510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Philips Integris Allura

K002016510(k) TraditionalCleared Sep 6, 200065-day reviewClass II21 CFR 892.1650Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Sep 6, 2000 after a 65-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 6 of the 77 devices in its family, counting itself, have a Class I recall. 76 later devices descend from it. It has been recalled 13 times.

Decided
Sep 6, 2000
Review
65 days
Class
Class II
Product code
OWB
Recalls
13
Predicates
0

76 later devices descend from it, 3 citing it directly. 6 of the 77 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2003: 2 descendants cleared, 1 with a Class I recall12010: 2 descendants cleared2006: 1 descendant cleared2009: 1 descendant cleared2013: 5 descendants cleared, 1 with a Class I recall12014: 6 descendants cleared2017: 4 descendants cleared, 2 with a Class I recall22018: 6 descendants cleared2021: 3 descendants cleared2016: 5 descendants cleared, 1 with a Class I recall12011: 1 descendant cleared2025: 11 descendants cleared2012: 1 descendant cleared2022: 6 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12023: 3 descendants cleared2026: 9 descendants cleared2024: 5 descendants cleared200020052010201520202025
  1. This deviceK002016 Philips Integris AlluraPhilips Medical Systems North America, Inc. · 2000
  2. Descendant · cites it directly · 51 of its ownK031836 Philips Fresco Release 1Philips Medical Systems North America Co. · 2003
  3. Descendant · 2 generations on · 49 of its ownK062650 EP Navigator SoftwarePhilips Medical Systems · 2006
  4. Descendant · 3 generations on · 41 of its ownK101311 EP Navigator R3Philips Medical Systems North America Co. · 2010
  5. Descendant · 4 generations on · 39 of its ownK111245 Heartnavigator Release 1Philips Medical Systems Nederland B.V. · 2011
  6. Descendant · 5 generations on · 32 of its ownK121781 Echonavigator Release 1Philips Medical Systems Nederland B.V. · 2012

Showing 5 of 76 descendants. See all descendants

Descendants cleared per year

2003: 22006: 12009: 12010: 22011: 12012: 12013: 52014: 62016: 52017: 42018: 62020: 52021: 32022: 62023: 32024: 52025: 112026: 9
2003high 11 (2025) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 76 · See all
  • K031836Philips Fresco Release 1Philips Medical Systems North America Co. · 2003 · cites it directly51descendants
  • K062650EP Navigator SoftwarePhilips Medical Systems · 2006 · 2 generations on49descendants
  • K101311EP Navigator R3Philips Medical Systems North America Co. · 2010 · 3 generations on41descendants
  • K111245Heartnavigator Release 1Philips Medical Systems Nederland B.V. · 2011 · 4 generations on39descendants
  • K121781Echonavigator Release 1Philips Medical Systems Nederland B.V. · 2012 · 5 generations on32descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K002016Philips Integris AlluraPhilips Medical Systems North America, Inc.2000—
+1K031836Philips Fresco Release 1Philips Medical Systems North America Co.2003—
+2K062650EP Navigator SoftwarePhilips Medical Systems2006—
+3K101311EP Navigator R3Philips Medical Systems North America Co.2010—
+4K111245Heartnavigator Release 1Philips Medical Systems Nederland B.V.2011—
+5K121781Echonavigator Release 1Philips Medical Systems Nederland B.V.2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200321
200610
200910
201020
201110
201210
201351
201460
201651
201742
201860
202051
202130
202260
202330
202450
2025110
202690
FIG. 1 — Lineage of K002016, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (76)

See all 76

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031836Philips Fresco Release 1Philips Medical Systems North America Co.200351
K033737Allura Xper FD20 Class IPhilips Medical Systems North America, Inc.200326
K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201024

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 22, 2024Class IIZ-3167-2024Device DesignPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Oct 4, 2023Class IIZ-0227-2024Device DesignPhilips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
Dec 30, 2019Class IIZ-1487-2020Component design/selectionA capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Dec 30, 2019Class IIZ-1488-2020Component design/selectionA capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Nov 3, 2015Class IIZ-0726-2016Device DesignThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Nov 3, 2015Class IIZ-0727-2016Device DesignThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Nov 3, 2015Class IIZ-0728-2016Device DesignThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Nov 3, 2015Class IIZ-0729-2016Device DesignThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Aug 8, 2014Class IIZ-0022-2015Radiation Control for Health and Safety ActPhilips had discovered through a field service testing where a device failed to comply with a Federal standard.
Dec 14, 2006Not classifiedZ-0366-2007OtherMovement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
Oct 12, 2005Not classifiedZ-0206-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.
Oct 12, 2005Not classifiedZ-0207-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.
Jul 27, 2005Not classifiedZ-0022-06Other1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · Cites it directly · ◆ in the 2003 descendants bar of FIG. 113 Class I recalls
  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 7 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • Of its descendants, K031836 (Philips Medical Systems North America Co., 2003) has gone furthest, with 51 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—
Cite this page

510(k) Report. “K002016: Philips Integris Allura.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K002016

Report an extraction error