LTV 1000

K984056510(k) TraditionalCleared Apr 12, 1999150-day reviewClass II21 CFR 868.5895Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pulmonetic Systems, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Apr 12, 1999 as substantially equivalent to K981971 (Bird Products Corp., 1998) after a 150-day review. Its family tree reaches back 2 generations to 1984. 16 of the 45 devices in its family, counting itself, have a Class I recall. 40 later devices descend from it. It has been recalled 12 times.

Decided
Apr 12, 1999
Review
150 days
Class
Class II
Product code
CBK
Recalls
12
Predicates
1

Its longest line of descent runs back 2 generations to K841279 (1984). 4 ancestors in all. 40 later devices descend from it, 7 citing it directly. 16 of the 45 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1990: 1 other ancestor2007: 1 descendant cleared, 1 with a Class I recall12004: 1 descendant cleared2009: 3 descendants cleared, 2 with a Class I recall22008: 2 descendants cleared, 1 with a Class I recall12003: 1 descendant cleared2017: 2 descendants cleared, 2 with a Class I recall22010: 2 descendants cleared, 1 with a Class I recall12012: 6 descendants cleared, 1 with a Class I recall12016: 4 descendants cleared, 1 with a Class I recall12011: 3 descendants cleared, 2 with a Class I recall22019: 3 descendants cleared, 2 with a Class I recall22014: 2 descendants cleared2013: 2 descendants cleared, 1 with a Class I recall12024: 4 descendants cleared2025: 2 descendants cleared2021: 1 descendant cleared, 1 with a Class I recall12018: 1 descendant cleared, 1 with a Class I recall1Reference device: K962517 · Model 7700 Quantum Pressure Support Ventilator PSV · Healthdyne Technologies, Inc. · 1997Reference device: K811102 · Siemens-Elema AB Servo Ventilator 900 C · Siemens Elema AB · 1981Reference device: K905244 · Model LP6 Ventilator, Modification · Aequitron Medical, Inc. · 1990K841279 · BEAR 33 · Bear Medical Systems, Inc. · 1984K981971 · Tbird VS/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume Ventilator with Pressure Limit Model 00-004130, Bear 3 · Bird Products Corp. · 199819801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 7 descendants, 4 with a Class I recall42 generations forward: 12 descendants, 5 with a Class I recall53 generations forward: 13 descendants, 4 with a Class I recall44 generations forward: 6 descendants, 2 with a Class I recall25 generations forward: 2 descendants, 1 with a Class I recall1 K981971 · Tbird VS/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume Ventilator with Pressure Limit Model 00-004130, Bear 3 · Bird Products Corp. · 1998K841279 · BEAR 33 · Bear Medical Systems, Inc. · 19840+5
  1. Root · 2 generations backK841279 BEAR 33Bear Medical Systems, Inc. · 1984
  2. 1 generation backK981971 Tbird VS/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume…Bird Products Corp. · 1998
  3. This deviceK984056 LTV 1000Pulmonetic Systems, Inc. · 1999
  4. Descendant · cites it directly · 29 of its ownK022750 PLV Continuum Ventilator, Model P1000Respironics California, Inc. · 2003 · Recalled
  5. Descendant · cites it directly · 26 of its ownK083526 Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · Recalled (Class I)
  6. Descendant · 2 generations on · 19 of its ownK093416 Trilogy 200 VentilatorRespironics, Inc. · 2010 · Recalled (Class I)
  7. Descendant · cites it directly · 12 of its ownK070594 Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · Recalled (Class I)
  8. Descendant · 2 generations on · 11 of its ownK111610 Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · Recalled (Class I)

Showing 5 of 40 descendants. See all descendants

Descendants cleared per year

2003: 12004: 12007: 12008: 22009: 32010: 22011: 32012: 62013: 22014: 22016: 42017: 22018: 12019: 32021: 12024: 42025: 2
2003high 6 (2012) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 40 · See all
  • K022750PLV Continuum Ventilator, Model P1000Respironics California, Inc. · 2003 · cites it directly29descendants
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · cites it directly26descendants
  • K093416Trilogy 200 VentilatorRespironics, Inc. · 2010 · 2 generations on19descendants
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · cites it directly12descendants
  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · 2 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K841279BEAR 33Bear Medical Systems, Inc.1984—
−1K981971Tbird VS/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume Ventilator…Bird Products Corp.1998—
0 (this device)K984056LTV 1000Pulmonetic Systems, Inc.1999—
+1K022750PLV Continuum Ventilator, Model P1000Respironics California, Inc.20032
+1K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group2009Class I
+2K093416Trilogy 200 VentilatorRespironics, Inc.2010Class I
+1K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc.2007Class I
+2K111610Trilogy Series Ventilator with OximetryRespironics, Inc.2011Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200711
200821
200932
201021
201132
201261
201321
201420
201641
201722
201811
201932
202111
202440
202520
FIG. 1 — Lineage of K984056, 1984–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

16 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K981971Tbird VS/Avs Volume Ventilator, Models 15568, 15570 , LP10 Volume Ventilator with…
Summary page 1
…LTV 1000 Common Name Ventilator Classification Name: Ventilator, Continuous (Respirator) PREDICATE DEVICE The primary predicate device is the TBird Homecare Ventilator (K981971), manufactured by Bird Products Corp. of Palm Springs, CA. DESCRIPTION The LTV1000 ventilator is intended to provide continuous…
Bird Products Corp.1998—
†K962517Model 7700 Quantum Pressure Support Ventilator PSVHealthdyne Technologies, Inc.1997—
†K811102Siemens-Elema AB Servo Ventilator 900 CSiemens Elema AB1981—
†K905244Model LP6 Ventilator, ModificationAequitron Medical, Inc.1990—

Also cleared against K981971: 1 other device →

What descends from it? (40)

See all 40

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022750PLV Continuum Ventilator, Model P1000Respironics California, Inc.200329
K083526Trilogy 100 Ventilator Class IRespironics Inc., Sleep & Home Respiratory Group200926
K070594Palmtop Ventilator, Models PTV-8, PTV-10 Class IPulmonetic Systems, Inc.200712
K082724Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc.20089
K090888Newport HT70 Family of Ventilators Class INewport Medical Instruments, Inc.20092
K034032PLV Continuum Ventilator, Model P2000Respironics California, Inc.20041
K072885Carina System, Model 8419350; Carina Basic Device, Model 5704110 Class IDraeger Medical AG & Co. KG20080

Has anything in its family been recalled? (16 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 22, 2008Not classifiedZ-1463-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1469-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1468-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1467-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1464-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1470-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1466-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1465-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Nov 9, 2005Not classifiedZ-0184-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0186-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0185-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Sep 8, 2004Not classifiedZ-1485-04OtherVentilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K072885Carina System, Model 8419350; Carina Basic Device, Model 5704110Draeger Medical AG & Co. KG · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 15 Class I recalls
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K093416Trilogy 200 VentilatorRespironics, Inc. · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K121623Bipap a 40 Ventilatory Support SystemRespironics, Inc. · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 13 Class I recalls
  • K181170Trilogy Evo UniversalRespironics, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 112 Class I recalls
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls

Nearest 12 of 16 shown.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 14.3 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K022750 (Respironics California, Inc., 2003) has gone furthest, with 29 descendants of its own.
  • 1 other device also cites its predicate K981971.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 Ibs), who require the following types of ventilatory support:

* Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).

* Assist/Control, SIMV, or CPAP modes of ventilation.

The ventilator is suitable for use in institutional, home, and transport settings.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K984056: LTV 1000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K984056

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026