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 3 generations backK023189 Vista Basic with FM SystemB.Braun Medical, Inc. · 2003
  4. 2 generations backK062700 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Recalled (Class I)
  5. 1 generation backK083689 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · Recalled (Class I)
  6. This deviceK121613 Agilia Infusion SystemFresenius Kabi AG · 2013
  7. Descendant · cites it directly · 0 of its ownK210075 Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG · 2022
  8. Descendant · cites it directly · 0 of its ownK210073 Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat…Fresenius Kabi AG · 2022

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K210075Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG · 2022 · cites it directly0descendants
  • K210073Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration…Fresenius Kabi AG · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−4K994375Horizon LightB.Braun Medical, Inc.2000Class I
−3K023189Vista Basic with FM SystemB.Braun Medical, Inc.2003—
−2K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2006Class I
−1K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2009Class I
0 (this device)K121613Agilia Infusion SystemFresenius Kabi AG2013—
+1K210075Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG2022—
+1K210073Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines…Fresenius Kabi AG2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198420
198610
198910
199110
199220
199310
199430
199511
199710
200111
200320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202220
FIG. 1 — Lineage of K121613, 1990–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062700Infusomat Space Volumetric Infusion Pump System
Summary page 1
…system Classification Name: Infusion pumps Intravascular administration sets 3. PREDICATE DEVICES K062700/K083689 B. Braun.Medical, Inc., Infusomat® Space Volumetric Infusion Pump System 4. DEVICE DESCRIPTION The…
B.Braun Medical, Inc.2006Class I
*K083689Infusomat Space Volumetric Infusion Pump System
Summary page 1
…Classification Name: Infusion pumps Intravascular administration sets 3. PREDICATE DEVICES K062700/K083689 B. Braun.Medical, Inc., Infusomat® Space Volumetric Infusion Pump System 4. DEVICE DESCRIPTION The Agilia…
B.Braun Medical, Inc.2009Class I

Also cleared against K062700: 4 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210075Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG20220
K210073Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration…Fresenius Kabi AG20220

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 24, 2019Class IZ-2064-2019OtherThe firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
Dec 2, 2017Class IIZ-0362-2019Software designFresenius Kabi initiated a mandatory software upgrade of the Volumat MC Agilia Volumetric Infusion Pump because it was designed to improve customer experience while preventing common software malfunctions.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 2 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 3 companies; a new generation was cleared every 3.6 years on median.
  • 13 of the 22 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K210075 (Fresenius Kabi AG, 2022) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K062700.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Agilia Infusion System is a transportable equipment intended to be used in

healthcare facilities environment by healthcare trained professionals according to

hospital protocols on adults, pediatrics and neonates human patients to administer via a

single channel or mounted on a multiple channels rack accessory:

-Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral

nutrition) and medications (including but not limited to diluted drugs, chemotherapy)

through clinically accepted IV routes of administration.

- Transfusion of blood and blood derivatives products. Note that for transfusion

procedures, use of a special administration set is required.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121613: Agilia Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121613

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.